Literature DB >> 20187854

Evaluation of the new body fluid mode on the Sysmex XE-5000 for counting leukocytes and erythrocytes in cerebrospinal fluid and other body fluids.

Robert de Jonge1, Rob Brouwer, Marieke T de Graaf, Ronald L Luitwieler, Cherina Fleming, Magda de Frankrijker-Merkestijn, Peter A E Sillevis Smitt, Joke G Boonstra, Jan Lindemans.   

Abstract

BACKGROUND: We evaluated the body fluid (BF) mode on the new Sysmex XE-5000 analyzer.
METHODS: Red (RBC) and white blood cell (WBC) (differential) counts of BFs (139 patient samples and 87 normal samples) were measured and compared to the Fuchs-Rosenthal chamber and stained cytospin slides.
RESULTS: Extended cell counting using the BF mode was noted to have an improved WBC detection limit (CV(20)%) of 10 x 10(6)/L. Excellent agreement with the manual method was observed for most BFs [mean bias +2 to 6 x 10(6)/L for cerebrospinal fluid (CSF) and -1 to 12 x 10(6)/L for other fluids]. In CSF, the BF-mode counted more WBC (polymorphic nuclear cells) compared with the manual method (mean bias +5 to 6 x 10(6)/L), especially in samples with low cell counts (<20 x 10(6)/L). Carry over was negligible (mostly <0.17%) and linearity was excellent (mean bias <5%). The reference ranges for CSF (n=87) were RBC 0 x 10(6)/L, WBC and mononuclear <7 x 10(6)/L, and polymorph nucleated cells <3 x 10(6)/L.
CONCLUSIONS: The BF mode on the Sysmex XE-5000 offers rapid and accurate RBC and WBC (differential) counts in clinically relevant concentration ranges in CSF and other fluids. In addition, the exclusion of high fluorescent cells, such as mesothelial cells and macrophages from WBC counting may reduce the number of manual analyses in pleural fluids and ascites.

Entities:  

Mesh:

Year:  2010        PMID: 20187854     DOI: 10.1515/CCLM.2010.108

Source DB:  PubMed          Journal:  Clin Chem Lab Med        ISSN: 1434-6621            Impact factor:   3.694


  10 in total

1.  Automated Cerebrospinal Fluid Cell Counts Using the New Body Fluid Mode of Sysmex UF-1000i.

Authors:  Sabrina Buoro; Sara Apassiti Esposito; MariaGrazia Alessio; Alberto Crippa; Cosimo Ottomano; Giuseppe Lippi
Journal:  J Clin Lab Anal       Date:  2015-08-24       Impact factor: 2.352

2.  Technology and new fluorescence flow cytometry parameters in hematological analyzers.

Authors:  Borros M Arneth; Mario Menschikowki
Journal:  J Clin Lab Anal       Date:  2014-05-05       Impact factor: 2.352

3.  Automated analysis for differentiating leukocytes in body fluids using the software "biological liquid application" on ADVIA2120/2120i hematology analyzer.

Authors:  Alessandra Falda; Paolo Doretto
Journal:  J Clin Lab Anal       Date:  2018-05-25       Impact factor: 2.352

4.  Use of the Sysmex XT-4000i hematology analyzer in the differentiation of cerebrospinal fluid cells in children.

Authors:  Beata Zelazowska-Rutkowska; Janusz Zak; Malgorzata Wojtkowska; Justyna Zaworonek; Bogdan Cylwik
Journal:  J Clin Lab Anal       Date:  2018-11-28       Impact factor: 2.352

5.  Effect of sample processing and time delay on cell count and chemistry tests in cerebrospinal fluid collected from drainage systems.

Authors:  Ana Mlinarić; Željka Vogrinc; Zrinka Drenšek
Journal:  Biochem Med (Zagreb)       Date:  2018-10-15       Impact factor: 2.313

6.  A comparison of the analysis of 3 types of body fluids using the XN-350 hematology analyzer versus light microscopy assessment.

Authors:  Jiwon Lee; Younggeun Cho; Han-Sung Kim; Hee Jung Kang; Miyoung Kim; Young Kyung Lee
Journal:  Medicine (Baltimore)       Date:  2021-03-19       Impact factor: 1.817

7.  Screening for malignant tumor cells in serous effusions with an automatic hematology analyzer using a novel diagnostic algorithm.

Authors:  Dehua Sun; Xinru Mao; Taixue An; Xiaojing He; Kai Qiu; Yuhong Luo; Zheyuan Qin; Yongjian He; Tie Xiong; Houmei Feng; Jin Li; Lei Zheng
Journal:  Ann Transl Med       Date:  2022-03

8.  Automated and manual microscopic analyses for leukocyte differential counts in exudative pleural effusions: Real-world disagreement and clinical application.

Authors:  Jaehee Lee; Yu Kyung Kim; Ji Eun Park; Yong Hoon Lee; Sun Ha Choi; Hyewon Seo; Seung Soo Yoo; Shin Yup Lee; Seung-Ick Cha; Jae Yong Park; Chang Ho Kim
Journal:  Medicine (Baltimore)       Date:  2022-09-16       Impact factor: 1.817

9.  Automated Analysis of Cerebrospinal Fluid Cells Using Commercially Available Blood Cell Analysis Devices-A Critical Appraisal.

Authors:  Manfred Wick; Catharina C Gross; Hayrettin Tumani; Brigitte Wildemann; Martin Stangel
Journal:  Cells       Date:  2021-05-18       Impact factor: 6.600

10.  Evaluation of cell count and classification capabilities in body fluids using a fully automated Sysmex XN equipped with high-sensitive Analysis (hsA) mode and DI-60 hematology analyzer system.

Authors:  Hiroyuki Takemura; Tomohiko Ai; Konobu Kimura; Kaori Nagasaka; Toshihiro Takahashi; Koji Tsuchiya; Haeun Yang; Aya Konishi; Kinya Uchihashi; Takashi Horii; Yoko Tabe; Akimichi Ohsaka
Journal:  PLoS One       Date:  2018-04-26       Impact factor: 3.240

  10 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.