Performance evaluation of the digital cell imaging analyzer DI-60 integrated into the fully automated Sysmex XN hematology analyzer system.

Yoko Tabe, Takemasa Yamamoto, Imiko Maenou, Rie Nakai, Mayumi Idei, Takashi Horii, Takashi Miida, Akimichi Ohsaka.   

Abstract

BACKGROUND: The XN-Series (Sysmex, Kobe, Japan) have been equipped with the automated digital cell imaging analyzer DI-60, which provides complete automation of the sample processing with automated complete blood counts (CBC), slide making/staining, and digital scanning with cell pre-classification. The aim of this study was to evaluate the efficacy of the XN-Series as an integrated blood cell analysis system.
METHODS: White blood cell (WBC) morphological analysis by the DI-60 was evaluated using 232 blood samples from patients. Routine analysis of a total of 2000 blood samples has been performed to evaluate the processing ability of the XN-Series connected to the DI-60.
RESULTS: The overall analysis accuracy of pre-classification of WBC by the DI-60 was 88.4%. Good correlation was observed between final results of the DI-60 analysis and manual differentiation with high sensitivity and specificity for blasts and immature granulocytes. The sample processing time of the XN-Series, from automated CBC to cell pre-classification, was 38±1 min/single run and 165±12 min/500 CBC samples run (slide preparation rate 15.6%) with no sample hold-up at the DI-60.
CONCLUSIONS: The automated morphological analysis capability of the DI-60 has potential usefulness in the integrated automated hematology analysis system of XN-Series.

Entities:  

Mesh:

Year:  2015        PMID: 25153399     DOI: 10.1515/cclm-2014-0445

Source DB:  PubMed          Journal:  Clin Chem Lab Med        ISSN: 1434-6621            Impact factor:   3.694


  4 in total

1.  Digital Morphology Analyzer Sysmex DI-60 vs. Manual Counting for White Blood Cell Differentials in Leukopenic Samples: A Comparative Assessment of Risk and Turnaround Time.

Authors:  Minjeong Nam; Sumi Yoon; Mina Hur; Gun Hyuk Lee; Hanah Kim; Mikyoung Park; Hyeong Nyeon Kim
Journal:  Ann Lab Med       Date:  2022-07-01       Impact factor: 4.941

2.  Red and white blood cell morphology characterization and hands-on time analysis by the digital cell imaging analyzer DI-60.

Authors:  Oh Joo Kweon; Yong Kwan Lim; Mi-Kyung Lee; Hye Ryoun Kim
Journal:  PLoS One       Date:  2022-04-27       Impact factor: 3.240

3.  Evaluation of cell count and classification capabilities in body fluids using a fully automated Sysmex XN equipped with high-sensitive Analysis (hsA) mode and DI-60 hematology analyzer system.

Authors:  Hiroyuki Takemura; Tomohiko Ai; Konobu Kimura; Kaori Nagasaka; Toshihiro Takahashi; Koji Tsuchiya; Haeun Yang; Aya Konishi; Kinya Uchihashi; Takashi Horii; Yoko Tabe; Akimichi Ohsaka
Journal:  PLoS One       Date:  2018-04-26       Impact factor: 3.240

4.  How Reproducible Is the Data from Sysmex DI-60 in Leukopenic Samples?

Authors:  Sumi Yoon; Mina Hur; Gun Hyuk Lee; Minjeong Nam; Hanah Kim
Journal:  Diagnostics (Basel)       Date:  2021-11-23
  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.