Mélanie Machiels1, Peng Jin2, Christianne H van Gurp2, Jeanin E van Hooft3, Tanja Alderliesten2, Maarten C C M Hulshof2.
Abstract
BACKGROUND: To investigate the feasibility and geometric accuracy of carina-based registration for CBCT-guided setup verification in esophageal cancer IGRT, compared with current practice bony anatomy-based registration.Entities:
Keywords: Carina-based registration; Esophageal cancer; Fiducial markers; Image-guided radiotherapy; Radiotherapy; Setup verification
Mesh:
Year: 2018 PMID: 29562919 PMCID: PMC5861603 DOI: 10.1186/s13014-018-0986-1
Source DB: PubMed Journal: Radiat Oncol ISSN: 1748-717X Impact factor: 3.481
Patient, tumor and marker characteristics
| Patient | Age (yr) | Sex | Tumor length (cm) | TNM | Histology | Tumor location | Marker type | No. of CBCTs | No. of markers | ||
|---|---|---|---|---|---|---|---|---|---|---|---|
| At placement | Visible in pCT | Visible in CBCTs | |||||||||
| 1 | 56 | M | 5 | T3N1M0 | AD | distal | SM | 7 | 2 | 1 | 1 |
| 2 | 79 | M | 8 | T3N1M0 | AD | distal | SM | 7 | 3 | 2 | 1 |
| 3 | 62 | M | 8 | T3N1M0 | AD | distal | VM | 7 | 3 | 3 | 1 |
| 4 | 61 | M | 4 | T3N1M0 | PDC | distal | SM | 7 | 3 | 3 | 3 → 2 |
| 5a | 73 | M | 3 | T2N1M0 | SCC | distal | HG | 8a | 0 | 0 | 0a |
| 6 | 57 | V | 3 | T2N0M0 | SCC | distal | SM | 8 | 4 | 4 | 4 |
| 7 | 63 | M | 2 | T3N2M0 | AD | distal | SM | 8 | 4 | 4 | 4 |
| 8a | 83 | M | 8 | T2N2M0 | AD | distal | VM | 28a | 5 | 5 | 0a |
| 9 | 65 | V | 4 | T2N0M0 | SCC | mid | VM | 8 | 4 | 3 | 3 |
| 10a | 57 | M | 5 | T2N0M0 | AD | distal | VM | 7a | 3 | 2 | 0a |
| 11 | 64 | M | 5 | T3N2M0 | SCC | distal | SM | 12 | 3 | 2 | 2 |
| 12 | 70 | M | 13 | T2N1M0 | SCC | proximal | SM | 8 | 3 | 3 | 3 |
| 13 | 67 | V | 5 | T3N2M0 | SCC | mid | VM | 23 | 4 | 4 | 4 |
| 14 | 73 | M | 4 | T2N0M0 | SCC | proximal | VM | 25 | 3 | 3 | 1 |
| 15a | 71 | M | 6 | T4aN2M0 | SCC | distal | HG | 8a | 6 | 5 | 0a |
| 16 | 84 | M | 5 | T3N1M0 | AD | distal | HG | 8 | 3 | 3 | 3 → 1 |
| 17 | 61 | M | 7 | T3N1M0 | AD | distal | VM | 8 | 3 | 3 | 3 |
| 18a | 69 | M | 7 | T3N1M0 | AD | distal | HG | 23a | 3 | 3 | 0a |
| 19 | 45 | M | 6 | T2N1M0 | AD | distal | SM | 7 | 3 | 3 | 3 → 2 |
| 20 | 63 | M | 3 | T2N0M0 | AD | distal | SM | 11 | 3 | 2 | 2 |
| 21 | 79 | V | 6 | T3N2M0 | AD | distal | VM | 8 | 4 | 4 | 3 |
| 22 | 59 | V | 10 | T3N2M0 | AD | distal | VM | 9 | 4 | 4 | 4 |
| 23 | 61 | M | 6 | T3N2M0 | AD | distal | VM | 12 | 3 | 3 | 3 |
| 24 | 69 | M | 5 | T3N0M0 | AD | distal | VM | 12 | 4 | 4 | 2 |
| 25 | 69 | M | 8 | T2N1MO | AD | distal | VM | 8 | 4 | 4 | 4 |
| 26a | 76 | M | 3 | T2N1M0 | SCC | distal | HG | 8a | 3 | 2 | 0a |
| 27 | 67 | V | 4.5 | T1N0M0 | SCC | proximal | SM | 8 | 3 | 3 | 3 |
| 28 | 65 | M | 4 | T3N2M0 | SCC | mid | VM | 9 | 5 | 4 | 4 |
| 29 | 75 | M | 5 | T3N1M0 | SCC | distal | VM | 9 | 3 | 2 | 2 |
| 30 | 51 | M | 17 | T3N2M0 | AD | prox-mid-dist | SM | 7 | 3 | 3 | 2 |
| Total | 318 → 236 | 101 | 91 | 65 → 61 | |||||||
Patient numbering is consistent with that in Jin et al. [4]
Abbreviations: AD adenocarcinoma, SCC squamous cell carcinoma, PDC poorly differentiated carcinoma, SM solid, rigid golden marker, VM Visicoil marker, HG radiopaque hydrogel, pCT planning computed tomography, CBCT cone-beam computed tomography
aMarker not visible on CBCT, subsequent exclusion of patient from data analysis
Fig. 1Illustration of the systematic error (SE, the length of the arrow) and the SD of random errors (sdRE, the length of semi-major/minor axis of the ellipse) of each marker position relative to the carina, projected onto the coronal (left) and sagittal (right) views of the esophagus. Note: the amplitudes of the errors are not scaled to the esophagus
Fig. 2Overlay of the pCT (purple) and CBCT (green) in XVI for patient 9. Arrows indicate fiducial markers, registered on the bony anatomy (left column) and the carina (right column). In axial (a), sagittal (b), and coronal view (c). First row = general carina image in coronal view. A = anterior, R = right, L = left, and F = inferior. Red contour = PTV, blue contour = CTV
Fig. 3Example of inadequate target coverage after registration. Overlay of the pCT (purple) and CBCT (green) in XVI for patient 20, registered on the bony anatomy (left column) and the carina (right column). In axial (first row) and coronal view (second row). A = anterior, R = right, and F = inferior. Red contour = PTV, blue contour = CTV
Fig. 4Absolute systematic error (SE) and standard deviation (SD) of random errors (sdRE) of the marker position relative to the bony anatomy and carina. The SD of SEs (Σ) and the root mean square of sdREs (σ) are also depicted
The group mean (M), SD of systematic errors (Σ), root mean square of SDs of random errors (σ), and the estimated margin required to compensate for the interfractional tumor position variation in each orthogonal direction for the four subgroups of markers and all markers (Overall)
| LR (mm) | CC (mm) | AP (mm) | |||||
|---|---|---|---|---|---|---|---|
| Bony anatomy | Carina | Bony anatomy | Carina | Bony anatomy | Carina | ||
| Proximal ( | M | 0.1 | −1.3 | −1.2 | 0.5 | − 0.7 | 1.0 |
| Σ | 1.5 | 1.7 | 4.1 | 3.3 | 1.9 | 3.3 | |
| σ | 1.3 | 1.6 | 1.5 | 1.9 | 1.2 | 1.6 | |
| 2.5 Σ + 0.7σ | 4.4 | 5.5 | 11.3 | 9.7 | 5.5 | 9.3 | |
| Mid-thoracic ( | M | −1.1 | −1.5 | −2.3 | −1.0 | 1.2 | 0.8 |
| Σ | 3.1 | 2.7 | 2.9 | 1.7 | 3.2 | 1.8 | |
| σ | 1.5 | 1.6 | 2.0 | 2.8 | 2.3 | 2.3 | |
| 2.5 Σ + 0.7σ | 8.9 | 8.0 | 8.6 | 6.1 | 9.6 | 6.1 | |
| Distal ( | M | 0.4 | 0.8 | −0.8 | 0.5 | 0.5 | −0.8 |
| Σ | 1.9 | 2.2 | 4.2 | 4.1 | 1.9 | 4.5 | |
| σ | 1.9 | 2.7 | 2.5 | 2.2 | 1.4 | 3.1 | |
| 2.5 Σ + 0.7σ | 6.1 | 7.4 | 12.1 | 11.7 | 5.7 | 13.4 | |
| Cardia ( | M | −1.1 | −1.2 | −1.2 | 0.9 | −0.1 | 1.7 |
| Σ | 5.4 | 6.1 | 4.9 | 4.0 | 1.9 | 3.6 | |
| σ | 4.3 | 5.1 | 3.2 | 3.7 | 2.4 | 4.7 | |
| 2.5 Σ + 0.7σ | 16.4 | 18.9 | 14.6 | 12.5 | 6.4 | 12.2 | |
| Overall ( | M | −0.2 | −0.3 | −1.2 | 0.3 | 0.3 | 0.2 |
| Σ | 2.9 | 3.3 | 4.1 | 3.6 | 2.2 | 3.9 | |
| σ | 2.4 | 3.0 | 2.4 | 2.6 | 1.8 | 3.1 | |
| 2.5 Σ + 0.7σ | 8.9 | 10.4 | 11.9 | 10.8 | 6.8 | 11.8 | |