| Literature DB >> 29520938 |
Dennis N Stanley1, Karl Rasmussen1, Neil Kirby1, Nikos Papanikolaou1, Alonso N Gutiérrez1,2.
Abstract
INTRODUCTION: A robust image quality assurance and analysis methodology for image-guided localization systems is crucial to ensure the accurate localization and visualization of target tumors. In this study, the long-term stability of selected image parameters was assessed and evaluated for the cone-beam computed tomography (CBCT) mode, planar radiographic kV mode, and the radiographic MV mode of an Elekta VersaHD.Entities:
Keywords: Elekta XVI; Elekta iView; image Quality; image stability; temporal trending
Mesh:
Year: 2018 PMID: 29520938 PMCID: PMC5978980 DOI: 10.1002/acm2.12289
Source DB: PubMed Journal: J Appl Clin Med Phys ISSN: 1526-9914 Impact factor: 2.102
Figure 1The X‐ray volume imaging (XVI) guidance system and iViewGT image system of the Elekta VersaHD radiation delivery system are shown. (a) a‐Si flat panel detector of the XVI, (b) kV X‐ray source of the XVI and (c) iViewGT imaging panel.
Figure 2The a‐Si flat panel detector of the Elekta XVI system is shown with the (a) QCkV‐1 image quality phantom.
Figure 3The iVieiwGT flat panel detector is shown with the (a) QC‐3 image quality phantom in measurement position.
Image quality scanning parameters for the Elekta VersaHD XVI kV‐CBCT
| Small | Medium | |
|---|---|---|
| CBCT mode | Head and Neck S20 | Pelvis M20 |
| Start angle | 25 | 180 |
| Stop angle | 180 | 180 |
| Reconstructed volume | 512 × 512 | 512 × 512 |
| kV collimator | S20 | M20 |
| kV filter | F0 | F1 |
| kV | 100 | 120 |
| mA per frame | 10 | 40 |
| ms per frame | 10 | 40 |
| Frames | 183 | 660 |
Normalized standard deviations for all evaluated metrics
| Planar radiographic | |||||||
|---|---|---|---|---|---|---|---|
| kV | MV | ||||||
| f30(lp/mm) | 0.004 | Noise | 0.048 | f30(lp/mm) | 0.015 | Noise | 0.005 |
| f40(lp/mm) | 0.003 | CNR | 0.024 | f40(lp/mm) | 0.009 | CNR | 0.021 |
| f50(lp/mm) | 0.003 | f50(lp/mm) | 0.017 | ||||
| kVp | 0.006 | Dose(μGy) | 0.030 | ||||
| CBCT | |||||||
| Small | |||||||
| High contrast spatial resolution (lp/mm) | HU constancy | Noise | Uniformity | ||||
| f30 | 0.083 | Lung(PMP) | 0.049 | Mean | 0.029 | Mean | 0.028 |
| f40 | 0.058 | Water(Poly) | 0.021 | Sigma | 0.059 | Sigma | 0.053 |
| f50 | 0.056 | Air | 0.025 | ||||
| Medium | |||||||
| High contrast spatial resolution (lp/mm) | HU constancy | Noise | Uniformity | ||||
| f30 | 0.093 | Lung(PMP) | 0.010 | Mean | 0.010 | Mean | 0.011 |
| f40 | 0.043 | Water(Poly) | 0.016 | Sigma | 0.041 | Sigma | 0.032 |
| f50 | 0.037 | Air | 0.006 | ||||
Figure 4(top) The run chart of the normalized f50 values measured with the QCkV‐1 phantom and PIPSpro software is shown for the kV planar radiographic mode of the Elekta XVI. (bottom) The run chart of the normalized image dose values measured with the Unfors RaySafe Xi R/F detector is shown for the kV planar radiographic mode of the Elekta XVI.
Image quality consistency thresholds for the planar radiographic modalities
| kV | MV | ||||
|---|---|---|---|---|---|
| Warning (%) | Action (%) | Warning (%) | Action (%) | ||
| f30(lp/mm) | 1 | 3 | f30(lp/mm) | 2 | 4 |
| f40(lp/mm) | 1 | 3 | f40(lp/mm) | 1 | 3 |
| f50(lp/mm) | 1 | 3 | f50(lp/mm) | 2 | 4 |
| Noise | 5 | 10 | Noise | 1 | 2 |
| CNR | 3 | 6 | CNR | 2 | 4 |
| Dose | 3 | 6 | |||
| kVp | 1 | 2 | |||
Sample size of 30 measurements.
Image quality consistency thresholds for the small(S20) CBCT
| Warning (%) | Action (%) | |
|---|---|---|
| High contrast spatial resolution(lp/mm) | ||
| f30 | 8 | 16 |
| f40 | 5 | 11 |
| f50 | 5 | 11 |
| HU constancy | ||
| Lung(PMP) | 5 | 9 |
| Water(Poly) | 3 | 6 |
| Air | 2 | 5 |
| Uniformity | ||
| Mean | 3 | 6 |
| Sigma | 5 | 10 |
| Noise | ||
| Mean | 3 | 6 |
| Sigma | 6 | 11 |
Sample size of 30 measurements.
Image quality consistency thresholds for the Medium (M20) CBCT
| Warning (%) | Action (%) | |
|---|---|---|
| High contrast spatial resolution(lp/mm) | ||
| f30 | 9 | 18 |
| f40 | 4 | 9 |
| f50 | 4 | 8 |
| HU constancy | ||
| Lung(PMP) | 2 | 4 |
| Water(Poly) | 2 | 4 |
| Air | 1 | 3 |
| Uniformity | ||
| Mean | 1 | 3 |
| Sigma | 3 | 6 |
| Noise | ||
| Mean | 1 | 3 |
| Sigma | 4 | 8 |
Sample size of 30 measurements.
Comparison of Image quality action thresholds for the planar radiographic modalities between the XVI and OBI
| XVI (%) | OBI | |
|---|---|---|
| kV | ||
| High contrast spatial resolution(lp/mm) | ||
| f30 | 3 | 4 |
| f40 | 3 | 3 |
| f50 | 3 | 3 |
| Dose | 6 | 2 |
| kVp | 2 | 2 |
| MV | ||
| High contrast spatial resolution(lp/mm) | ||
| f30 | 4 | 4 |
| f40 | 3 | 3 |
| f50 | 4 | 3 |
*Data from Stanley et al.7
Comparison of image quality action thresholds for the Small (S20) and Full‐Fan CBCT
| XVI (%) | OBI | |
|---|---|---|
| High contrast spatial resolution(lp/mm) | ||
| f30 | 16 | 18 |
| f40 | 11 | 18 |
| f50 | 11 | 16 |
| HU constancy | ||
| Lung(PMP) | 9 | 6 |
| Water(Poly) | 6 | 12 |
| Uniformity | 10 | 12 |
| Noise | 11 | 12 |
*Data from Stanley et al.7
Comparison of image quality action thresholds for the Medium (M20) and Half‐Fan CBCT
| XVI (%) | OBI | |
|---|---|---|
| High contrast spatial resolution(lp/mm) | ||
| f30 | 18 | 20 |
| f40 | 9 | 22 |
| f50 | 8 | 34 |
| HU constancy | ||
| Lung(PMP) | 4 | 8 |
| Water(Poly) | 4 | 12 |
| Uniformity | 6 | 17 |
| Noise | 8 | 16 |
*Data from Stanley et al.7