Suresh Rana1, Colton Eckert1, Hardev Singh1, Yuanshui Zheng2, Michael Chacko1, Mark Storey3, John Chang3.
Abstract
PURPOSE: The purpose of this study was twofold: (a) report the long-term monthly quality assurance (QA) dosimetry results of the uniform scanning beam delivery system, and (b) derive the machine-specific tolerances based on the statistic process control (SPC) methodology and compare them against the AAPM TG224 recommended tolerances.Entities:
Keywords: monthly QA; proton therapy; statistical analysis
Mesh:
Substances:
Year: 2020 PMID: 32741135 PMCID: PMC7497929 DOI: 10.1002/acm2.12990
Source DB: PubMed Journal: J Appl Clin Med Phys ISSN: 1526-9914 Impact factor: 2.102
Fig. 1The I chart and mR chart of dose output difference in TR1 and TR4.
Fig. 2The I chart of flatness difference for selected monthly quality assurance beams in different treatment rooms.
Fig. 3The I chart of symmetry difference for selected monthly quality assurance beams in different treatment rooms.
Fig. 4The I chart of range difference for selected monthly quality assurance beams in different treatment rooms.
Monthly quality assurance (QA) results of TG224 recommended dosimetry parameters. The results included the measurements performed in four treatment rooms (TR1, TR2, TR3, and TR4) over a period of 6 yr (Feb 2014–Jan 2020). The 2σ and 3σ values were calculated using the SPC method, which included individuals and moving range control charts.
| Dosimetry parameters | Mean | 2σ | 3σ | TG224 |
|---|---|---|---|---|
| Dose output | ||||
| TR1 | −0.3% | ±0.9% | ±1.3% | ±2% |
| TR2 | 0.0% | ±1.4% | ±2.1% | ±2% |
| TR3 | −0.2% | ±1.0% | ±1.6% | ±2% |
| TR4 | −0.5% | ±0.9% | ±1.3% | ±2% |
| Range 10 cm | ||||
| TR1 | 0.9 mm | ±0.2 mm | ±0.3 mm | ±1 mm |
| TR2 | 1.0 mm | ±0.2 mm | ±0.3 mm | ±1 mm |
| TR3 | 0.9 mm | ±0.2 mm | ±0.3 mm | ±1 mm |
| TR4 | 0.6 mm | ±0.3 mm | ±0.4 mm | ±1 mm |
| Range 16 cm | ||||
| TR1 | 0.2 mm | ±0.3 mm | ±0.4 mm | ±1 mm |
| TR2 | 0.2 mm | ±0.6 mm | ±0.8 mm | ±1 mm |
| TR3 | −0.2 mm | ±0.4 mm | ±0.6 mm | ±1 mm |
| TR4 | 0.4 mm | ±0.3 mm | ±0.4 mm | ±1 mm |
| Range 22 | ||||
| TR1 | 0.2 mm | ±0.3 mm | ±0.5 mm | ±1 mm |
| TR2 | 0.5 mm | ±0.3 mm | ±0.5 mm | ±1 mm |
| TR3 | 0.1 mm | ±0.5 mm | ±0.8 mm | ±1 mm |
| TR4 | 0.2 mm | ±0.5 mm | ±0.7 mm | ±1 mm |
| Range 28 | ||||
| TR1 | 0.0 mm | ±0.6 mm | ±0.9 mm | ±1 mm |
| TR2 | 0.2 mm | ±1.3 mm | ±1.9 mm | ±1 mm |
| TR3 | −0.3 mm | ±0.9 mm | ±1.4 mm | ±1 mm |
| TR4 | 0.2 mm | ±0.6 mm | ±0.9 mm | ±1 mm |
| Flatness R10M6 | ||||
| TR1 | −0.3% | ±0.4% | ±0.5% | ±2% |
| TR2 | −0.1% | ±0.5% | ±0.7% | ±2% |
| TR3 | −0.7% | ±0.5% | ±0.8% | ±2% |
| TR4 | −0.3% | ±0.9% | ±0.5% | ±2% |
| Flatness R16M10 | ||||
| TR1 | −0.1% | ±0.5% | ±0.7% | ±2% |
| TR2 | −0.1% | ±0.5% | ±0.7% | ±2% |
| TR3 | 0.0% | ±0.4% | ±0.7% | ±2% |
| TR4 | −0.6% | ±0.5% | ±0.8% | ±2% |
| Flatness R22M8 | ||||
| TR1 | −0.1% | ±0.4% | ±0.6% | ±2% |
| TR2 | −0.5% | ±0.6% | ±0.9% | ±2% |
| TR3 | 0.0% | ±0.6% | ±0.9% | ±2% |
| TR4 | −0.4% | ±0.5% | ±0.7% | ±2% |
| Flatness R28M14 | ||||
| TR1 | −0.5% | ±0.6% | ±0.9% | ±2% |
| TR2 | −0.9% | ±0.7% | ±1.0% | ±2% |
| TR3 | −0.2% | ±0.8% | ±1.2% | ±2% |
| TR4 | −0.5% | ±0.8% | ±1.1% | ±2% |
| Symmetry R10M6 | ||||
| TR1 | −0.8% | ±0.3% | ±0.4% | ±2% |
| TR2 | −0.1% | ±0.5% | ±0.7% | ±2% |
| TR3 | 0.2% | ±0.8% | ±1.2% | ±2% |
| TR4 | 0.4% | ±0.3% | ±0.5% | ±2% |
| Symmetry R16M10 | ||||
| TR1 | 0.0% | ±0.5% | ±0.8% | ±2% |
| TR2 | 0.1% | ±0.7% | ±1.1% | ±2% |
| TR3 | −0.1% | ±0.6% | ±0.9% | ±2% |
| TR4 | 0.7% | ±0.6% | ±0.9% | ±2% |
| Symmetry R22M8 | ||||
| TR1 | 0.5% | ±0.6% | ±0.8% | ±2% |
| TR2 | 0.2% | ±0.9% | ±1.4% | ±2% |
| TR3 | 0.4% | ±0.6% | ±0.9% | ±2% |
| TR4 | 0.7% | ±0.6% | ±1.0% | ±±2% |
| Symmetry R28M14 | ||||
| TR1 | 0.3% | ±0.5% | ±0.7% | ±2% |
| TR2 | 0.2% | ±0.8% | ±1.2% | ±2% |
| TR3 | 0.6% | ±0.7% | ±1.1% | ±2% |
| TR4 | 0.4% | ±0.7% | ±1.0% | ±2% |
Fig. 5A box‐whisker plot showing the difference in range of four monthly quality assurance beams. For a given range, the results are combined from all treatment rooms.
Fig. 6Histograms showing the difference in dose output, range, flatness, and symmetry of all monthly quality assurance beams in four treatment rooms.