| Literature DB >> 29500480 |
Andres Kaalep1, Terez Sera2,3, Sjoerd Rijnsdorp4, Maqsood Yaqub5, Anne Talsma6, Martin A Lodge7, Ronald Boellaard8,9,10.
Abstract
PURPOSE: The objective of this study was to explore the feasibility of harmonising performance for PET/CT systems equipped with time-of-flight (ToF) and resolution modelling/point spread function (PSF) technologies. A second aim was producing a working prototype of new harmonising criteria with higher contrast recoveries than current EARL standards using various SUV metrics.Entities:
Keywords: EARL accreditation; Harmonisation; PET/CT; Performance; Quantification
Mesh:
Year: 2018 PMID: 29500480 PMCID: PMC5993859 DOI: 10.1007/s00259-018-3977-4
Source DB: PubMed Journal: Eur J Nucl Med Mol Imaging ISSN: 1619-7070 Impact factor: 9.236
Acquisition and reconstruction settings for the initial 15 reconstruction modes
| Reconstruction mode | Post filter width (mm) | Q.Clear β value | Pixel size (mm) | Slice thickness (mm) | Long frame duration (s) | Medium frame duration (s) | Short frame duration (s) |
|---|---|---|---|---|---|---|---|
| GE - A | N/A | 200 | 2.73 | 3.27 | 300 | 120 | 60 |
| GE - B | 0 | N/A | 2.73 | 3.27 | 300 | 120 | 60 |
| GE - C | N/A | 350 | 2.73 | 3.27 | 300 | 120 | 60 |
| GE - D | 3 | N/A | 2.73 | 3.27 | 300 | 120 | 60 |
| GE - E | N/A | 800 | 2.73 | 3.27 | 300 | 120 | 60 |
| GE - F | 5 | N/A | 2.73 | 3.27 | 300 | 120 | 60 |
| GE - G | 6.4 | N/A | 2.73 | 3.27 | 300 | 120 | 60 |
| Philips - A | N/A | N/A | 2.00 | 2.00 | 301 | 120 | 60 |
| Philips - B | N/A | N/A | 4.00 | 4.00 | 301 | 120 | 60 |
| Siemens 1 - A | 0 | N/A | 2.04 | 2.00 | 300 | 120 | 60 |
| Siemens 1 - B | 0 | N/A | 1.59 | 2.00 | 300 | 120 | 60 |
| Siemens 1 - C | 3 | N/A | 2.04 | 2.00 | 300 | 120 | 60 |
| Siemens 1 - D | 5 | N/A | 2.04 | 2.00 | 300 | 120 | 60 |
| Siemens 1 - E | 6.5 | N/A | 3.18 | 2.00 | 300 | 120 | 60 |
| Siemens 2 - A | 5 | N/A | 4.07 | 5.00 | 223 | 111 | 56 |
Fig. 1RC curves derived from 15 initial reconstruction modes using SUVmean (a), SUVmax (b) and SUVpeak (c) quantitative metrics. Only long acquisition time frame curves are displayed. GE (Q.Clear) – blue dashed lines, GE (non-Q.Clear) – blue solid lines, Philips – red solid lines, Siemens 1 – orange solid lines, Siemens 2 – green solid lines, current EARL specifications – black solid lines
Description of quantitative metrics used
| Metric | Description of metric |
|---|---|
| SUVmean | Ratio of image derived average radioactivity concentration within a region of interest and the whole body concentration of the injected radioactivity |
| SUVmax | Ratio of image derived maximum (single pixel) radioactivity concentration within a region of interest and the whole body concentration of the injected radioactivity |
| SUVpeak | Ratio of image derived average radioactivity concentration within a 12 mm diameter spherical volume within the region of interest, positioned to yield the highest uptake, and the whole body concentration of the injected radioactivity |
| RC | Recovery Coefficient - the ratio between image derived and expected activity concentration |
| MCR* | Mean Contrast Recovery - mean RC of all spheres in corresponding reconstruction mode’s long duration acquisition. Parameter is indicative of reconstruction mode’s overall contrast recovery potential. |
| CoVMCR | Coefficient of Variation (SD/mean*100, %) of a group of MCR values. Parameter is indicative of RC curves’ alignment within a group. |
| CRVmedium* | Contrast Recovery Variability - Mean deviation of medium duration acquisition spheres’ RCs from the corresponding values of long duration aquisition. |
| CRVshort* | Contrast Recovery Variability - Mean deviation of short duration acquisition spheres’ RCs from the corresponding values of long duration aquisition. |
| CoVBG* | Coefficient of Variation (SD/mean *100, %) of measured activity concentration within the uniform background compartment of the phantom. Parameter is indicative of the noise present in the images. |
| Curvature | Long acquisition duration root-mean-square deviation of spheres’ RC values from RC value of the largest (37 mm) sphere. Parameter characterises the deviation of smaller spheres’ RC values which usually cause the RC-object size relation to assume a curved shape. |
| Absolute error | Long acquisition duration root-mean-square deviation of spheres’ RC values from unity. The parameter characterises the reconstruction mode’s ability to report accurate activity concentration values. |
| Curvature (excl. 10 mm sphere) | Same as "curvature" but excluding the smallest (10 mm) sphere. |
| Absolute error (excl. 10 mm sphere) | Same as "absolute error" but excluding the smallest (10 mm) sphere. |
*Quantitative metrics that were retrospectively used to determine harmonising cut-off criteria
Retrospectively determined quantitative cut-off criteria for the harmonising reconstructions
| SUVmean | SUVmax | |
|---|---|---|
| MCR | ±11% (0.77–0.96) | ±13% (1.01–1.31) |
| CRVmedium | 6% | 8% |
| CRVshort | 7% | 9% |
| Visual analysis | No excessive Gibbs and partial volume artefacts | |
| Noise | Background CoV ≤15% (high statistics acquisition) | |
Fig. 2RC curves derived from suggested harmonising reconstruction modes using SUVmean (a), SUVmax (b) and SUVpeak (c) quantitative metrics along with current EARL and possible new specifications. Only long acquisition time frame curves are displayed. GE (Q.Clear) – blue dashed lines, GE (non-Q.Clear) – blue solid lines, Philips – red solid lines, Siemens 1 – orange solid lines, Siemens 2 – green solid lines, current EARL specifications – black solid lines, possible new EARL specifications – black dashed lines
Analysis results of 15 initial reconstruction modes using a SUVmean quantitative metric. Values found to be outside of acceptable range during retrospective quantitative analysis, are coloured red
Analysis results of 15 initial reconstruction modes using a SUVmax quantitative metric. Values found to be outside of acceptable range during retrospective quantitative analysis, are coloured red
Analysis results of 15 initial reconstruction modes using SUVpeak quantitative metric
| Reconstruction mode | MCR | CRVmedium | CRVshort | Curvature | Absolute error | Curvature (excl. 10 mm sphere) | Absolute error (excl. 10 mm sphere) |
|---|---|---|---|---|---|---|---|
| GE - A | 0.848 | 3.9% | 3.7% | 0.334 | 0.287 | 0.187 | 0.153 |
| GE - B | 0.833 | 3.4% | 5.7% | 0.381 | 0.310 | 0.237 | 0.179 |
| GE - C | 0.840 | 2.3% | 3.6% | 0.359 | 0.302 | 0.211 | 0.166 |
| GE - D | 0.823 | 3.9% | 6.3% | 0.389 | 0.320 | 0.248 | 0.191 |
| GE - E | 0.821 | 2.9% | 4.1% | 0.400 | 0.339 | 0.250 | 0.203 |
| GE - F | 0.784 | 3.3% | 5.8% | 0.404 | 0.346 | 0.272 | 0.223 |
| GE - G | 0.757 | 3.1% | 5.9% | 0.413 | 0.367 | 0.287 | 0.248 |
| Philips - A | 0.874 | 3.2% | 3.4% | 0.328 | 0.281 | 0.192 | 0.161 |
| Philips - B | 0.796 | 2.8% | 2.9% | 0.383 | 0.341 | 0.263 | 0.229 |
| Siemens 1 - A | 0.901 | 4.5% | 6.3% | 0.305 | 0.232 | 0.148 | 0.090 |
| Siemens 1 - B | 0.929 | 1.6% | 4.2% | 0.325 | 0.240 | 0.154 | 0.103 |
| Siemens 1 - C | 0.872 | 3.3% | 5.0% | 0.308 | 0.251 | 0.151 | 0.107 |
| Siemens 1 - D | 0.823 | 3.0% | 4.5% | 0.350 | 0.291 | 0.204 | 0.155 |
| Siemens 1 - E | 0.754 | 3.9% | 2.7% | 0.382 | 0.346 | 0.255 | 0.226 |
| Siemens 2 - A | 0.789 | 2.9% | 4.9% | 0.355 | 0.323 | 0.240 | 0.214 |
| Min | 0.754 | 1.6% | 2.7% | 0.305 | 0.232 | 0.148 | 0.090 |
| Max | 0.929 | 4.5% | 6.3% | 0.413 | 0.367 | 0.287 | 0.248 |
| Average | 0.830 | 3.2% | 4.6% | 0.361 | 0.305 | 0.220 | 0.177 |
| COVMCR | 6.0% |
Results of the analysis of five reconstruction modes considered for harmonisation using the SUVmean quantitative metric
| Reconstruction mode | MCR | CRVmedium | CRVshort | Curvature | Absolute error | Curvature (excl. 10 mm sphere) | Absolute error (excl. 10 mm sphere) |
|---|---|---|---|---|---|---|---|
| GE - E | 0.806 | 5.2% | 6.2% | 0.218 | 0.253 | 0.075 | 0.134 |
| GE - F | 0.770 | 5.3% | 5.3% | 0.228 | 0.277 | 0.120 | 0.183 |
| Philips - B | 0.800 | 2.7% | 2.3% | 0.236 | 0.271 | 0.124 | 0.165 |
| Siemens 1 - D | 0.816 | 2.9% | 5.1% | 0.197 | 0.222 | 0.097 | 0.138 |
| Siemens 2 - A | 0.804 | 3.0% | 4.4% | 0.203 | 0.238 | 0.100 | 0.150 |
| Min | 0.770 | 2.7% | 2.3% | 0.197 | 0.222 | 0.075 | 0.134 |
| Max | 0,816 | 5.3% | 6.2% | 0.236 | 0.277 | 0.124 | 0.183 |
| Average | 0.799 | 3.8% | 4.6% | 0.216 | 0.252 | 0.103 | 0.154 |
| COVMCR | 2.2% | ||||||
| EARL min | 0.570 | N/A | N/A | 0.282 | 0.466 | 0.198 | 0.393 |
| EARL max | 0.710 | N/A | N/A | 0.277 | 0.342 | 0.176 | 0.251 |
| EARL Average | 0.640 | N/A | N/A | 0.279 | 0.403 | 0.187 | 0.321 |
Results of the analysis of five reconstruction modes considered for harmonisation using the SUVmax quantitative metric
| Reconstruction mode | MCR | CRVmedium | CRVshort | Curvature | Absolute error | Curvature (excl. 10 mm sphere) | Absolute error (excl. 10 mm sphere) |
|---|---|---|---|---|---|---|---|
| GE - E | 1.036 | 7.2% | 7.7% | 0.212 | 0.178 | 0.041 | 0.119 |
| GE - F | 1.013 | 8.0% | 9.2% | 0.235 | 0.170 | 0.085 | 0.099 |
| Philips - B | 1.061 | 3.7% | 5.2% | 0.267 | 0.232 | 0.150 | 0.197 |
| Siemens 1 - D | 1.094 | 3.9% | 7.9% | 0.218 | 0.179 | 0.080 | 0.165 |
| Siemens 2 - A | 1.045 | 3.7% | 5.4% | 0.246 | 0.184 | 0.104 | 0.138 |
| Min | 1.013 | 3.7% | 5.2% | 0.212 | 0.170 | 0.041 | 0.099 |
| Max | 1.094 | 8.0% | 9.2% | 0.267 | 0.232 | 0.150 | 0.197 |
| Average | 1.050 | 5.3% | 7.1% | 0.236 | 0.189 | 0.092 | 0.144 |
| COVMCR | 2.9% | ||||||
| EARL min | 0.730 | N/A | N/A | 0.347 | 0.355 | 0.220 | 0.237 |
| EARL max | 0.970 | N/A | N/A | 0.339 | 0.236 | 0.176 | 0.121 |
| EARL Average | 0.850 | N/A | N/A | 0.342 | 0.277 | 0.198 | 0.142 |
Results of the analysis of five reconstruction modes considered for harmonisation using the SUVpeak quantitative metric
| Reconstruction mode | MCR | CRVmedium | CRVshort | Curvature | Absolute error | Curvature (excl. 10 mm sphere) | Absolute error (excl. 10 mm sphere) |
|---|---|---|---|---|---|---|---|
| GE - E | 0.821 | 2.9% | 4.1% | 0.400 | 0.339 | 0.250 | 0.203 |
| GE - F | 0.784 | 3.3% | 5.8% | 0.404 | 0.346 | 0.272 | 0.223 |
| Philips - B | 0.796 | 2.8% | 2.9% | 0.383 | 0.341 | 0.263 | 0.229 |
| Siemens 1 - D | 0.823 | 3.0% | 4.5% | 0.350 | 0.291 | 0.204 | 0.155 |
| Siemens 2 - A | 0.789 | 2.9% | 4.9% | 0.355 | 0.323 | 0.240 | 0.214 |
| Min | 0.784 | 2.8% | 2.9% | 0.350 | 0.291 | 0.204 | 0.155 |
| Max | 0.823 | 3.3% | 5.8% | 0.404 | 0.346 | 0.272 | 0.229 |
| Average | 0.803 | 3.0% | 4.4% | 0.378 | 0.328 | 0.246 | 0.205 |
| COVMCR | 2.2% |
Fig. 3RC curves derived from additional reconstructions using SUVmean (a), SUVmax (b) and SUVpeak (c) quantitative metrics along with proposed new specifications. GE (Q.Clear) – blue dashed lines, GE (non-Q.Clear) – blue solid lines, Philips – red solid lines, Siemens – orange solid lines, possible new EARL specifications – black dashed lines
Analysis results of 23 additional reconstructions using the SUVmean quantitative metric
| PET/CT system | MCR | CRVmedium | CRVshort | Curvature | Absolute error | Curvature (excl. 10 mm sphere) | Absolute error (excl. 10 mm sphere) |
|---|---|---|---|---|---|---|---|
| Ingenuity 1 | 0.820 | N/A | N/A | 0.213 | 0.249 | 0.106 | 0.145 |
| Ingenuity 2 | 0.694 | N/A | N/A | 0.276 | 0.365 | 0.164 | 0.263 |
| mCT Flow 1 | 0.691 | N/A | N/A | 0.303 | 0.368 | 0.196 | 0.270 |
| mCT Flow 2 | 0.711 | N/A | N/A | 0.298 | 0.339 | 0.190 | 0.242 |
| mCT Flow 3 | 0.816 | N/A | N/A | 0.193 | 0.231 | 0.079 | 0.136 |
| mCT 1 | 0.847 | N/A | N/A | 0.176 | 0.194 | 0.080 | 0.112 |
| mCT 2 | 0.786 | N/A | N/A | 0.194 | 0.250 | 0.115 | 0.181 |
| mCT 3 | 0.825 | N/A | N/A | 0.188 | 0.208 | 0.113 | 0.142 |
| mCT 4 | 0.765 | N/A | N/A | 0.174 | 0.262 | 0.091 | 0.195 |
| mCT 5 | 0.786 | N/A | N/A | 0.195 | 0.245 | 0.119 | 0.179 |
| mCT 6 | 0.811 | N/A | N/A | 0.136 | 0.207 | 0.078 | 0.161 |
| Discovery 710 1 | 0.847 | N/A | N/A | 0.153 | 0.182 | 0.079 | 0.120 |
| Discovery 710 2 | 0.793 | N/A | N/A | 0.217 | 0.254 | 0.129 | 0.174 |
| Discovery 710 1 Q.Clear 1 | 0.887 | N/A | N/A | 0.120 | 0.145 | 0.027 | 0.074 |
| Discovery 710 2 Q.Clear 2 | 0.817 | N/A | N/A | 0.211 | 0.236 | 0.110 | 0.146 |
| Discovery 710 3 Q.Clear 3 | 0.895 | N/A | N/A | 0.121 | 0.144 | 0.042 | 0.073 |
| GE Discovery MI 1 | 0.794 | N/A | N/A | 0.150 | 0.228 | 0.099 | 0.182 |
| GE Discovery MI 2 | 0.813 | N/A | N/A | 0.171 | 0.214 | 0.102 | 0.155 |
| GE Discovery MI 1 Q.Clear 1 | 0.857 | N/A | N/A | 0.081 | 0.151 | 0.055 | 0.129 |
| GE Discovery MI 2 Q.Clear 2 | 0.869 | N/A | N/A | 0.118 | 0.156 | 0.039 | 0.096 |
| GE Discovery IQ 1 | 0.817 | N/A | N/A | 0.219 | 0.244 | 0.077 | 0.123 |
| GE Discovery IQ 1 Q.Clear 1 | 0.818 | N/A | N/A | 0.221 | 0.246 | 0.069 | 0.118 |
| Vereos 1 | 0.757 | N/A | N/A | 0.191 | 0.277 | 0.087 | 0.195 |
| Min | 0.691 | 0.081 | 0.144 | 0.027 | 0.073 | ||
| Max | 0.895 | 0.303 | 0.368 | 0.196 | 0.270 | ||
| Average | 0.805 | 0.188 | 0.235 | 0.098 | 0.157 | ||
| COVMCR | 6.6% |
Analysis results of 23 additional reconstructions using the SUVmax quantitative metric
| PET/CT system | MCR | CRVmedium | CRVshort | Curvature | Absolute error | Curvature (excl. 10 mm sphere) | Absolute error (excl. 10 mm sphere) |
|---|---|---|---|---|---|---|---|
| Ingenuity 1 | 1.094 | N/A | N/A | 0.278 | 0.264 | 0.143 | 0.228 |
| Ingenuity 2 | 0.917 | N/A | N/A | 0.334 | 0.288 | 0.188 | 0.167 |
| mCT Flow 1 | 0.911 | N/A | N/A | 0.347 | 0.270 | 0.207 | 0.159 |
| mCT Flow 2 | 0.943 | N/A | N/A | 0.350 | 0.234 | 0.187 | 0.109 |
| mCT Flow 3 | 1.071 | N/A | N/A | 0.237 | 0.211 | 0.110 | 0.179 |
| mCT 1 | 1.118 | N/A | N/A | 0.185 | 0.179 | 0.057 | 0.179 |
| mCT 2 | 1.038 | N/A | N/A | 0.173 | 0.140 | 0.065 | 0.108 |
| mCT 3 | 1.098 | N/A | N/A | 0.168 | 0.148 | 0.082 | 0.151 |
| mCT 4 | 1.019 | N/A | N/A | 0.160 | 0.130 | 0.041 | 0.082 |
| mCT 5 | 1.033 | N/A | N/A | 0.176 | 0.127 | 0.067 | 0.092 |
| mCT 6 | 1.067 | N/A | N/A | 0.113 | 0.107 | 0.033 | 0.105 |
| Discovery 710 1 | 1.139 | N/A | N/A | 0.151 | 0.176 | 0.051 | 0.188 |
| Discovery 710 2 | 1.045 | N/A | N/A | 0.213 | 0.168 | 0.086 | 0.130 |
| Discovery 710 1 Q.Clear 1 | 1.172 | N/A | N/A | 0.085 | 0.189 | 0.054 | 0.207 |
| Discovery 710 2 Q.Clear 2 | 1.049 | N/A | N/A | 0.204 | 0.172 | 0.064 | 0.131 |
| Discovery 710 3 Q.Clear 3 | 1.154 | N/A | N/A | 0.114 | 0.184 | 0.042 | 0.200 |
| GE Discovery MI 1 | 1.055 | N/A | N/A | 0.105 | 0.100 | 0.032 | 0.095 |
| GE Discovery MI 2 | 1.066 | N/A | N/A | 0.179 | 0.142 | 0.065 | 0.125 |
| GE Discovery MI 1 Q.Clear 1 | 1.119 | N/A | N/A | 0.040 | 0.123 | 0.017 | 0.108 |
| GE Discovery MI 2 Q.Clear 2 | 1.124 | N/A | N/A | 0.107 | 0.157 | 0.039 | 0.168 |
| GE Discovery IQ 1 | 1.102 | N/A | N/A | 0.255 | 0.240 | 0.047 | 0.201 |
| GE Discovery IQ 1 Q.Clear 1 | 1.083 | N/A | N/A | 0.234 | 0.219 | 0.052 | 0.177 |
| Vereos 1 | 1.029 | N/A | N/A | 0.230 | 0.176 | 0.074 | 0.115 |
| Min | 0.911 | 0.040 | 0.100 | 0.017 | 0.082 | ||
| Max | 1.172 | 0.350 | 0.288 | 0.207 | 0.228 | ||
| Average | 1.063 | 0.193 | 0.180 | 0.078 | 0.148 | ||
| COVMCR | 6.3% |
Analysis results of 23 additional reconstructions using SUVpeak quantitative metric
| PET/CT system | MCR | CRVmedium | CRVshort | Curvature | Absolute error | Curvature (excl. 10 mm sphere) | Absolute error (excl. 10 mm sphere) |
|---|---|---|---|---|---|---|---|
| Ingenuity 1 | 0.789 | N/A | N/A | 0.376 | 0.341 | 0.246 | 0.218 |
| Ingenuity 2 | 0.736 | N/A | N/A | 0.405 | 0.383 | 0.284 | 0.267 |
| mCT Flow 1 | 0.737 | N/A | N/A | 0.439 | 0.390 | 0.324 | 0.280 |
| mCT Flow 2 | 0.750 | N/A | N/A | 0.476 | 0.379 | 0.353 | 0.263 |
| mCT Flow 3 | 0.797 | N/A | N/A | 0.393 | 0.328 | 0.274 | 0.217 |
| mCT 1 | 0.858 | N/A | N/A | 0.348 | 0.282 | 0.214 | 0.162 |
| mCT 2 | 0.812 | N/A | N/A | 0.347 | 0.302 | 0.225 | 0.188 |
| mCT 3 | 0.847 | N/A | N/A | 0.365 | 0.281 | 0.242 | 0.169 |
| mCT 4 | 0.781 | N/A | N/A | 0.326 | 0.313 | 0.198 | 0.192 |
| mCT 5 | 0.803 | N/A | N/A | 0.355 | 0.304 | 0.243 | 0.199 |
| mCT 6 | 0.827 | N/A | N/A | 0.297 | 0.269 | 0.184 | 0.163 |
| Discovery 710 1 | 0.829 | N/A | N/A | 0.357 | 0.301 | 0.234 | 0.188 |
| Discovery 710 2 | 0.794 | N/A | N/A | 0.398 | 0.342 | 0.274 | 0.227 |
| Discovery 710 1 Q.Clear 1 | 0.867 | N/A | N/A | 0.372 | 0.294 | 0.231 | 0.171 |
| Discovery 710 2 Q.Clear 2 | 0.824 | N/A | N/A | 0.413 | 0.344 | 0.276 | 0.221 |
| Discovery 710 3 Q.Clear 3 | 0.884 | N/A | N/A | 0.370 | 0.298 | 0.212 | 0.166 |
| GE Discovery MI 1 | 0.797 | N/A | N/A | 0.351 | 0.313 | 0.233 | 0.202 |
| GE Discovery MI 2 | 0.819 | N/A | N/A | 0.375 | 0.308 | 0.237 | 0.180 |
| GE Discovery MI 1 Q.Clear 1 | 0.838 | N/A | N/A | 0.328 | 0.285 | 0.200 | 0.166 |
| GE Discovery MI 2 Q.Clear 2 | 0.859 | N/A | N/A | 0.356 | 0.294 | 0.202 | 0.157 |
| GE Discovery IQ 1 | 0.814 | N/A | N/A | 0.407 | 0.342 | 0.263 | 0.210 |
| GE Discovery IQ 1 Q.Clear 1 | 0.831 | N/A | N/A | 0.412 | 0.336 | 0.258 | 0.199 |
| Vereos 1 | 0.803 | N/A | N/A | 0.381 | 0.320 | 0.251 | 0.199 |
| Min | 0.736 | 0.297 | 0.269 | 0.184 | 0.157 | ||
| Max | 0.884 | 0.476 | 0.390 | 0.353 | 0.280 | ||
| Average | 0.813 | 0.376 | 0.320 | 0.246 | 0.200 | ||
| COVMCR | 4.7% |
SUVmean, SUVmax and SUVpeak specifications proposed for harmonisation along with current EARL specifications
| Sphere diameter (mm) | Current EARL RC bandwidth | Proposed RC bandwidth | ||||
|---|---|---|---|---|---|---|
| SUVmean | SUVmax | SUVpeak | SUVmean | SUVmax | SUVpeak | |
| 37 | 0.76–0.89 | 0.95–1.16 | N/A | 0.85–1,00 | 1.05–1.29 | 0.99–1.07 |
| 28 | 0.72–0.85 | 0.91–1.13 | N/A | 0.82–0.97 | 1.01–1.26 | 0.95–1.07 |
| 22 | 0.63–0.78 | 0.83–1.09 | N/A | 0.80–0.99 | 1.01–1.32 | 0.90–1.09 |
| 17 | 0.57–0.73 | 0.73–1.01 | N/A | 0.76–0.97 | 1.00–1.38 | 0.75–0.99 |
| 13 | 0.44–0.60 | 0.59–0.85 | N/A | 0.63–0.86 | 0.85–1.22 | 0.45–0.69 |
| 10 | 0.27–0.43 | 0.34–0.57 | N/A | 0.39–0.61 | 0.52–0.88 | 0.27–0.41 |