Literature DB >> 25525180

Quantitative PET/CT scanner performance characterization based upon the society of nuclear medicine and molecular imaging clinical trials network oncology clinical simulator phantom.

John J Sunderland1, Paul E Christian2.   

Abstract

UNLABELLED: The Clinical Trials Network (CTN) of the Society of Nuclear Medicine and Molecular Imaging (SNMMI) operates a PET/CT phantom imaging program using the CTN's oncology clinical simulator phantom, designed to validate scanners at sites that wish to participate in oncology clinical trials. Since its inception in 2008, the CTN has collected 406 well-characterized phantom datasets from 237 scanners at 170 imaging sites covering the spectrum of commercially available PET/CT systems. The combined and collated phantom data describe a global profile of quantitative performance and variability of PET/CT data used in both clinical practice and clinical trials.
METHODS: Individual sites filled and imaged the CTN oncology PET phantom according to detailed instructions. Standard clinical reconstructions were requested and submitted. The phantom itself contains uniform regions suitable for scanner calibration assessment, lung fields, and 6 hot spheric lesions with diameters ranging from 7 to 20 mm at a 4:1 contrast ratio with primary background. The CTN Phantom Imaging Core evaluated the quality of the phantom fill and imaging and measured background standardized uptake values to assess scanner calibration and maximum standardized uptake values of all 6 lesions to review quantitative performance. Scanner make-and-model-specific measurements were pooled and then subdivided by reconstruction to create scanner-specific quantitative profiles.
RESULTS: Different makes and models of scanners predictably demonstrated different quantitative performance profiles including, in some cases, small calibration bias. Differences in site-specific reconstruction parameters increased the quantitative variability among similar scanners, with postreconstruction smoothing filters being the most influential parameter. Quantitative assessment of this intrascanner variability over this large collection of phantom data gives, for the first time, estimates of reconstruction variance introduced into trials from allowing trial sites to use their preferred reconstruction methodologies. Predictably, time-of-flight-enabled scanners exhibited less size-based partial-volume bias than non-time-of-flight scanners.
CONCLUSION: The CTN scanner validation experience over the past 5 y has generated a rich, well-curated phantom dataset from which PET/CT make-and-model and reconstruction-dependent quantitative behaviors were characterized for the purposes of understanding and estimating scanner-based variances in clinical trials. These results should make it possible to identify and recommend make-and-model-specific reconstruction strategies to minimize measurement variability in cancer clinical trials.
© 2015 by the Society of Nuclear Medicine and Molecular Imaging, Inc.

Entities:  

Keywords:  PET quantitation; multicenter clinical trials; phantom; scanner calibration

Mesh:

Year:  2014        PMID: 25525180     DOI: 10.2967/jnumed.114.148056

Source DB:  PubMed          Journal:  J Nucl Med        ISSN: 0161-5505            Impact factor:   10.057


  49 in total

1.  The engagement of FDG PET/CT image quality and harmonized quantification: from competitive to complementary.

Authors:  Ronald Boellaard
Journal:  Eur J Nucl Med Mol Imaging       Date:  2016-01       Impact factor: 9.236

2.  Comparative analysis of iterative reconstruction algorithms with resolution recovery and time of flight modeling for 18F-FDG cardiac PET: A multi-center phantom study.

Authors:  Roberta Matheoud; Michela Lecchi; Domenico Lizio; Camilla Scabbio; Claudio Marcassa; Lucia Leva; Angelo Del Sole; Carlo Rodella; Luca Indovina; Christian Bracco; Marco Brambilla; Orazio Zoccarato
Journal:  J Nucl Cardiol       Date:  2016-01-12       Impact factor: 5.952

3.  mDixon-Based Synthetic CT Generation for PET Attenuation Correction on Abdomen and Pelvis Jointly Using Transfer Fuzzy Clustering and Active Learning-Based Classification.

Authors:  Pengjiang Qian; Yangyang Chen; Jung-Wen Kuo; Yu-Dong Zhang; Yizhang Jiang; Kaifa Zhao; Rose Al Helo; Harry Friel; Atallah Baydoun; Feifei Zhou; Jin Uk Heo; Norbert Avril; Karin Herrmann; Rodney Ellis; Bryan Traughber; Robert S Jones; Shitong Wang; Kuan-Hao Su; Raymond F Muzic
Journal:  IEEE Trans Med Imaging       Date:  2019-08-16       Impact factor: 10.048

4.  Measuring PET Spatial Resolution Using a Cylinder Phantom Positioned at an Oblique Angle.

Authors:  Martin A Lodge; Jeffrey P Leal; Arman Rahmim; John J Sunderland; Eric C Frey
Journal:  J Nucl Med       Date:  2018-06-14       Impact factor: 10.057

5.  18F-FDG PET/CT heterogeneity quantification through textural features in the era of harmonisation programs: a focus on lung cancer.

Authors:  Charline Lasnon; Mohamed Majdoub; Brice Lavigne; Pascal Do; Jeannick Madelaine; Dimitris Visvikis; Mathieu Hatt; Nicolas Aide
Journal:  Eur J Nucl Med Mol Imaging       Date:  2016-06-21       Impact factor: 9.236

Review 6.  Towards enhanced PET quantification in clinical oncology.

Authors:  Habib Zaidi; Nicolas Karakatsanis
Journal:  Br J Radiol       Date:  2017-11-22       Impact factor: 3.039

7.  General Biomarker Recommendations for Lymphoma.

Authors:  Lisa Rimsza; Yuri Fedoriw; Louis M Staudt; Ari Melnick; Randy Gascoyne; Michael Crump; Lawrence Baizer; Kai Fu; Eric Hsi; John W C Chan; Lisa McShane; John P Leonard; Brad S Kahl; Richard F Little; Jonathan W Friedberg; Lale Kostakoglu
Journal:  J Natl Cancer Inst       Date:  2016-12-16       Impact factor: 13.506

8.  Standard OSEM vs. regularized PET image reconstruction: qualitative and quantitative comparison using phantom data and various clinical radiopharmaceuticals.

Authors:  Judit Lantos; Erik S Mittra; Craig S Levin; Andrei Iagaru
Journal:  Am J Nucl Med Mol Imaging       Date:  2018-04-25

9.  Multicenter survey of PET/CT protocol parameters that affect standardized uptake values.

Authors:  Darrin Byrd; Rebecca Christopfel; John Buatti; Eduardo Moros; Sadek Nehmeh; Adam Opanowski; Paul Kinahan
Journal:  J Med Imaging (Bellingham)       Date:  2017-12-08

10.  Whole-body PET estimation from low count statistics using cycle-consistent generative adversarial networks.

Authors:  Yang Lei; Xue Dong; Tonghe Wang; Kristin Higgins; Tian Liu; Walter J Curran; Hui Mao; Jonathon A Nye; Xiaofeng Yang
Journal:  Phys Med Biol       Date:  2019-11-04       Impact factor: 3.609

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.