| Literature DB >> 29467359 |
Fabian Kording1,2, Bjoern P Schoennagel1, Manuela Tavares de Sousa3, Kai Fehrs1,2, Gerhard Adam1, Jin Yamamura1, Christian Ruprecht1,2.
Abstract
PURPOSE: Fetal cardiac MRI has the potential to play an important role in the assessment of fetal cardiac pathologies, but it is up to now not feasible due to a missing gating method. The purpose of this work was the evaluation of Doppler ultrasound (DUS) for external fetal cardiac gating with regard to compatibility, functionality, and reliability. Preliminary results were assessed performing fetal cardiac MRI.Entities:
Keywords: Doppler ultrasound; cardiovascular system; fetal heart; magnetic resonance imaging
Mesh:
Year: 2018 PMID: 29467359 PMCID: PMC6196307 DOI: 10.2463/mrms.mp.2017-0100
Source DB: PubMed Journal: Magn Reson Med Sci ISSN: 1347-3182 Impact factor: 2.471
Fig. 1(A) Used Doppler ultrasound (DUS) system with the ultrasound transducer (1) connected to the transmission line containing the cable traps, which are covered by a surrounding isolating sleeve (2). The ultrasound transducer is finally connected to the DUS device (3). The application of the DUS device to derive a fetal cardiac gating signal is shown schematically in (B). The transducer is placed above the fetal heart and is connected with the DUS electronic via a cable including cable traps to attenuate radiofrequency (RF) interferences. After signal acquisition, the signal processing algorithm is outlined schematically. The gating output is finally connected to the physiologic unit of the MRI.
Fig. 2The setup and dimensions of the used biventricular heart phantom are shown in (a–c). An example of the measured Doppler ultrasound signal from the moving heart phantom mimicking fetal heart movement is shown in (d) with corresponding detected gating signals (BPM = 118).
Results of cable trap attenuation
| Cable trap | 1 | 2 | 3 | 4 |
|---|---|---|---|---|
| 64 MHz (S21 dB) | 48 | 50 | 47 | 45 |
| 128 MHz (S21 dB) | 42.42 | 47.18 | 39.15 | 26.22 |
The values shown in the table are the S21 measurement results in dB for each cable trap at 1.5T (64 MHz) and 128 MHz (3T) using a network analyzer and current probes. Each cable trap fulfilled the minimum requirement of 20 dB.
Fig. 3Shown are coronal B1 flip angle maps and axial gradient echo images at 1.5T (upper row) and at 3T (bottom row) for cases with transducer inside the scanner (dashed line) on top of the phantom in (b) and (f) and without transducer inside the magnetic resonance scanner in (a) and (e). No effects on B1 field homogeneity due to the cable or transducer are visible for 1.5T or for 3T as shown in the image difference in (c) and (g). No geometric distortions or signal loss due to common-mode currents were visible with the transducer inside the MRI using Gradient echo images for 1.5T (d) and for 3T (h).
Fig. 4Shown are exemplary steady-state free precession (SSFP) images of the fetal cardiac phantom at 1.5T (upper row) and at 3T (bottom row). Images during mid-diastole are shown in (a) and in (c) with their corresponding projection over one whole cardiac cycle in (b) and (d).
DUS gating results
| RR intervals | RR length (ms) | Difference in RR length (msec) | Variability (msec) | Sensitivity (%) | ||
|---|---|---|---|---|---|---|
| 1.5T | Heart phantom | 2924 | 408 ± 127 | – | 0.1 ± 0.01 | 100 |
| DUS device | 2924 | 404 ± 126 | 5.5 ± 4.5 | 6.4 ± 2.1 | 100 | |
| 3T | Heart phantom | 2789 | 444 ± 147 | – | 0.1 ± 0.01 | 100 |
| DUS device | 2789 | 443 ± 147 | 5.6 ± 4.6 | 6.5 ± 2.8 | 100 | |
| Fetal verification | Heart cycle | 1876 | 434 ± 32 | – | 22.0 ± 7.6 | 100 |
| DUS device | 1860 | 434 ± 31 | 0.01 ± 11 | 22.2 ± 7.1 | 99.1 |
The values shown in the tables are the Doppler ultrasound (DUS) gating results for the phantom study at 1.5T and 3T and for the in vivo verification study outside the MRI room. RR, beat-to-beat.
Fig. 5The results of the Doppler ultrasound (DUS) gating validation using the heart phantom during MRI acquisition for heart rates ranging from 75 to 200 BPM is shown in (a) and (b), for the in vivo validation outside of MRI in (c) and (d). The dashed lines of the Bland Altman plot in (b) and (d) represent the confidence interval of ±1.96 of the standard deviation.
Fig. 6Exemplary the Doppler ultrasound (DUS) signal during MRI acquisition for one fetus at 1.5T. The dashed lines represent the detected gating time points of DUS device which are sent to the physiologic unit of the MRI scanner for gating.
Fig. 7Preliminary fetal cardiac steady-state free precession (SSFP) cine images in four-chamber view for 1.5T (upper row) and 3T (bottom row). Gated images are shown in diastole in (a) and (d), for systole in (b) and (e). The corresponding projection over one cardiac cycle is shown in (c) and (f).