Hazardous situation in the MR bore: induction in ECG leads causes fire.

Harald Kugel1, Christoph Bremer, Marco Püschel, Roman Fischbach, Horst Lenzen, Bernd Tombach, Hugo Van Aken, Walter Heindel.   

Abstract

There is a potential hazard during examinations of patients with attached or implanted long conductors, e.g. ECG leads: an MR exam of the lumbar spine of a patient was performed in a 1.5-T scanner under ECG monitoring using equipment marked as MR compatible. Standard cabling of 370-cm length was guided without loops from the electrodes through the caudal opening of the magnet bore. During a sagittal T1-weighted turbo-spin-echo scan suddenly a flame of approximately 3 cm arose from the patient's shirt, close to the position of the electrodes. The supervising anaesthesiologist extinguished the flames with his hands. A subsequent physical examination revealed second- to third-degree burns. The analysis of the incident revealed that high voltages can be induced in straight conductors without loops as ECG cables by coupling with the electric component of the HF field. Local heating or sparking can cause an open flame at the position of the electrodes. This danger exists even with ECG equipment that is specifically marked as MR compatible.

Entities:  

Mesh:

Year:  2003        PMID: 12664104     DOI: 10.1007/s00330-003-1841-8

Source DB:  PubMed          Journal:  Eur Radiol        ISSN: 0938-7994            Impact factor:   5.315


  10 in total

1.  Neurostimulation systems for deep brain stimulation: in vitro evaluation of magnetic resonance imaging-related heating at 1.5 tesla.

Authors:  Ali R Rezai; Daniel Finelli; John A Nyenhuis; Greg Hrdlicka; Jean Tkach; Ashwini Sharan; Paul Rugieri; Paul H Stypulkowski; Frank G Shellock
Journal:  J Magn Reson Imaging       Date:  2002-03       Impact factor: 4.813

2.  On the heating of linear conductive structures as guide wires and catheters in interventional MRI.

Authors:  W R Nitz; A Oppelt; W Renz; C Manke; M Lenhart; J Link
Journal:  J Magn Reson Imaging       Date:  2001-01       Impact factor: 4.813

Review 3.  Safety considerations in MR imaging.

Authors:  E Kanal; F G Shellock; L Talagala
Journal:  Radiology       Date:  1990-09       Impact factor: 11.105

4.  Unusual burns of the lower extremities caused by a closed conducting loop in a patient at MR imaging.

Authors:  M V Knopp; M Essig; J Debus; H J Zabel; G van Kaick
Journal:  Radiology       Date:  1996-08       Impact factor: 11.105

5.  Effects of magnetic resonance imaging on cardiac pacemakers and electrodes.

Authors:  S Achenbach; W Moshage; B Diem; T Bieberle; V Schibgilla; K Bachmann
Journal:  Am Heart J       Date:  1997-09       Impact factor: 4.749

6.  Transesophageal cardiac pacing during magnetic resonance imaging: feasibility and safety considerations.

Authors:  M B Hofman; C C de Cock; J C van der Linden; A C van Rossum; F C Visser; M Sprenger; N Westerhof
Journal:  Magn Reson Med       Date:  1996-03       Impact factor: 4.668

7.  Severe burn of the finger caused by using a pulse oximeter during MR imaging.

Authors:  F G Shellock; G L Slimp
Journal:  AJR Am J Roentgenol       Date:  1989-11       Impact factor: 3.959

8.  Burns associated with clinical MR examinations.

Authors:  E Kanal; F G Shellock
Journal:  Radiology       Date:  1990-05       Impact factor: 11.105

9.  The electrical characteristics of some commercial ECG electrodes.

Authors:  R P Patterson
Journal:  J Electrocardiol       Date:  1978-01       Impact factor: 1.438

Review 10.  Patient monitoring during clinical MR imaging.

Authors:  E Kanal; F G Shellock
Journal:  Radiology       Date:  1992-12       Impact factor: 11.105

  10 in total
  21 in total

1.  Self-gated cardiac cine MRI.

Authors:  Andrew C Larson; Richard D White; Gerhard Laub; Elliot R McVeigh; Debiao Li; Orlando P Simonetti
Journal:  Magn Reson Med       Date:  2004-01       Impact factor: 4.668

2.  MRI quality control: six imagers studied using eleven unified image quality parameters.

Authors:  T Ihalainen; O Sipilä; S Savolainen
Journal:  Eur Radiol       Date:  2004-03-03       Impact factor: 5.315

3.  Adaptive removal of gradients-induced artefacts on ECG in MRI: a performance analysis of RLS filtering.

Authors:  Mario Sansone; Luciano Mirarchi; Marcello Bracale
Journal:  Med Biol Eng Comput       Date:  2010-03-18       Impact factor: 2.602

4.  Assessing the Electromagnetic Fields Generated By a Radiofrequency MRI Body Coil at 64 MHz: Defeaturing Versus Accuracy.

Authors:  Elena Lucano; Micaela Liberti; Gonzalo G Mendoza; Tom Lloyd; Maria Ida Iacono; Francesca Apollonio; Steve Wedan; Wolfgang Kainz; Leonardo M Angelone
Journal:  IEEE Trans Biomed Eng       Date:  2015-12-17       Impact factor: 4.538

5.  Flow-compensated self-gating.

Authors:  Jessica Schulz; Matthias Korn; Michael Deimling; Wolfhard Semmler; Michael Bock
Journal:  MAGMA       Date:  2008-07-31       Impact factor: 2.310

6.  Integrating fMRI with psychophysiological measurements in the study of decision-making.

Authors:  Savio W H Wong; Gui Xue; Antoine Bechara
Journal:  J Neurosci Psychol Econ       Date:  2011

7.  Comparison of left ventricular function assessment using phonocardiogram- and electrocardiogram-triggered 2D SSFP CINE MR imaging at 1.5 T and 3.0 T.

Authors:  Meike Becker; Tobias Frauenrath; Fabian Hezel; Gabriele A Krombach; Ute Kremer; Benedikt Koppers; Christoph Butenweg; Andreas Goemmel; Jane F Utting; Jeanette Schulz-Menger; Thoralf Niendorf
Journal:  Eur Radiol       Date:  2009-12-16       Impact factor: 5.315

8.  RF Heating of Gold Cup and Conductive Plastic Electrodes during Simultaneous EEG and MRI.

Authors:  Mukund Balasubramanian; William M Wells; John R Ives; Patrick Britz; Robert V Mulkern; Darren B Orbach
Journal:  Neurodiagn J       Date:  2017

9.  Acoustic cardiac triggering: a practical solution for synchronization and gating of cardiovascular magnetic resonance at 7 Tesla.

Authors:  Tobias Frauenrath; Fabian Hezel; Wolfgang Renz; Thibaut de Geyer d'Orth; Matthias Dieringer; Florian von Knobelsdorff-Brenkenhoff; Marcel Prothmann; Jeanette Schulz Menger; Thoralf Niendorf
Journal:  J Cardiovasc Magn Reson       Date:  2010-11-16       Impact factor: 5.364

Review 10.  Physiological recordings: basic concepts and implementation during functional magnetic resonance imaging.

Authors:  Marcus A Gray; Ludovico Minati; Neil A Harrison; Peter J Gianaros; Vitaly Napadow; Hugo D Critchley
Journal:  Neuroimage       Date:  2009-05-19       Impact factor: 6.556

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.