Daisuke Kawahara1,2, Shuichi Ozawa3,4, Akito Saito3, Tomoki Kimura3, Tatsuhiko Suzuki3, Masato Tsuneda5, Sodai Tanaka6, Takeo Nakashima1, Yoshimi Ohno1, Yuji Murakami3, Yasushi Nagata3,4.
Abstract
PURPOSE: Lipiodol, which was used in transcatheter arterial chemoembolization before liver stereotactic body radiation therapy (SBRT), remains in SBRT. Previous we reported the dose enhancement in Lipiodol using 10 MV (10×) FFF beam. In this study, we compared the dose enhancement in Lipiodol and evaluated the probability of electron generation (PEG) for the dose enhancement using flattening filter (FF) and flattening filter free (FFF) beams.Entities:
Keywords: zzm321990PEGzzm321990; Lipiodol; Monte Carlo calculation; energy spectrum
Mesh:
Substances:
Year: 2018 PMID: 29450985 PMCID: PMC5849857 DOI: 10.1002/acm2.12282
Source DB: PubMed Journal: J Appl Clin Med Phys ISSN: 1526-9914 Impact factor: 2.102
Figure 1A geometric scheme of Lipiodol was located at depth of 5.0 cm in a water‐equivalent phantom (20 × 20 × 20 cm3).
Figure 2The validation of PDD curves comparing the film measurement with MC calculation with FFF beam (left) and FF beam (right) for 6× and 10×.
Figure 3MC‐calculated PDD curves for the virtual phantoms with (asterisk) and without (closed circles) Lipiodol with (a) 6× FFF, (b) 6× FF, (c) 10× FFF and (d) 10× FF beams.
Figure 4The dose enhancement ratio with the virtual water phantom which Lipiodol was located at depth of 5.0 cm for 6× FFF (open diamond), 6× FF (asterisk), 10× FFF (open circle), and 10× FF (closed circle) beams.
Figure 5The disrubutions of photon energy spectrum and energy at the center of Lipiodol (depth, 6.5 cm), for irradiation with FFF and FF beams with 6× and 10×. (a, b) Photon spectra for irradiation with 6× and 10× beams, respectively. (c, d) Electron spectra for irradiation with 6× and 10× beams, respectively.
Figure 6PEG at the center of the Lipiodol (6.5 cm depth) with FFF and FF for 6× and 10× beams. (a, b) PEG by photoelectric effect for 6× and 10×, respectively. (c, d) PEG by Compton scattering for 6× and 10×, respectively.
Figure 7Total PEG at the center of the Lipiodol (6.5 cm depth) with FFF and FF for 6× and 10× beams.