Literature DB >> 28102954

Dosimetric impact of Lipiodol in stereotactic body radiation therapy on liver after trans-arterial chemoembolization.

Daisuke Kawahara1,2, Shuichi Ozawa3, Akito Saito3, Teiji Nishio3, Tomoki Kimura3, Tatsuhiko Suzuki3, Kazunari Hioki1, Takeo Nakashima1, Yoshimi Ohno1, Yuji Murakami3, Yasushi Nagata3.   

Abstract

PURPOSE: Stereotactic body radiation therapy (SBRT) combining trans-arterial chemoembolization (TACE) with Lipiodol is expected to improve local control. This study is aimed to estimate the dose enhancement in Lipiodol's proximity and to evaluate the dose calculation accuracy of the Acuros XB (AXB) algorithm and anisotropic analytical algorithm (AAA) in the Eclipse treatment planning system (TPS) (ver. 11, Varian Medical Systems, Palo Alto, USA), compared with that of the Monte Carlo (MC) calculation (using BEAMnrc/DOSXYZnrc code) for a virtual phantom and a treatment plan for liver SBRT after TACE.
METHODS: The MC calculation accuracy was validated by comparing its results with the percent depth dose (PDD) and the off-axis ratio (OAR) measured using a water-equivalent phantom containing Lipiodol. The dose difference in Lipiodol's proximity and the inhomogeneity correction accuracies of the AAA, AXB algorithm, and MC calculation were evaluated by calculating the PDDs and OARs for the virtual phantom with Lipiodol and the lateral profile for the clinical plan data.
RESULTS: The measured data and the MC results agreed within 3%. The average dose in the Lipiodol uptake region was higher by 8.1% for the virtual phantom and 6.0% for the clinical case compared to that in regions without Lipiodol uptake. For the virtual phantom, compared with the MC calculation, the AAA and the AXB algorithm underestimated the doses immediately upstream of the Lipiodol region by 5.0% and 4.2%, in the Lipiodol region by 7.4% and 9.8%, and downstream of the Lipiodol region by 5.5% and 3.9% respectively. These discrepancy between the AXB and MC calculations were due to the incorrect assignment of Lipiodol material properties. Namely, the bone material was assigned automatically by the AXB algorithm as the materials for the AXB algorithm do not contain iodine, which is the main constituent of Lipiodol.
CONCLUSIONS: The MC calculation indicated a larger and more accurate dose increase in Lipiodol compared with the TPS algorithms. The observed dose enhancement in the tumor area could be clinically significant.
© 2016 American Association of Physicists in Medicine.

Entities:  

Keywords:  Lipiodol; Monte Carlo; dose enhancement

Mesh:

Substances:

Year:  2017        PMID: 28102954     DOI: 10.1002/mp.12028

Source DB:  PubMed          Journal:  Med Phys        ISSN: 0094-2405            Impact factor:   4.071


  6 in total

1.  Evaluation of raw-data-based and calculated electron density for contrast media with a dual-energy CT technique.

Authors:  Daisuke Kawahara; Shuichi Ozawa; Kazushi Yokomachi; Toru Higaki; Takehiro Shiinoki; Yoshimi Ohno; Yuji Murakami; Kazuo Awai; Yasushi Nagata
Journal:  Rep Pract Oncol Radiother       Date:  2019-08-20

2.  Energy spectrum and dose enhancement due to the depth of the Lipiodol position using flattened and unflattened beams.

Authors:  Daisuke Kawahara; Shuichi Ozawa; Akito Saito; Tomoki Kimura; Tatsuhiko Suzuki; Masato Tsuneda; Sodai Tanaka; Kazunari Hioki; Takeo Nakashima; Yoshimi Ohno; Yuji Murakami; Yasushi Nagata
Journal:  Rep Pract Oncol Radiother       Date:  2018-01-12

3.  DNA strand breaks based on Monte Carlo simulation in and around the Lipiodol with flattening filter and flattening filter-free photon beams.

Authors:  Daisuke Kawahara; Akito Saito; Hisashi Nakano; Yasushi Nagata
Journal:  Rep Pract Oncol Radiother       Date:  2022-07-29

4.  Biological dose-enhancement analysis with Monte Carlo simulation for Lipiodol for photon beams.

Authors:  Daisuke Kawahara; Shuichi Ozawa; Hisashi Nakano; Katsumaro Kubo; Takehiro Shiinoki; Tomoki Kimura; Yasushi Nagata
Journal:  Rep Pract Oncol Radiother       Date:  2019-11-08

5.  Effect of secondary electron generation on dose enhancement in Lipiodol with and without a flattening filter.

Authors:  Daisuke Kawahara; Shuichi Ozawa; Akito Saito; Tomoki Kimura; Tatsuhiko Suzuki; Masato Tsuneda; Sodai Tanaka; Takeo Nakashima; Yoshimi Ohno; Yuji Murakami; Yasushi Nagata
Journal:  J Appl Clin Med Phys       Date:  2018-02-15       Impact factor: 2.102

6.  Comparison of Stereotactic Body Radiation Therapy Combined With or Without Transcatheter Arterial Chemoembolization for Patients With Small Hepatocellular Carcinoma Ineligible for Resection or Ablation Therapies.

Authors:  Tomoki Kimura; Hiroshi Aikata; Yoshiko Doi; Nobuki Imano; Yuki Takeuchi; Ippei Takahashi; Ikuno Nishibuchi; Tsuyoshi Katsuta; Masahiro Kenjo; Yuji Murakami; Kazuo Awai; Kazuaki Chayama; Yasushi Nagata
Journal:  Technol Cancer Res Treat       Date:  2018-01-01
  6 in total