Literature DB >> 29436841

Fat suppression in magnetic resonance imaging of the head and neck region: is the two-point DIXON technique superior to spectral fat suppression?

Christina M Wendl1, Johannes Eiglsperger2, Lena-Marie Dendl1, Harald Brodoefel1,3, Karl-Michael Schebesch4, Christian Stroszczynski1, Claudia Fellner1.   

Abstract

OBJECTIVE: The aim of our study was to systematically compare two-point Dixon fat suppression (FS) and spectral FS techniques in contrast enhanced imaging of the head and neck region.
METHODS: Three independent readers analysed coronal T1 weighted images recorded after contrast medium injection with Dixon and spectral FS techniques with regard to FS homogeneity, motion artefacts, lesion contrast, image sharpness and overall image quality.
RESULTS: 85 patients were prospectively enrolled in the study. Images generated with Dixon-FS technique were of higher overall image quality and had a more homogenous FS over the whole field of view compared with the standard spectral fat-suppressed images (p < 0.001). Concerning motion artefacts, flow artefacts, lesion contrast and image sharpness no statistically significant difference was observed.
CONCLUSION: The Dixon-FS technique is superior to the spectral technique due to improved homogeneity of FS and overall image quality while maintaining lesion contrast. Advances in knowledge: T1 with Dixon FS technique offers, compared to spectral FS, significantly improved FS homogeneity and over all image quality in imaging of the head and neck region.

Entities:  

Mesh:

Substances:

Year:  2018        PMID: 29436841      PMCID: PMC6190774          DOI: 10.1259/bjr.20170078

Source DB:  PubMed          Journal:  Br J Radiol        ISSN: 0007-1285            Impact factor:   3.039


  20 in total

1.  T1-weighted fat-suppressed imaging of the pelvis with a dual-echo Dixon technique: initial clinical experience.

Authors:  Peter Beddy; R Deepa Rangarajan; Masako Kataoka; Penelope Moyle; Martin J Graves; Evis Sala
Journal:  Radiology       Date:  2010-11-15       Impact factor: 11.105

Review 2.  Head and neck imaging: the role of CT and MRI.

Authors:  Franz J Wippold
Journal:  J Magn Reson Imaging       Date:  2007-03       Impact factor: 4.813

3.  Chemical shift artifact: dependence on shape and orientation of the lipid-water interface.

Authors:  R C Smith; R C Lange; S M McCarthy
Journal:  Radiology       Date:  1991-10       Impact factor: 11.105

Review 4.  Dixon techniques for water and fat imaging.

Authors:  Jingfei Ma
Journal:  J Magn Reson Imaging       Date:  2008-09       Impact factor: 4.813

5.  Fat-suppression failure artifacts simulating pathology on frequency-selective fat-suppression MR images of the head and neck.

Authors:  Y Anzai; R B Lufkin; B A Jabour; W N Hanafee
Journal:  AJNR Am J Neuroradiol       Date:  1992 May-Jun       Impact factor: 3.825

6.  MR Neurography of Brachial Plexus at 3.0 T with Robust Fat and Blood Suppression.

Authors:  Xinzeng Wang; Crystal Harrison; Yogesh K Mariappan; Karthik Gopalakrishnan; Avneesh Chhabra; Robert E Lenkinski; Ananth J Madhuranthakam
Journal:  Radiology       Date:  2016-12-22       Impact factor: 11.105

Review 7.  Chemical shift encoding-based water-fat separation methods.

Authors:  Holger Eggers; Peter Börnert
Journal:  J Magn Reson Imaging       Date:  2014-01-21       Impact factor: 4.813

8.  Simple proton spectroscopic imaging.

Authors:  W T Dixon
Journal:  Radiology       Date:  1984-10       Impact factor: 11.105

Review 9.  Fat suppression in MR imaging: techniques and pitfalls.

Authors:  E M Delfaut; J Beltran; G Johnson; J Rousseau; X Marchandise; A Cotten
Journal:  Radiographics       Date:  1999 Mar-Apr       Impact factor: 5.333

10.  Comparing T1-weighted and T2-weighted three-point Dixon technique with conventional T1-weighted fat-saturation and short-tau inversion recovery (STIR) techniques for the study of the lumbar spine in a short-bore MRI machine.

Authors:  S Brandão; D Seixas; M Ayres-Basto; S Castro; J Neto; C Martins; J C Ferreira; F Parada
Journal:  Clin Radiol       Date:  2013-08-06       Impact factor: 2.350

View more
  5 in total

Review 1.  Target Definition in MR-Guided Adaptive Radiotherapy for Head and Neck Cancer.

Authors:  Mischa de Ridder; Cornelis P J Raaijmakers; Frank A Pameijer; Remco de Bree; Floris C J Reinders; Patricia A H Doornaert; Chris H J Terhaard; Marielle E P Philippens
Journal:  Cancers (Basel)       Date:  2022-06-20       Impact factor: 6.575

2.  Implementation of a sagittal T2-weighted DIXON turbo spin-echo sequence may shorten MRI acquisitions in the emergency setting of suspected spinal bleeding.

Authors:  Nico Sollmann; Charlotte Rüther; Simon Schön; Claus Zimmer; Thomas Baum; Jan S Kirschke
Journal:  Eur Radiol Exp       Date:  2021-05-12

3.  Is fat suppression in T1 and T2 FSE with mDixon superior to the frequency selection-based SPAIR technique in musculoskeletal tumor imaging?

Authors:  Willemijn H F Huijgen; Catherina S P van Rijswijk; Johan L Bloem
Journal:  Skeletal Radiol       Date:  2019-06-01       Impact factor: 2.199

4.  Occult Disco-Ligamentous Lesions of the Subaxial c-Spine-A Comparison of Preoperative Imaging Findings and Intraoperative Site Inspection.

Authors:  Insa Janssen; Nico Sollmann; Melanie Barz; Thomas Baum; Karl Schaller; Claus Zimmer; Yu-Mi Ryang; Jan S Kirschke; Bernhard Meyer
Journal:  Diagnostics (Basel)       Date:  2021-03-05

5.  Regularized joint water-fat separation with B0 map estimation in image space for 2D-navigated interleaved EPI based diffusion MRI.

Authors:  Yiming Dong; Kirsten Koolstra; Malte Riedel; Matthias J P van Osch; Peter Börnert
Journal:  Magn Reson Med       Date:  2021-07-13       Impact factor: 4.668

  5 in total