Literature DB >> 29415203

Left atrial appendage patency and device-related thrombus after percutaneous left atrial appendage occlusion: a computed tomography study.

Hubert Cochet1,2, Xavier Iriart3, Soumaya Sridi1, Claudia Camaioni1, Olivier Corneloup1, Michel Montaudon1,2, François Laurent1,2, Wieme Selmi3, Pauline Renou3, Zakaria Jalal2,3, Jean-Benoît Thambo2,3.   

Abstract

Aims: Transoesophageal echocardiography studies have reported frequent peri-device leaks and device-related thrombi (DRT) after percutaneous left atrial appendage (LAA) occlusion. We assessed the prevalence, characteristics and correlates of leaks and DRT on cardiac computed tomography (CT) after LAA occlusion. Methods and results: Consecutive patients underwent cardiac CT before LAA occlusion to assess left atrial (LA) volume, LAA shape, and landing zone diameter. Follow-up CT was performed after >3 months to assess device implantation criteria, device leaks and DRT. CT findings were related to patient and device characteristics, as well as to outcome during follow-up. One-hundred and seventeen patients (age 74 ± 9, 37% women, CHA2DS2VASc 4.4 ± 1.3, and HASBLED 3.5 ± 1.0) were implanted with Amplatzer cardiac plug (ACP)/Amulet (71%) or Watchman (29%). LAA patency was detected in 44% on arterial phase CT images and 69% on venous phase images. The most common leak location was postero-inferior. LAA patency related to LA dilatation, left ventricular ejection fraction impairment, non-chicken wing LAA shape, large landing zone diameter, incomplete device lobe thrombosis, and disc/lobe misalignment in patients with ACP/Amulet. DRT were detected in 19 (16%), most being laminated and of antero-superior location. DRT did not relate to clinical or imaging characteristics nor to implantation criteria, but to total thrombosis of device lobe. Over a mean 13 months follow-up, stroke/transient ischaemic attack occurred in eight patients, unrelated to DRT or LAA patency.
Conclusion: LAA patency on CT is common after LAA occlusion, particularly on venous phase images. Leaks relate to LA/LAA size at baseline, and device malposition and incomplete thrombosis at follow-up. DRT is also quite common but poorly predicted by patient and device-related factors.

Entities:  

Mesh:

Year:  2018        PMID: 29415203     DOI: 10.1093/ehjci/jey010

Source DB:  PubMed          Journal:  Eur Heart J Cardiovasc Imaging        ISSN: 2047-2404            Impact factor:   6.875


  9 in total

1.  Assessment of the healing process after percutaneous implantation of a cardiovascular device: a systematic review.

Authors:  Elodie Perdreau; Zakaria Jalal; Richard D Walton; Jérôme Naulin; Julie Magat; Bruno Quesson; Hubert Cochet; Olivier Bernus; Jean-Benoît Thambo
Journal:  Int J Cardiovasc Imaging       Date:  2019-11-19       Impact factor: 2.357

2.  Short-Term Antiplatelet Versus Anticoagulant Therapy After Left Atrial Appendage Occlusion: A Systematic Review and Meta-Analysis.

Authors:  Mohammed Osman; Tatiana Busu; Khansa Osman; Safi U Khan; Matthew Daniels; David R Holmes; Mohamad Alkhouli
Journal:  JACC Clin Electrophysiol       Date:  2020-01-29

3.  Clinical and echocardiographic risk factors for device-related thrombus after left atrial appendage closure: an analysis from the multicenter EUROC-DRT registry.

Authors:  Vivian Vij; Kerstin Piayda; Dominik Nelles; Steffen Gloekler; Roberto Galea; Monika Fürholz; Bernhard Meier; Marco Valgimigli; Gilles O'Hara; Dabit Arzamendi; Victor Agudelo; Lluis Asmarats; Xavier Freixa; Eduardo Flores-Umanzor; Ole De Backer; Lars Sondergaard; Luis Nombela-Franco; Angela McInerney; Kasper Korsholm; Jens Erik Nielsen-Kudsk; Shazia Afzal; Tobias Zeus; Felix Operhalski; Boris Schmidt; Gilles Montalescot; Paul Guedeney; Xavier Iriart; Noelie Miton; Jacqueline Saw; Thomas Gilhofer; Laurent Fauchier; Egzon Veliqi; Felix Meincke; Nils Petri; Peter Nordbeck; Dmitrii Ognerubov; Evgeny Merkulov; Ignacio Cruz-González; Rocio Gonzalez-Ferreiro; Deepak L Bhatt; Alessandra Laricchia; Antonio Mangieri; Heyder Omran; Jan Wilko Schrickel; Josep Rodes-Cabau; Horst Sievert; Georg Nickenig; Alexander Sedaghat
Journal:  Clin Res Cardiol       Date:  2022-07-18       Impact factor: 6.138

Review 4.  Sex Differences in Atrial Fibrillation-Update on Risk Assessment, Treatment, and Long-Term Risk.

Authors:  Charlotte J Bai; Nidhi Madan; Shaza Alshahrani; Neelum T Aggarwal; Annabelle Santos Volgman
Journal:  Curr Treat Options Cardiovasc Med       Date:  2018-08-27

5.  Design and Rationale of the Swiss-Apero Randomized Clinical Trial: Comparison of Amplatzer Amulet vs Watchman Device in Patients Undergoing Left Atrial Appendage Closure.

Authors:  Roberto Galea; Federico De Marco; Adel Aminian; Nicolas Meneveau; Frederic Anselme; Christoph Gräni; Adrian T Huber; Emmanuel Teiger; Xavier Iriart; Marco Angelillis; Nicolas Brugger; Alessandro Spirito; Noé Corpataux; Anna Franzone; Pascal Vranckx; Urs Fischer; Giovanni Pedrazzini; Francesco Bedogni; Stephan Windecker; Lorenz Räber; Marco Valgimigli
Journal:  J Cardiovasc Transl Res       Date:  2021-04-21       Impact factor: 4.132

6.  Short-term non-vitamin K antagonist oral anticoagulants vs. warfarin in preventing device-related thrombosis after left atrial appendage closure.

Authors:  Yuyi Chen; Yonghua Zhang; Lulu Qu; Weiping Huang; Xi Su; Yanhong Chen
Journal:  J Thromb Thrombolysis       Date:  2021-03-05       Impact factor: 2.300

7.  Reduced plasma level of basic fibroblast growth factor is associated with incomplete device endothelialization at six months following left atrial appendage closure.

Authors:  Jing Xu; Xin Gong; Chuanzhi Chen; Jun Xing; Qi Wang; Weifeng Shen; Qi Zhang
Journal:  BMC Cardiovasc Disord       Date:  2021-05-16       Impact factor: 2.298

8.  Device neo-endothelialization after left atrial appendage closure: the role of cardiac computed tomography angiography.

Authors:  Roberto Galea; Christoph Gräni
Journal:  Int J Cardiovasc Imaging       Date:  2021-03-17       Impact factor: 2.357

9.  Temporal changes and clinical significance of peridevice leak following left atrial appendage occlusion with Amplatzer devices.

Authors:  Kasper Korsholm; Jesper M Jensen; Bjarne L Nørgaard; Jens E Nielsen-Kudsk
Journal:  Catheter Cardiovasc Interv       Date:  2022-05-18       Impact factor: 2.585

  9 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.