Polyethylene Wear Increases in Liners Articulating With Scratched Oxidized Zirconium Femoral Heads.

Alberto Carli1, Chelsea N Koch, Christina I Esposito, Timothy M Wright, Douglas E Padgett.   

Abstract

BACKGROUND: Oxidized zirconium (OxZr) femoral heads were introduced in hip arthroplasty to reduce wear of the polyethylene compared with metallic heads and to reduce fracture risk compared with ceramic heads. Severe scratches have been reported on OxZr heads in patients undergoing revision for instability, but whether these scratches contribute to increased acetabular polyethylene wear remains unclear. QUESTIONS/PURPOSES: (1) How is the polyethylene of the acetabular liner affected by damage on the opposing OxZr head? (2) How does damage to the head affect the degree of polyethylene wear?
METHODS: We assessed damage and deformation on all retrieved highly crosslinked liners that had articulated against OxZr heads collected at one institution between 2006 and 2013. Two observers used a visual subjective scoring system to assess polyethylene damage on the surface of the 42 retrieved liners. Polyethylene components were also laser scanned to measure dimensional changes to the liner. These outcomes were compared with the severity of scratching on the surface of the articulating OxZr head. We also used a 12-station hip simulator to measure wear over five million cycles (MCs) of pristine liners articulating against nine retrieved OxZr heads with varying degrees of scratching representing a spectrum of little to severe damage and three retrieved ceramic heads with severe metal transfer.
RESULTS: Seventeen of the OxZr heads showed severe damage, of which 14 heads had been revised for dislocation. The retrieved liners that had articulated with these heads had greater damage scores for abrasion (mean score 0.4 versus 1.6; p = 0.008) and embedded debris (mean score 0.4 versus 1.4; p = 0.006) compared with liners that had articulated with less damaged heads. Four severely damaged OxZr heads wore at a higher rate than the others in the study with weight loss of 37.7 mg, 30.0 mg, 14.4 mg, and 2.6 mg after the first MC and a steady increase through testing to 5 MCs. Conversely, neither OxZr heads with less damage nor ceramic heads with severe metal transfer produced appreciable wear.
CONCLUSIONS: Surface scratching of OxZr heads from recurrent dislocation and reduction maneuvers leads to increased wear of the crosslinked polyethylene used as a bearing surface. CLINICAL RELEVANCE: Patients in whom such instability occurs in the presence of an OxZr head should be followed closely for the possibility of more severe wear.

Entities:  

Mesh:

Substances:

Year:  2018        PMID: 29389782      PMCID: PMC6259683          DOI: 10.1007/s11999.0000000000000040

Source DB:  PubMed          Journal:  Clin Orthop Relat Res        ISSN: 0009-921X            Impact factor:   4.176


  22 in total

1.  Effect of molecular weight, calcium stearate, and sterilization methods on the wear of ultra high molecular weight polyethylene acetabular cups in a hip joint simulator.

Authors:  H A McKellop; F W Shen; P Campbell; T Ota
Journal:  J Orthop Res       Date:  1999-05       Impact factor: 3.494

2.  Development of an extremely wear-resistant ultra high molecular weight polyethylene for total hip replacements.

Authors:  H McKellop; F W Shen; B Lu; P Campbell; R Salovey
Journal:  J Orthop Res       Date:  1999-03       Impact factor: 3.494

3.  Damage of oxinium femoral heads subsequent to hip arthroplasty dislocation three retrieval case studies.

Authors:  Alan M Kop; Colin Whitewood; Don J L Johnston
Journal:  J Arthroplasty       Date:  2007-04-20       Impact factor: 4.757

4.  Laser scanning as a useful tool in implant retrieval analysis: a demonstration using rotating platform and fixed bearing tibial inserts.

Authors:  Kirsten E Stoner; Nader A Nassif; Timothy M Wright; Douglas E Padgett
Journal:  J Arthroplasty       Date:  2013-08-13       Impact factor: 4.757

5.  Oxidized zirconium head on crosslinked polyethylene liner in total hip arthroplasty: a 7- to 12-year in vivo comparative wear study.

Authors:  George K Karidakis; Theofilos Karachalios
Journal:  Clin Orthop Relat Res       Date:  2015-12       Impact factor: 4.176

6.  Retrieval analysis of total knee prostheses: a method and its application to 48 total condylar prostheses.

Authors:  R W Hood; T M Wright; A H Burstein
Journal:  J Biomed Mater Res       Date:  1983-09

7.  Surface Damage Is Not Reduced With Highly Crosslinked Polyethylene Tibial Inserts at Short-term.

Authors:  Tong Liu; Christina Esposito; Marcella Elpers; Timothy Wright
Journal:  Clin Orthop Relat Res       Date:  2016-01       Impact factor: 4.176

8.  Tribocorrosion: Ceramic and Oxidized Zirconium vs Cobalt-Chromium Heads in Total Hip Arthroplasty.

Authors:  Sok Chuen Tan; Adrian C K Lau; Christopher Del Balso; James L Howard; Brent A Lanting; Matthew G Teeter
Journal:  J Arthroplasty       Date:  2016-02-27       Impact factor: 4.757

Review 9.  Fracture of ceramic bearing surfaces following total hip replacement: a systematic review.

Authors:  Francesco Traina; Marcello De Fine; Alberto Di Martino; Cesare Faldini
Journal:  Biomed Res Int       Date:  2013-06-13       Impact factor: 3.411

10.  Expected values for pedometer-determined physical activity in older populations.

Authors:  Catrine Tudor-Locke; Teresa L Hart; Tracy L Washington
Journal:  Int J Behav Nutr Phys Act       Date:  2009-08-25       Impact factor: 6.457

View more
  5 in total

1.  Editor's Spotlight/Take 5: Polyethylene Wear Increases in Liners Articulating With Scratched Oxidized Zirconium Femoral Heads.

Authors:  Seth S Leopold
Journal:  Clin Orthop Relat Res       Date:  2018-02       Impact factor: 4.176

2.  MRI of THA Correlates With Implant Wear and Tissue Reactions: A Cross-sectional Study.

Authors:  Matthew F Koff; Christina Esposito; Parina Shah; Mauro Miranda; Elexis Baral; Kara Fields; Thomas Bauer; Douglas E Padgett; Timothy Wright; Hollis G Potter
Journal:  Clin Orthop Relat Res       Date:  2019-01       Impact factor: 4.176

3.  The Oxinium Arthrogram: A Sign of Oxidized Zirconium Implant Failure.

Authors:  Benjamin M Frye; Kelsey R Laughery; Adam E Klein
Journal:  Arthroplast Today       Date:  2021-03-07

4.  Long-term performance of oxidized zirconium on conventional and highly cross-linked polyethylene in total hip arthroplasty.

Authors:  Alberto V Carli; Anay R Patel; Michael B Cross; David J Mayman; Kaitlin M Carroll; Paul M Pellicci; Seth A Jerabek
Journal:  SICOT J       Date:  2020-05-07

5.  Protrusion of a ceramic femoral head through the acetabular metallic cup in total-hip arthroplasty: A case report.

Authors:  Ding Zhao; De-Bao Zhang; Dong-Feng Han; Gui-Shan Gu
Journal:  Medicine (Baltimore)       Date:  2020-06-12       Impact factor: 1.817

  5 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.