Literature DB >> 10376720

Effect of molecular weight, calcium stearate, and sterilization methods on the wear of ultra high molecular weight polyethylene acetabular cups in a hip joint simulator.

H A McKellop1, F W Shen, P Campbell, T Ota.   

Abstract

Orthopaedic surgeons must currently choose from several types of ultra high molecular weight polyethylene acetabular cups that differ in their material properties and in the methods used for their sterilization. Information on the wear resistance of these different cups may help in the selection process. This study included two separate tests for wear run on a hip simulator to investigate the effect of molecular weight, calcium stearate, and sterilization methods on the wear resistance of ultra high molecular weight polyethylene acetabular cups. Test 1 revealed nearly identical wear rates for acetabular cups with molecular weights in two distinct ranges, as well as for cups with molecular weights in the same range but with or without calcium stearate added. In Test 2, cups that were sterilized in air with gamma irradiation exhibited lower rates of wear than those sterilized with ethylene oxide, presumably due to the crosslinking induced by the irradiation. In addition, cups that were irradiated while packed in a partial vacuum to minimize oxygen absorbed in the surface layer initially showed lower rates of wear than those irradiated in air, with the wear rates becoming similar as wear penetrated the more oxidized surface layer and the more crosslinked subsurface region. Because these tests were run a few months after the irradiation, the potential effects of long-term oxidation of any residual free radicals in the irradiated materials could not be taken into account. After artificial aging to accelerate oxidative degradation of the materials, the wear rates could be markedly different. Analyses performed after wear indicated that the irradiated (i.e., crosslinked) cups exhibited a smaller proportion of, as well as shorter, fibrils in the wear debris and an increased crystallinity and melting temperature and that gamma irradiation in the low-oxygen environment reduced the level of oxidation and increased the level of crosslinking in the surface region of the cups.

Entities:  

Mesh:

Substances:

Year:  1999        PMID: 10376720     DOI: 10.1002/jor.1100170306

Source DB:  PubMed          Journal:  J Orthop Res        ISSN: 0736-0266            Impact factor:   3.494


  7 in total

1.  Oxidation and wear of 100-Mrad cross-linked polyethylene shelf-aged for 30 years.

Authors:  Hironobu Oonishi; Sok Chol Kim; Hiroyuki Oonishi; Masayuki Kyomoto; Shingo Masuda
Journal:  Clin Orthop Relat Res       Date:  2008-01-10       Impact factor: 4.176

2.  Metal-on-conventional polyethylene total hip arthroplasty bearing surfaces have a higher risk of revision than metal-on-highly crosslinked polyethylene: results from a US registry.

Authors:  Elizabeth W Paxton; Maria C S Inacio; Robert S Namba; Rebecca Love; Steven M Kurtz
Journal:  Clin Orthop Relat Res       Date:  2015-03       Impact factor: 4.176

3.  Do first-generation highly crosslinked polyethylenes oxidize in vivo?

Authors:  Daniel MacDonald; Ashlyn Sakona; Allyson Ianuzzi; Clare M Rimnac; Steven M Kurtz
Journal:  Clin Orthop Relat Res       Date:  2011-08       Impact factor: 4.176

Review 4.  History and systematic review of wear and osteolysis outcomes for first-generation highly crosslinked polyethylene.

Authors:  Steven M Kurtz; Heather A Gawel; Jasmine D Patel
Journal:  Clin Orthop Relat Res       Date:  2011-08       Impact factor: 4.176

5.  Polyethylene Wear Increases in Liners Articulating With Scratched Oxidized Zirconium Femoral Heads.

Authors:  Alberto Carli; Chelsea N Koch; Christina I Esposito; Timothy M Wright; Douglas E Padgett
Journal:  Clin Orthop Relat Res       Date:  2018-02       Impact factor: 4.176

Review 6.  Ultra-High Molecular Weight Polyethylene: Influence of the Chemical, Physical and Mechanical Properties on the Wear Behavior. A Review.

Authors:  Pierangiola Bracco; Anuj Bellare; Alessandro Bistolfi; Saverio Affatato
Journal:  Materials (Basel)       Date:  2017-07-13       Impact factor: 3.623

7.  The influence of malalignment and ageing following sterilisation by gamma irradiation in an inert atmosphere on the wear of ultra-high-molecular-weight polyethylene in patellofemoral replacements.

Authors:  Raman Maiti; Raelene M Cowie; John Fisher; Louise M Jennings
Journal:  Proc Inst Mech Eng H       Date:  2017-03-25       Impact factor: 1.617