Tribocorrosion: Ceramic and Oxidized Zirconium vs Cobalt-Chromium Heads in Total Hip Arthroplasty.

Sok Chuen Tan1, Adrian C K Lau1, Christopher Del Balso1, James L Howard1, Brent A Lanting1, Matthew G Teeter2.   

Abstract

BACKGROUND: This matched-cohort study aims to compare tribocorrosion between matched ceramic and cobalt-chromium femoral head trunnions and between matched Oxinium and cobalt-chromium femoral head trunnions. Secondary objectives were to investigate whether taper design, depth of trunnion, implantation time, age, body mass index, and gender have an effect on fretting and corrosion.
METHODS: All hip prostheses retrieved between 1999 and 2015 at one center were reviewed, giving a total of 52 ceramic heads. These were matched to a cobalt-chromium cohort according to taper design, head size, neck length, and implantation time. The trunnions were examined by 2 observers using a 4-point scoring technique and scored in 3 zones: apex, middle, and base. The observers were blinded to clinical and manufacturing data where possible. A separate matched-cohort analysis was performed between 8 Oxinium heads and 8 cobalt-chromium heads, which were similarly scored.
RESULTS: Ceramic head trunnions demonstrated a lower median fretting and corrosion score at the base zone (P < .001), middle zone (P < .001), and in the combined score (P < .001). Taper design had a significant effect on fretting and corrosion in the apex zone (P = .04) of the ceramic group, as well as the cobalt-chromium group (P = .03). Between Oxinium heads and cobalt-chromium heads, there was no significant difference in the fretting and corrosion score across all 3 zones (base: P = .22; middle: P = .92; and apex: P = .71) and for the combined score (P = .67).
CONCLUSION: This study shows that ceramic head confers an advantage in trunnion fretting and corrosion. Taper design and implantation time were also significant factors for fretting and corrosion.
Copyright © 2016 Elsevier Inc. All rights reserved.

Entities:  

Keywords:  Oxinium heads; ceramic heads; cobalt-chromium heads; oxidized zirconium heads; taper corrosion; trunnionosis

Mesh:

Substances:

Year:  2016        PMID: 27067752     DOI: 10.1016/j.arth.2016.02.027

Source DB:  PubMed          Journal:  J Arthroplasty        ISSN: 0883-5403            Impact factor:   4.757


  14 in total

1.  Wear performance of cobalt chromium, ceramic, and oxidized zirconium on highly crosslinked polyethylene at mid-term follow-up.

Authors:  Matthew G Teeter; Colin J MacLean; Lyndsay E Somerville; James L Howard; Richard W McCalden; Brent A Lanting; Edward M Vasarhelyi
Journal:  J Orthop       Date:  2018-05-07

Review 2.  [Tribology in hip arthroplasty : Benefits of different materials].

Authors:  J Philippe Kretzer; Maximilian Uhler; Sebastian Jäger; Therese Bormann; Robert Sonntag; Mareike Schonhoff; Stefan Schröder
Journal:  Orthopade       Date:  2021-02-25       Impact factor: 1.087

3.  Do Well-functioning THAs Retrieved at Autopsy Exhibit Evidence of Fretting and Corrosion?

Authors:  Jeffrey Lange; Amanda Wach; Chelsea N Koch; Robert H Hopper; Henry Ho; Charles A Engh; Timothy M Wright; Douglas E Padgett
Journal:  Clin Orthop Relat Res       Date:  2018-10       Impact factor: 4.176

4.  Periprosthetic Joint Infection in Hip Arthroplasty: Is There an Association Between Infection and Bearing Surface Type?

Authors:  Rocco P Pitto; Laurent Sedel
Journal:  Clin Orthop Relat Res       Date:  2016-10       Impact factor: 4.176

5.  CORR Insights®: Polyethylene Wear Increases in Liners Articulating With Scratched Oxidized Zirconium Femoral Heads.

Authors:  Matthew G Teeter
Journal:  Clin Orthop Relat Res       Date:  2018-02       Impact factor: 4.176

Review 6.  Corrosion of the Head-Stem Taper Junction-Are We on the Verge of an Epidemic?: Review Article.

Authors:  Michael Morlock; Dennis Bünte; Julian Gührs; Nicholas Bishop
Journal:  HSS J       Date:  2016-09-20

Review 7.  Current status and future potential of wear-resistant coatings and articulating surfaces for hip and knee implants.

Authors:  Charlotte Skjöldebrand; Joanne L Tipper; Peter Hatto; Michael Bryant; Richard M Hall; Cecilia Persson
Journal:  Mater Today Bio       Date:  2022-04-30

8.  Polyethylene Wear Increases in Liners Articulating With Scratched Oxidized Zirconium Femoral Heads.

Authors:  Alberto Carli; Chelsea N Koch; Christina I Esposito; Timothy M Wright; Douglas E Padgett
Journal:  Clin Orthop Relat Res       Date:  2018-02       Impact factor: 4.176

Review 9.  Current Concepts of Using Large Femoral Heads in Total Hip Arthroplasty.

Authors:  Myung-Rae Cho; Won Kee Choi; Jae Jung Kim
Journal:  Hip Pelvis       Date:  2016-09-30

10.  Pseudotumor in ceramic-on-metal total hip arthroplasty.

Authors:  Yoni M Blau; Andrew J Meyers; Mauro Giordani; John P Meehan
Journal:  Arthroplast Today       Date:  2017-10-06