| Literature DB >> 29368796 |
Christopher P Oliver1, Duncan J Butler1, Viliami Takau1, Ivan Williams1.
Abstract
The Australian Radiation Protection and Nuclear Safety Agency (ARPANSA) held a comparison exercise in April 2016 where participants came to ARPANSA and measured the output factor of a nominal 5 mm cone attached to the ARPANSA Elekta Synergy (Elekta, Crawley, UK) linear accelerator. The goal of the exercise was to compare the consistency and methods used by independent medical physicists in measuring small-field output factors. ARPANSA provided a three-dimensional scanning tank for detector setup and positioning, but the participants were required to measure the output factor with their own detectors. No information regarding output factors previously measured was supplied to participants to make each result as independent as possible. Fifteen groups travelled to ARPANSA bringing a wide range of detectors and methods. A total of 30 measurements of the output factor were made. The standard deviation of the measurements (excluding one expected outlier from an uncorrected ionization chamber measurement) was 3.6%. The results provide an insight into the consistency of small-field dosimetry being performed in Australia and New Zealand at the present time.Entities:
Keywords: output factor; small-field dosimetry; stereotactic cone
Mesh:
Year: 2018 PMID: 29368796 PMCID: PMC5849830 DOI: 10.1002/acm2.12259
Source DB: PubMed Journal: J Appl Clin Med Phys ISSN: 1526-9914 Impact factor: 2.102
Figure 1Experimental setup for the small‐field output factor measurement indicating the geometry for the reference and small fields. In this example, the 10 × 10 cm field was being measured by a Farmer chamber and the small field by a diode; however, the detector types were chosen by each participant. The measurement by both detectors in an intermediate field to relate the two displayed measurements is not shown.
Participant organizations
| Organization |
|---|
| Andrew Love Cancer Centre, Geelong, Vic |
| ARPANSA, Vic |
| Calvary Mater Newcastle, Waratah, NSW |
| Casey Radiation Oncology Centre, Berwick Vic |
| Chris O'Brien Lifehouse, Camperdown, NSW |
| Olivia Newton‐John Cancer Wellness & Research Centre, Austin Health, Heidelberg, Vic |
| Peter MacCallum Cancer Centre, Sunshine, Vic |
| Prince of Wales Hospital, Randwick, NSW |
| Radiation Oncology Institute, Wahroonga, NSW |
| RMIT University, Melbourne, Vic |
| Royal Adelaide Hospital, Adelaide, SA |
| The Alfred Hospital, Melbourne, Vic |
| The Canberra Hospital, Garran, ACT |
| Townsville Cancer Centre, Douglas, QLD |
| Vector Lab, University of Sydney, NSW |
| Waikato Hospital, Hamilton, New Zealand |
Materials and dimensions of the active volumes of the detectors used in the comparison exercise
| Detector | Active detector material | Active detector dimensions perpendicular to the beam | Active detector length parallel to the beam | Active volume |
|---|---|---|---|---|
| IBA Razor detector | Unshielded p‐type silicon diode | 0.6 mm diameter | 0.02 mm | 0.006 mm3 |
| Sun nuclear edge detector | Silicon diode encapsulated in brass housing | 0.8 × 0.8 mm2 | 0.03 mm | 0.019 mm3 |
| IBA SFD stereotactic diode | Unshielded p‐type silicon diode | 0.6 mm diameter | 0.06 mm | 0.02 mm3 |
| IBA EFD electron diode | Unshielded p‐type silicon diode | 2 mm diameter | 0.06 mm | 0.2 mm3 |
| PTW 60019 microDiamond | Diamond Schottky diode | 2.2 mm diameter | 0.001 mm | 0.004 mm3 |
| PTW 60018 SRS diode | Unshielded p‐type silicon diode | 1.2 mm diameter | 0.25 mm | 0.3 mm3 |
| Aircore plastic scintillation detector | BC400 plastic scintillator | 1 mm diameter | 1 mm | 0.8 cm3 |
| PTW 60017 electron diode | Unshielded p‐type silicon diode | 1.2 mm diameter | 0.03 mm | 0.03 mm3 |
| Bragg peak chamber | PMMA‐walled ionization chamber vented to air | 82 mm diameter | 2 mm | 10.5 cm3 |
| PTW 31016 PinPoint 3D | PMMA‐walled ionization chamber vented to air | 2.9 mm diameter | 2.9 mm | 0.016 cm3 |
| PTW 31014 PinPoint | PMMA‐walled ionization chamber vented to air | 2 mm diameter | 5 mm | 0.015 cm3 |
| IBA CC04 | C552‐walled ionization chamber vented to air | 4 mm diameter | 3.6 mm | 0.04 cm3 |
| Standard imaging A1SL | C552‐walled ionization chamber vented to air | 4 mm diameter | 4.4 mm | 0.053 cm3 |
Summary of results for all active detectors
| 5 mm field detector | Intermediate field | 10 × 10 cm field detector | Uncorrected output factor | Stated uncertainty | Correction factor | Source of correction factor | Output factor | Stated uncertainty | Clinical use | Smallest clinical field |
|---|---|---|---|---|---|---|---|---|---|---|
| IBA Razor diode | NA | IBA Razor diode | 0.637 | 0.61% | None | NA | 0.637 | 0.61% | No | NA |
| IBA Razor diode | 5 × 5 cm | IBA FC65‐G | 0.653 | 3.02% | 0.974 |
| 0.636 | 3.62% | Yes | 1.5 cm2 |
| Sun nuclear edge | NA | Sun Nuclear Edge | 0.668 | 3.74% | 0.945 |
| 0.632 | 4.17% | Yes | 2 × 2 cm |
| PTW 31016 pinpoint 3D | NA | PTW 31016 Pinpoint 3D | 0.529 | 3.74% | 1.12 |
| 0.592 | 4.13% | Yes | 2 × 2 cm |
| IBA SFD stereotactic | 2 × 2 cm & 5 × 5 cm | IBA CC04 | 0.628 | 2.78% | None | NA | 0.628 | 2.78% | Yes | 4 mm cone |
| IBA EFD electron | 4 × 4 cm | IBA CC13 | 0.638 | 1.60% | 0.955 |
| 0.610 | 1.68% | Yes | 10 mm cone |
| IBA SFD stereotactic | 5 × 5 cm | PTW pinpoint 3D chamber | 0.639 | 1.46% | 0.961 |
| 0.614 | 2.48% | Yes | 1 × 1 cm |
| PTW 60019 microDiamond | NA | PTW microDiamond | 0.643 | 5.0% | 0.969 |
| 0.623 | 5.00% | No | NA |
| IBA RAZOR diode | 4 × 4 cm | IBA CC13 | 0.652 | 0.10% | None | NA | 0.652 | 0.10% | Yes | 5 mm cone |
| Sun nuclear edge | NA | Sun Nuclear Edge | 0.667 | 1.00% | None | NA | 0.667 | 1.00% | Yes | 1.5 × 1.5 cm |
| PTW 60019 microDiamond | NA | PTW 60019 microDiamond | 0.645 | 1.00% | 0.961 |
| 0.620 | 1.00% | Yes | 1.5 × 1.5 cm |
| Standard imaging A1SL | NA | Standard Imaging A1SL | 0.471 | 1.00% | None | NA | 0.471 | 1.00% | Yes | 2 × 2 cm |
| PTW T60018 SRS Diode | 3 × 3 cm & 5 × 5 cm | IBA CC04 | 0.658 | 1.50% | 0.929 |
| 0.612 | 3.00% | Yes | 1 × 1 cm |
| IBA CC04 | NA | IBA CC04 | 0.500 | 0.2% | 1.141 | – | 0.570 | 3.40% | Yes | 1 × 1 cm |
| IBA EFD electron | 3 × 3 cm | IBA CC13 | 0.623 | 1.00% | 0.974 |
| 0.606 | 2.90% | No | NA |
| IBA SFD stereotactic | 3 × 3 cm | IBA CC13 | 0.618 | 1.00% | 0.966 |
| 0.598 | 3.00% | No | NA |
| PTW 34070 Bragg Peak chamber + Gafchromic EBT3 film | 5 cm cone field | IBA FC65‐G | NA | NA | NA | NA | 0.645 | 4.90% | No | NA |
| IBA Razor diode | 3 × 3 cm | IBA CC04 | 0.641 | 0.16% | 1.002 | – | 0.642 | 0.16% | No | NA |
| Air core plastic scintillator | NA | Air core plastic scintillator | 0.599 | 0.67% | 1.007 | – | 0.603 | 0.66% | Yes | 4 mm cone |
| PTW 60017 electron diode | NA | PTW 60017 electron diode | 0.632 | 3.00% | 0.939 |
| 0.594 | 3.20% | No | NA |
| PTW 60019 microDiamond | NA | PTW 60019 microDiamond | 0.601 | 3.00% | 1.027 |
| 0.618 | 3.20% | No | NA |
| PTW 31014 pinpoint chamber | NA | PTW 31014 pinpoint chamber | 0.490 | 3.20% | 1.22 |
| 0.598 | 3.5% | No | NA |
Expanded uncertainty (k = 2) defining an interval having a level of confidence of approximately 95% in the output factor (or uncorrected output factor), as stated by the facility. The reasons for the large variations between these values are discussed in the text.
Volume averaging correction calculated from profile measurements.
Summary of all results using Gafchromic EBT3 film
| Output factor | Stated uncertainty | Linearity curve measured at ARPANSA | MU delivered for 5 mm field irradiations | Number of films irradiated in 5 mm field | MU delivered for 10 × 10 cm field irradiations | Number of films irradiated in 10 × 10 cm field | Size of region of interest for 5 mm cone field | Scanner Brand/Model |
|---|---|---|---|---|---|---|---|---|
| 0.601 | 3.35% | No | 500 | 3 | 500 | 3 | 0.35 mm2 | EPSON (Seiko Epson Corp., Nagano, Japan) 10000XL |
| 0.639 | 6.00% | No | 500 | 1 | 500 | 1 | 1 × 0.5 mm | EPSON V700 red channel |
| 0.611 | 0.80% | Yes | 200 | 3 | 200 | 3 | 1.4 × 1.4 mm | EPSON Expression 10000XL |
| 0.610 | 3.40% | Yes | 434 | 1 | Multiple exposures, 200–300 MU | 4 | 4 pixels | EPSON V700 |
| 0.570 | 4.56% | No | 154 | 5 | 92 | 5 | 0.7 × 0.7 mm | EPSON 10000XL |
| 0.615 | 3.25% | No | 285 | 3 | 200 | 3 | Zero area extrapolation technique | EPSON 10000XL |
| 0.621 | 4.50% | Yes | 400 | 1 | 100 MU FC‐65G used to get CAX dose | NA, IBA FC‐65G ionization chamber used | 1 mm diameter | EPSON 10000XL red channel |
| 0.599 | 3.67% | Yes | 300 | 5 | 300 | 5 | 0.7 × 0.7 mm | EPSON 10000XL |
Expanded uncertainty (k = 2) in the output factor (or uncorrected output factor), as stated by the facility.
Figure 2All reported output factors grouped according to detector type with their associated average value shown as a horizontal line.