| Literature DB >> 29333245 |
Fredrik Leijonhufvud1, Fredrik Jöneby1, Jan G Jakobsson2.
Abstract
Low-flow anaesthesia is considered beneficial for the patient and the environment, and it is cost reducing due to reduced anaesthetic gas consumption. An initial high-flow to saturate the circle system ( wash-in) is desirable from a clinical point of view. We measured the wash-in and wash-out times (time to saturate and to eliminate the anaesthetic agent, AA), for sevoflurane and desflurane, in a test-lung with fixed 3 MAC vaporizer setting at different fresh gas flow (FGF) and calculated the consumption of AA. We tried to find an optimal flow rate for speed and gas consumption, comparing two anaesthesia machines (AMs): Aisys and Flow-i. Time to reach 1 minimal alveolar concentration (MAC) (wash-in) decreased (p<0.05) at higher flow rates (1 - 2 - 4) but plateaued at 4-4.8 l/min. The consumption of AA was at its lowest around 4-4.8 l/min (optimal flow) for all but the Aisys /desflurane group. Wash-out times decreased as FGF increased, until reaching plateau at FGF of 4-6 l/min. Aisys had generally shorter wash-in times at flow rates < 4 l/min as well as lower consumption of AA. At higher flow rates there were little difference between the AMs. The "optimal FGF" for wash-out, elimination of gas from the test-lung and circle system, plateaued with no increase in speed beyond 6 l/min. A fresh gas flow of 4 l/min. seems "optimal" taking speed to reach a 1 MAC ET and gas consumption into account during wash-in with a fixed 3 MAC vaporizer setting, and increasing fresh gas flow beyond 6 l/min does not seem to confirm major benefit during wash-out.Entities:
Keywords: desflurane; fresh gas flow; inhalational anaesthetics; low-flow anaesthesia; sevoflurane; wash-in; wash-out
Year: 2017 PMID: 29333245 PMCID: PMC5754744 DOI: 10.12688/f1000research.13064.2
Source DB: PubMed Journal: F1000Res ISSN: 2046-1402
Figure 1. Schematic view of the experimental set-up.
The fresh-gas flow is defined in this project as the gas-mixture introduced to the circle i.e. after the vaporizer. The functional residual capacity (FRC) consisted of two reservoir bags. The tidal volume (TV) consisted of one self-deflating bag. A small volume of gas gets diverted through the sidestream sensor but is reintroduced into the circle afterwards. 1. Vaporizer; 2. One-way valves; 3. Sidestream sensor; 4. Mainstream sensor; 5. Spill-valve; 6. CO 2 Absorber.
Figure 2. Time to reach 1 MAC and the calculated AA consumption for each fresh gas flows.
( a) Flow-i /sevoflurane. ( b) Flow-i /desflurane. ( c) Aisys /sevoflurane. ( d) Aisys/desflurane. Time(s) is on the Y-axis. FGF (l/min) is on the X-axis. AA (ml) is on the secondary Y-axis.
Times at which wash-out plateaued, and fastest wash-out times at optimal flow rate.
| Flow-i | Aisys | All (mean) | |
|---|---|---|---|
|
| 349 s | 317 s | 333 s |
|
| 1013 s | 797 s | 905 s |
|
| 681 s | 557 s |
Figure 3. Wash-out times comparing Flow-i and Aisys AMs.
( a) sevoflurane ( b) desflurane. Time (s) is on the Y-axis and FGF (l/min) is on the X-axis.