| Literature DB >> 35435423 |
Martin Bellgardt1, Vladimir Vinnikov1, Adrian Iustin Georgevici1, Livia Procopiuc2, Thomas Peter Weber1, Andreas Meiser3, Jennifer Herzog-Niescery1, Dominik Drees1.
Abstract
The use of volatile anesthetics as sedatives in the intensive care unit is relevant to the patient's outcome. We compared anesthetic gas consumption of the conventional semi-closed Aisys CSTM with the MIRUSTM system, which is the first anesthetic gas reflector system that can administer desflurane in addition to isoflurane and sevoflurane. We connected an artificial lung model to either a MIRUSTM system and a Puritan BennettTM 840 ventilator or an Aisys CSTM anesthesia machine. We found that consumption of 0.5% isoflurane, which corresponds to the target concentration 0.5 MAC, was averaged to 2 mL/h in the MIRUSTM system, which is identical to the Aisys CSTM at a fresh gas flow (FGF) of 1.0 L/min. MIRUSTM consumption of 1% sevoflurane was averaged to 10 mL/h, which corresponds to 8.4 mL/h at FGF 2.5 L/min. The MIRUSTM system consumed 3% or 4% desflurane at an average of 13.0 mL/h or 21.3 mL/h, which is between the consumption at 1.0 L/min and 2.5 L/min FGF. Thus, the MIRUSTM system can effectively deliver volatile anesthetics in clinically relevant concentrations in a similar rate as a conventional circular breathing system at FGFs between 1.0 L/min and 2.5 L/min.Entities:
Keywords: MIRUS; anesthesia; circle system; consumption; desflurane; isoflurane; reflection; sevoflurane; volatile anesthetics; wash-out
Mesh:
Substances:
Year: 2022 PMID: 35435423 PMCID: PMC9074978 DOI: 10.4103/2045-9912.337991
Source DB: PubMed Journal: Med Gas Res ISSN: 2045-9912
Wash-in time of isoflurane, sevoflurane, and desflurane in the Aisys CSTM fresh gas flow and the MIRUSTM system
| System | Isoflurane (0% to 2.5%) | Sevoflurane (0% to 2.5%) | Desflurane (0% to 6.0%) |
|---|---|---|---|
| 0.5 L/min FGF | 7:38±0:16 | 10:41±0:40 | 12:31±0:45 |
| 1.0 L/min FGF | 7:34±0:32 | 8:45±0:27 | 11:39±0:16 |
| 2.5 L/min FGF | 7:00±0:05 | 8:06±0:18 | 10:53±0:42 |
| 5.0 L/min FGF | 7:02±0:08 | 7:15±0:44 | 10:01±0:08 |
| MIRUSTM MV 5L | 12:27±1:09 | 20:04±1:20 | 17:39±1:20 |
Note: Wash-in times were presented in the format hour:minute:second. Data of three independent experiments are presented as mean ± SD, and were analyzed by Kruskal-Wallis test. FGF: Fresh gas flow of the Aisys CSTM; MV: minute volume.
Wash-in of isoflurane (0% to 2.5%), sevoflurane (0% to 2.5%) and desflurane (0% to 6%) of the Aisys CS™ fresh gas flow compared to the MIRUS™ system.
| System | Mean difference (h:min) | 95% bias-corrected and accelerated-confidence interval | Statistics |
|---|---|---|---|
| Isoflurane | |||
| MIRUS™ MV 5L | |||
| 0.5 L/min | -4:49±0:35 | -6:10; -3:49 | |
| 1.0 L/min | -4:53±0:38 | -6:16; -3:46 | |
| 2.5 L/min | -5:26±0:35 | -6:45; -4:29 | |
| 5.0 L/min | -5:24±0:35 | -6:43; -4:26 | |
| Sevoflurane | |||
| MIRUS™ MV 5L | |||
| 0.5 L/min | -9:22±0:40 | -10:43; -7:54 | |
| 1.0 L/min | -11:18±0:38 | -12:30; -10:00 | |
| 2.5 L/min | -11:58±0:38 | -13:22; -10:38 | |
| 5.0 L/min | -12:49±0:41 | -14:07; -11:24 | |
| Desflurane | |||
| MIRUS™ MV 5L | |||
| 0.5 L/min | -5:08±0:41 | -6:33; -3:47 | |
| FGF | |||
| 1.0 L/min | -6:00±0:34 | -7:07; -4:42 | |
| 2.5 L/min | -6:46±0:40 | -08:06; -5:22 | |
| 5.0 L/min | -7:38±0:34 | -8:45; -6:25 | |
| FGF |
Bootstrapped data are presented as mean ± SEM and 95% bias-corrected and accelerated-confidence interval. Dunnet t-test was performed as post hoc analyses with Mirus™ MV 5L as reference category. According to univariate analysis of variance with 1000 sample bootstrapping, the MIRUS™ system shows significantly higher wash-in times. FGF: Fresh gas flow of the Aisys CS™; MV: minute volume; Rc2: corrected R2.
Consumption (mL/h) of isoflurane, sevoflurane, and desflurane related to the Aisys CSTM fresh gas flow, and the MIRUSTM system
| Concentration | 0.5 L/min FGF | 1.0 L/min FGF | 2.5 L/min FGF | 5.0 L/min FGF | MIRUSTM MV 5 L | ||
|---|---|---|---|---|---|---|---|
| Isoflurane | |||||||
| 0.50% | 0.93±0.15 | 2.00±0.10 | 4.00±0.10 | 7.77±0.25 | 2.00±0.00 | <0.001 | >0.999 |
| 1.00% | 1.97±0.06 | 3.40±0.10 | 8.13±0.21 | 15.80±0.20 | 4.00±0.00# | <0.001 | – |
| 1.50% | 2.90±0.10 | 5.50±0.27 | 12.10±0.10 | 23.80±0.27 | 8.00±0.00# | <0.001 | – |
| 2.00% | 3.70±0.10 | 7.27±0.35 | 15.77±0.15 | 30.77±0.25 | 16.00±0.00 | <0.001 | 0.897 |
| 2.50% | 4.97±0.15 | 8.67±0.32 | 19.57±0.47 | 38.43±0.40 | 24.67±2.31† | <0.001 | – |
| Sevoflurane | |||||||
| 0.50% | 1.00±0.00 | 2.00±0.00 | 4.34±0.25 | 8.33±0.58 | 4.67±2.68 | <0.001 | 0.999 |
| 1.00% | 2.33±0.58 | 4.00±0.00 | 8.37±0.29 | 16.00±0.00 | 10.00±2.00 | <0.001 | 0.457 |
| 1.50% | 3.00±0.00 | 5.33±0.58 | 12.47±0.31 | 24.33±0.58 | 14.67±1.15† | <0.001 | – |
| 2.00% | 4.33±0.58 | 8.00±0.00 | 16.43±0.45 | 32.67±0.58 | 22.67±3.06† | <0.001 | – |
| 2.50% | 5.33±0.58 | 9.33±0.58 | 20.37±0.31 | 42.00±0.00 | 30.67±1.15† | <0.001 | – |
| Desflurane | |||||||
| 1.00% | 1.57±0.12 | 3.20±0.26 | 8.03±0.21 | 14.83±0.15 | 4.13±2.72 | <0.001 | 0.998 |
| 2.00% | 3.40±0.53 | 6.33±0.58 | 15.07±0.55 | 29.07±0.40 | 6.88±1.03 | <0.001 | 0.982 |
| 3.00% | 5.00±0.00 | 9.67±0.58 | 22.80±1.05 | 43.00±1.00 | 13.00±1.15# | <0.001 | – |
| 4.00% | 6.17±0.29 | 11.33±0.58 | 30.27±0.72 | 57.00±1.00 | 21.25±1.89# | <0.001 | – |
| 5.00% | 8.50±1.32 | 15.50±0.50 | 37.43±0.60 | 71.83±1.26 | 31.00±5.77 | <0.001 | 0.192 |
| 6.00% | 10.83±1.89 | 19.04±1.07 | 44.77±0.35 | 86.50±1.32 | 40.70±5.31 | <0.001 | 0.598 |
Note: Data of three independent experiments are presented as mean ± SD. Statistically insignificant (i.e., “comparable”) consumption is indicated in bold. *Sidak correction was performed for post-hoc analyses as a follow up to univariate analysis of variance (main effect “device” performed for each concentration) with MIRUSTM tested against 0.5, 1, 2.5, and 5 L/min FGF of the Aisys CSTM). #P < 0.05, vs. 1.0 L/min FGF; †P < 0.05, vs. 2.5 L/min and 5.0 L/min FGF. FGF: Fresh gas flow of the Aisys CSTM.
Consumption time of isoflurane, sevoflurane and desflurane in the Aisys CS™ fresh gas flow, and the MIRUS™ system.
| System | Consumption (h) | 95% bias-corrected andaccelerated-confidence interval | Statistics |
|---|---|---|---|
|
| |||
|
| |||
| 0.5 L/min | 0.93±0.82 | 0.80-1.10 | |
| 1.0 L/min | 2.00±0.53 | 1.90-2.10 |
|
| 2.5 L/min | 4.00±0.05 | 3.93-4.08 |
|
| 5.0 L/min | 7.77±0.14 | 7.50-8.00 |
|
| MIRUS™ MV 5L | 2.00±0.00 | 2.00-2.00 | |
|
| |||
| 0.5L/min | 1.97±0.03 | 1.90-2.00 | |
| 1.0 L/min | 3.40±0.05 | 3.30-3.50 |
|
| 2.5 L/min | 8.13±0.11 | 7.90-8.30 |
|
| 5.0 L/min | 15.80±0.11 | 15.60-16.00 |
|
| MIRUS™ MV 5L | 4.00±0.00 | 4.00-4.00 | |
|
| |||
| 0.5 L/min | 2.90±0.05 | 2.80-3.00 | |
| 1.0 L/min | 5.50±0.14 | 5.30-5.80 |
|
| 2.5 L/min | 12.10±0.55 | 12.00-12.20 |
|
| 5.0 L/min | 23.80±0.14 | 23.50-24.00 |
|
| MIRUS™ MV 5L | 8.00±0.00 | 8.00-8.00 | |
|
| |||
| 0.5 L/min | 3.70±0.54 | 3.60-3.80 | |
| 1.0 L/min | 7.27±0.19 | 6.90-7.60 |
|
| 2.5 L/min | 15.77±0.08 | 15.60-15.90 |
|
| 5.0 L/min | 30.77±0.13 | 30.50-31.00 |
|
| MIRUS™ MV 5L | 16.00±0.00 | 16.00-16.00 | |
|
| |||
| 0.5 L/min | 4.97±0.08 | 4.80-5.10 | |
| 1.0 L/min | 8.67±0.17 | 8.30-8.90 |
|
| 2.5 L/min | 19.57±0.26 | 19.20-20.10 |
|
| 5.0 L/min | 38.43±0.22 | 38.00-38.80 |
|
| MIRUS™ MV 5L | 24.67±1.23 | 22.00-26.00 | |
|
| |||
|
| |||
| 0.5 L/min | 1.00±0.00 | 1.00-1.00 | |
| 1.0 L/min | 2.00±0.00 | 2.00-2.00 |
|
| 2.5 L/min | 4.33±0.13 | 4.10-4.60 |
|
| 5.0 L/min | 8.33±0.30 | 8.00-9.00 |
|
| MIRUS™ MV 5L | 4.67±0.62 | 4.00-6.00 | |
|
| |||
| 0.5 L/min | 2.33±0.32 | 2.00-3.00 | |
| 1.0 L/min | 4.00±0.00 | 4.00-4.00 |
|
| 2.5 L/min | 8.37±0.16 | 8.20-8.70 |
|
| 5.0 L/min | 16.00±0.00 | 16.00-16.00 |
|
| MIRUS™ MV 5L | 10.00±1.08 | 8.67-11.50 | |
|
| |||
| 0.5 L/min | 3.00±0.00 | 3.00-3.00 | |
| 1.0 L/min | 5.33±0.31 | 5.00-6.00 |
|
| 2.5 L/min | 12.47±0.16 | 12.20-12.80 |
|
| 5.0 L/min | 24.33±0.31 | 24.00-25.00 |
|
| MIRUS™ MV 5L | 14.67±0.62 | 14.00-16.00 | |
|
| |||
| 0.5 L/min | 4.33±0.30 | 4.00-5.00 | F(4,10) = 190.81, p < 0.001, |
| FGF | Rc2 > 0.982 | ||
| 1.0 L/min | 8.00±0.00 | 8.00-8.00 | p = 0.100* |
| 2.5 L/min | 16.43±0.24 | 16.00-16.90 |
|
| 5.0 L/min | 32.67±0.32 | 32.00-33.00 |
|
| MIRUS™ MV 5L | 22.67±1.62 | 20.00-26.00 | |
|
| |||
| 0.5 L/min | 5.33±0.32 | 5.00-6.00 | |
| 1.0 L/min | 9.33±0.32 | 9.00-10.00 |
|
| 2.5 L/min | 20.37±0.17 | 20.10-20.70 |
|
| 5.0L/min | 42.00±0.00 | 42.00-42.00 |
|
| MIRUS™ MV 5L | 30.67±0.62 | 30.00-32.00 | |
|
| |||
|
| |||
| 0.5 L/min | 1.57±0.06 | 1.50-1.63 | |
| 1.0 L/min | 3.20±0.14 | 3.00-3.50 |
|
| 2.5 L/min | 8.03±0.11 | 7.80-8.20 |
|
| 5.0 L/min | 14.83±0.08 | 14.7-15.00 |
|
| MIRUS™ MV 5L | 4.13±1.33 | 2.17-7.00 | |
|
| |||
| 0.5 L/min | 3.4±0.29 | 3.00-4.00 | |
| 1.0 L/min | 6.33±0.30 | 6.00-7.00 |
|
| 2.5 L/min | 15.07±0.29 | 14.50-15.60 |
|
| 5.0 L/min | 29.07±0.21 | 28.70-29.50 |
|
| MIRUS™ MV 5L | 6.88±0.50 | 6.00-8.00 | |
|
| |||
| 0.5 L/min | 5.00±0.00 | 5.00-5.00 | |
| 1.0 L/min | 9.67±0.31 | 9.00-10.00 |
|
| 2.5 L/min | 22.80±0.55 | 21.70-23.80 |
|
| 5.0 L/min | 43.00±0.53 | 42.00-44.00 |
|
| MIRUS™ MV 5L | 13.00±0.55 | 12.00-14.00 | |
|
| |||
| 0.5 L/min | 6.17±0.15 | 6.00-6.50 | |
| 1.0 L/min | 11.33±0.30 | 11.00-12.00 |
|
| 2.5 L/min | 30.27±0.39 | 29.80-31.10 |
|
| 5.0 L/min | 57.00±0.52 | 56.36-57.60 |
|
| MIRUS™ | 21.25±0.96 | 20.00-24.00 | |
| MV 5 L | |||
|
| |||
| 0.5 L/min | 8.50±0.72 | 7.50-10.00 | |
| 1.0 L/min | 15.50±0.27 | 15.08-15.90 |
|
| 2.5 L/min | 37.43±0.32 | 36.80-38.00 |
|
| 5.0 L/min | 71.83±0.67 | 70.50-73.00 |
|
| MIRUS™ MV 5L | 31.00±2.80 | 25.52-38.00 | |
|
| |||
| 0.5 L/min | 10.83±1.00 | 9.50-13.00 | |
| 1.0 L/minFGF | 19.04±0.49 | 18.00-20.27 |
|
| 2.5 L/min | 44.77±0.19 | 44.40-45.10 |
|
| 5.0 L/min | 86.50±0.72 | 85.00-87.50 |
|
| MIRUS™ MV 5L | 40.70±2.38 | 34.41-45.09 |
Bootstrapped data are presented as mean ± SEM and 95% bias-corrected and accelerated-confidence interval. Sidak correction was performed for post hoc analyses as a follow up to univariate analysis of variance (main effect “device” for isoflurane, with adjacent FGFs tested against each other (0.5 vs. 1 L/min FGF; 1 vs. 2.5 L/min FGF, 2.5 vs. 5 L/min FGF). FGF: Fresh gas flow of the Aisys CS™; MV: minute volume; Rc2: corrected R2.
Wash-out time of isoflurane, sevoflurane, and desflurane in the Aisys CSTM fresh gas flow and the MIRUSTM system
| System | Isoflurane (2.5% to 0%) | Sevoflurane (2.5% to 0%) | Desflurane (6.0% to 0%) |
|---|---|---|---|
| 0.5 L/min | 1:36:36±0:01:30 | 1:23:21±0:03:30 | 1:39:29±0:06:31 |
| FGF | |||
| 1.0 L/min | 0:48:43±0:00:41 | 0:50:21±0:06:40 | 0:59:00±0:05:08 |
| FGF | |||
| 2.5 L/min | 0:21:17±0:00:18 | 0:20:22±0:00:34 | 0:23:33±0:00:22 |
| FGF | |||
| 5.0 L/min | 0:10:44±0:00:05 | 0:11:19±0:00:56 | 0:12:34±0:00:46 |
| FGF | |||
| MirusTM | 0:36:18±0:02:47# | 0:55:16±0:00:16# | 0:59:46±0:03:00# |
| MV 5L | |||
| MirusTM | 0:09:38±0:00:05‡ | 0:12:39±0:00:30 | 0:14:15±0:00:26 |
| MV 20 L |
Note: Wash-out times are presented as hour:minute:second. Data of three independent experiments are presented as mean ± SD. Univariate analysis of variance with Sidak-corrected post-hoc tests and 1000 sample bootstrapping were used for statistical comparison. #P < 0.05, vs. 1.0 L/min and 2.5 L/min FGF; $P < 0.05, vs. 0.5 L/min and 1.0 L/min FGF; ‡P < 0.05, vs. 1.0 L/min FGF. FGF: Fresh gas flow of the Aisys CSTM; MV: minute volume.
Wash-out time of isoflurane (2.5% to 0%), sevoflurane (2.5% to 0 %) and desflurane (6 % to 0 %) in the Aisys CS™ fresh gas flow, and the MIRUS™ system.
| System | Wash-out time | 95% bias-corrected and accelerated-confidence interval | Statistics |
|---|---|---|---|
|
| |||
| 0.5 L/min FGF | 1:36:36±0:01:30 | 1:33:40-1:39:27 | |
| 1.0 L/min FGF | 0:48:43±0:00:41 | 0:47:17-0:49:38 |
|
| 2.5 L/min FGF | 0:21:17±0:00:18 | 0:20:38-0:21:43 |
|
| 5.0 L/min FGF | 0:10:44±0:00:05 | 0:10:33-0:10:51 |
|
| MIRUS™ MV 5L | 0:36:18±0:02:47 | 0:33:18-0:42:12 | |
| MIRUS™ MV 20L | 0:09:38±0:00:05 | 0:09:26-0:09:48 | |
|
| |||
| 0.5 L/min FGF | 1:23:21±0:01:53 | 1:20:00-1:27:00 | |
| 1.0 L/min FGF | 0:50:21±0:03:39 | 0:46:01-0:58:02 |
|
| 2.5 L/min FGF | 0:20:22±0:00:17 | 0:20:02-0:21:02 |
|
| 5.0 L/min FGF | 0:11:19±0:00:28 | 0:10:05-0:12:03 |
|
| MIRUS™ MV 5L | 0:55:16±0:00:09 | 0:54:58-0:55:29 | |
| MIRUS™ MV 20L | 0:12:39±0:00:10 | 0:12:18-0:12:58 | |
|
| |||
| 0.5 L/min FGF | 1:39:29±0:03:14 | 1:33:47-1:45:47 | |
| 1.0 L/min FGF | 0:59:00±0:02:20 | 0:54:31-1:04:07 |
|
| 2.5 L/min FGF | 0:23:33±0:00:12 | 0:23:07-0:23:49 |
|
| 5.0 L/min FGF | 0:12:34±0:00:24 | 0:12:01-0:13:27 |
|
| MIRUS™ MV 5L | 0:59:46±0:01:35 | 0:56:18-1:01:42 | |
| MIRUS™ MV20L | 0:14:15±0:00:1 | 0:13:56-0:14:34 |
Wash-out times were presented in the format hour:minute:second. Bootstrapped data are presented as mean±SEM and 95% bias-corrected and accelerated-confidence interval. *Sidak correction was performed for post-hoc analysis as follow-up to univariate analysis of variance (F(5,18) = 959.04, P < 0.001, Rc2 = 0.996). **Sidak correction was performed for post-hoc analysis as follow-up to univariate ANOVA (F(5,18) = 474.30 P < 0.001, R 0.992). ***Sidak correction was performed for post-hoc analysis as follow-up to univariate ANOVA (F(5,17) = 351.02, p < 0.001, Rc2 = 0.990). All Sidak correction was with adjacent FGFs tested against each other (0.5 vs. 1 L/min FGF; 1 vs. 2.5 L/min FGF; 2.5 vs. 5 L/min FGF). FGF: Fresh gas flow of the Aisys CS™; MV: minute volume; Rc2: corrected R2.