| Literature DB >> 28103806 |
Hye Won Shin1, Hae Na Yu2, Go Eun Bae2, Hyub Huh2, Ji Yong Park2, Ji Young Kim2.
Abstract
BACKGROUND: Anesthesia machines have been developed by the application of new technology for rapid and easier control of anesthetic concentration. In this study, we used a test lung to investigate whether the time taken to reach the target sevoflurane concentration varies with the rate of fresh gas flow (FGF) and type of anesthesia machine (AM).Entities:
Keywords: Equipment design; Inhalation anesthetics; Ventilation
Mesh:
Substances:
Year: 2017 PMID: 28103806 PMCID: PMC5248460 DOI: 10.1186/s12871-016-0294-y
Source DB: PubMed Journal: BMC Anesthesiol ISSN: 1471-2253 Impact factor: 2.217
The differences in the compositions and configurations of the anesthesia machines; Primus®, Perseus®, and Zeus® [12, 13, 15]
| Primus® | Perseus® | Zeus® | |
|---|---|---|---|
| Mode of anesthesia | Fresh gas mode | Fresh gas mode | Fresh gas mode (Zeus®-F) |
| Type of ventilator | Piston | Blower-driven | Blower-driven |
| Internal volume of anesthesia machine | 4.7 L | 2.1 L | 2.0 L |
| Position of the fresh gas inlet (the rank for proximity to patient) | Between ventilator and absorber (2nd) | In front of ventilator and absorber (1st) | Next to ventilator and absorber (3rd) |
| Presence of a decoupling system | Yes | No | No |
| Type of vaporizer | Bypass | Bypass | DIVA |
| Operation mode of vaporizer | Out-of-circle | Out-of-circle | Out-of-circle (Zeus®-F) |
| Existence of TCA with a feedback control system | No | No | Yes |
DIVA direct injection of volatile anesthetics, TCA target-controlled anesthesia
The times taken to reach the target sevoflurane concentrations using a test lung (A) at variable rates of fresh gas flow (FGF) and different types of anesthesia machine (Primus®, Perseus®, and Zeus®-fresh gas mode), (B) at variable rates of FGF and different modes in Zeus® (Zeus®-F; Zeus® fresh gas mode, Zeus®-A; Zeus® auto-mode) using a lung model. The data are presented as median (95% confidence interval) in seconds. At different FGF rates, but in the same type of AM (Perseus®, Primus®, and Zeus®-F), the time to target concentration shortened as the FGF rate increased (P < 0.05; for simplicity, no statistical remarks are included in the table). Identical data are shown for Zeus®-F mode in Tables A and B. **P < 0.05 for comparisons with Zeus®-F. TCA; target-controlled anesthesia
| A. | |||
| FGF rate | Primus® | Perseus® | Zeus®-F (fresh gas mode) |
| 0.5 L/min | 1165 (1150–1185)† | 920* (883–960) | 1590*† (1528–1639) |
| 1 L/min | 534 (505–538)† | 445* (428–474) | 705*† (690–729) |
| 3 L/min | 155 (149–164)† | 134* (121–137) | 255*† (247–260) |
| B. | |||
| FGF rate | Zeus ®-F (fresh gas mode) | FGF rate | Zeus®-A (auto-mode) |
| 0.5 L/min | 1590 (1528–1639) | Auto-control of FGF by TCA | 380 ** (374–389) |
| 1 L/min | 705 (690–729) | ||
| 3 L/min | 255 (247–260) | ||
*P < 0.05 for comparisons with Primus® . † P < 0.05 for comparisons with Perseus®
**P < 0.05 for comparisons with Zeus®-F
Fig. 1The Fi/Ft ratio curves at variable rate of fresh gas flow (FGF) and different types of anesthesia machine (Primus®, Perseus®, and Zeus® [Zeus®-F; Zeus® fresh gas mode, Zeus®-A; Zeus® auto-mode]). The Fi is the sevoflurane concentration measured at the Y-piece of breathing circuit, while the Ft is the target sevoflurane concentration.)
The circuit volume of anesthesia machine (AM) and time constant at variable rates of fresh gas flow (FGF) and different types of AM
| Primus® | Perseus® | Zeus®-F (fresh gas mode) | ||
|---|---|---|---|---|
| The circuit volume of AM (the internal volume of AM + the volume of breathing circuit 1.2 L) | 5.9 L | 3.3 L | 3.2 L | |
| Time constant (the circuit volume/FGF rate) | ||||
| FGF rate | 0.5 L/min | 11.8 min | 6.6 min | 6.4 min |
| 1 L/min | 5.9 min | 3.3 min | 3.2 min | |
| 3 L/min | 1.9 min | 1.1 min | 1.0 min | |