Otso Arponen1,2, Mazen Sudah3,4, Anna Sutela3, Mikko Taina3, Amro Masarwah3,4, Timo Liimatainen3, Ritva Vanninen3,4,5.
Abstract
To retrospectively evaluated the influence of administration of the gadolinium based intravenous contrast agent (G-CA) on apparent diffusion coefficient (ADC) values in ADC maps generated using multiple b value combinations. A total of 106 women underwent bilateral 3.0 T breast MRI. As an internal validation, diffusion-weighted imaging (b values of 0, 200, 400, 600, 800 s/mm2) was performed before and after the G-CA (gadoterate meglumine (0.2 ml/kg, 3 ml/s)). Whole lesion and fibroglandular tissue (FGT) covering region-of-interests (ROIs) were drawn on the b = 800 s/mm2 images; ROIs were then propagated to multiple retrospectively generated ADC maps. Twenty-seven patients (mean age 55.8 ± 10.8 years) with 32 mass-like enhancing breast lesions including 25 (78.1 %) histopathologically malignant lesions were enrolled. Lesion ADC values were statistically significantly higher in pre-G-CA than post-G-CA ADC maps (ADC0,200,400,600,800: 1.05 ± 0.35 × 10-3 mm2/s vs. 1.02 ± 0.36 × 10-3 mm2/s (P < 0.05); ADC0,200,400: 1.25 ± 0.42 × 10-3 mm2/s vs. 1.20 ± 0.35 × 10-3 mm2/s (P < 0.05)). ADC values between pre- and post-contrast maps were not statistically different when the maps were generated using other b value combinations. Contrast agent administration did not affect the FGT ADC values. G-CA statistically significantly reduced the ADC values of breast lesions on ADC maps generated using the clinically widely utilized b values.Entities:
Mesh:
Substances:
Year: 2018 PMID: 29311709 PMCID: PMC5758819 DOI: 10.1038/s41598-017-18035-0
Source DB: PubMed Journal: Sci Rep ISSN: 2045-2322 Impact factor: 4.379
Figure 1Pre- and post-contrast ADC measurements in three illustrative lesions. The lesions were first localized using T1-weighted, T2-weighted and dynamic contrast enhanced (DCE) images (DCE images shown in the left column). A whole lesion covering region of interest (ROI) was drawn on the pre-contrast b = 800 s/mm2 image. This ROI was then replicated on pre- and post-contrast ADC maps. Images in the right column show ROIs on the pre- and post-contrast ADC maps that were generated using the b values of 0, 200, 400, 600, and 800 s/mm2. (A) A large ductal carcinoma of 47 mm in diameter in the right breast. (B) A ductolobular carcinoma of 21 mm in diameter in the left breast. (C) An intraductal papilloma (red arrow) of 13 mm in diameter in the right breast. A circumscribed, oval hemorrhagic cyst (blue arrow) was detected lateral to the enhancing mass.
Tissue type and mean ± standard deviation apparent diffusion coefficient (ADC) values by using different b value combinations before and after the gadolinium based contrast agent administration (G-CA).
| b Value Protocol | Tissue type | |||||||
|---|---|---|---|---|---|---|---|---|
| Lesion (N = 32) | Fibroglandular tissue (N = 32) | |||||||
| Pre-G-CA ADC (x 10−3 mm2/s) | Post-G-CA ADC (x 10−3 mm2/s) | Mean change (%) | r/ | Pre-G-CA ADC (x 10−3 mm2/s) | Post-G-CA ADC (x 10−3 mm2/s) | Mean change (%) | r/ | |
| ADC0,200,400,600,800 | 1.05 ± 0.35 | 1.03 ± 0.36 | −1.58 ± 4.55 | 0.984/0.042 | 1.49 ± 0.28 | 1.48 ± 0.28 | −0.42 ± 4.47 | 0.950/ns. |
| ADC400,600,800 | 0.87 ± 0.30 | 0.88 ± 0.36 | 0.87 ± 13.7 | 0.930/ns. | 1.25 ± 0.31 | 1.28 ± 0.30 | 2.22 ± 12.23 | 0.876/ns. |
| ADC0,200,400 | 1.25 ± 0.42 | 1.20 ± 0.35 | −2.15 ± 8.07 | 0.968/0.016 | 1.73 ± 0.26 | 1.70 ± 0.26 | −1.51 ± 4.81 | 0.956/ns. |
| ADC0,200,800 | 1.03 ± 0.35 | 1.02 ± 0.35 | −1.55 ± 4.86 | 0.980/ns. | 1.46 ± 0.28 | 1.46 ± 0.28 | −0.24 ± 4.53 | 0.956/ns. |
| ADC0,400,800 | 1.06 ± 0.35 | 1.04 ± 0.35 | −1.46 ± 4.91 | 0.979/ns. | 1.49 ± 0.28 | 1.48 ± 0.27 | −0.24 ± 4.43 | 0.958/ns. |
| ADC0,200 | 1.38 ± 0.45 | 1.35 ± 0.44 | −0.70 ± 8.93 | 0.951/ns. | 1.86 ± 0.26 | 1.86 ± 0.27 | −0.01 ± 7.08 | 0.910/ns. |
| ADC0,800 | 1.06 ± 0.35 | 1.04 ± 0.35 | −1.46 ± 4.90 | 0.979/ns. | 1.49 ± 0.28 | 1.48 ± 0.27 | −0.24 ± 4.43 | 0.955/ns. |
| ADC200,800 | 0.83 ± 0.28 | 0.80 ± 0.33 | −2.53 ± 23.1 | 0.902/ns. | 1.17 ± 0.30 | 1.20 ± 0.30 | 3.72 ± 9.84 | 0.895/0.05 |
ADC = Apparent Diffusion Coefficient; G-CA = gadolinium based intravenous contrast agent; Pre-G-CA = measurement before the intravenously administered contrast agent; Post-G-CA = measurement after the intravenously administered contrast agent.
Figure 2Box-and-whiskers plots showing pre-contrast (white boxes) and post-contrast (gray boxes) breast lesion ADC values generated using different b value combinations. The boxes represent the interquartile range (the data from the 25th to the 75th percentile); the horizontal line in the box refers to the median ADC value and the whiskers show the range (minimum and maximum) of the mean ADC values.
Indicate lesion type and mean ± standard deviation apparent diffusion coefficient (ADC) values by using different b value combinations before and after the gadolinium based contrast agent administration (G-CA).
| b Value Protocol | Tissue type | |||||||
|---|---|---|---|---|---|---|---|---|
| Malignant (N = 25) | Benign (N = 7) | |||||||
| Pre-G-CA (x 10−3 mm2/s) | Post-G-CA (x 10−3 mm2/s) | Mean change (%) | r/ | Pre-G-CA (x 10−3 mm2/s) | Post-G-CA (x 10−3 mm2/s) | Mean change (%) | r/ | |
| ADC0,200,400,600,800 | 0.93 ± 0.28 | 0.92 ± 0.28 | −1.60 ± 4.51 | 0.975/0.042 | 1.47 ± 0.25 | 1.46 ± 0.30 | −1.49 ± 5.07 | 1.00/ns. |
| ADC400,600,800 | 0.76 ± 0.24 | 0.76 ± 0.27 | 0.21 ± 13.3 | 0.902/ns. | 1.25 ± 0.17 | 1.30 ± 0.33 | 3.23 ± 15.80 | 0.857/ns. |
| ADC0,200,400 | 1.12 ± 0.34 | 1.08 ± 0.28 | −1.43 ± 8.61 | 0.947/ns. | 1.71 ± 0.34 | 1.62 ± 0.29 | −4.73 ± 5.46 | 0.750/ns. |
| ADC0,200,800 | 0.92 ± 0.28 | 0.90 ± 0.27 | −1.57 ± 4.64 | 0.969/0.040 | 1.45 ± 0.24 | 1.44 ± 0.30 | −1.50 ± 6.00 | 1.000/ns. |
| ADC0,400,800 | 0.94 ± 0.28 | 0.92 ± 0.27 | −1.36 ± 4.82 | 0.970/ns. | 1.48 ± 0.25 | 1.46 ± 0.30 | −1.79 ± 5.61 | 0.964/ns. |
| ADC0,200 | 1.25 ± 0.37 | 1.23 ± 0.30 | 0.40 ± 9.60 | 0.925/ns. | 1.87 ± 0.35 | 1.78 ± 0.34 | −4.62 ± 4.55 | 0.929/ns. |
| ADC0,800 | 0.94 ± 0.28 | 0.92 ± 0.27 | −1.36 ± 4.82 | 0.970/ns. | 1.48 ± 0.25 | 1.46 ± 0.30 | −1.79 ± 5.61 | 0.964/ns. |
| ADC200,800 | 0.73 ± 0.23 | 0.69 ± 0.23 | −3.32 ± 24.0 | 0.875/ns. | 1.17 ± 0.18 | 1.19 ± 0.35 | 0.31 ± 20.8 | 0.929/ns. |
ADC = Apparent Diffusion Coefficient; G-CA = gadolinium based intravenous contrast agent; Pre-G-CA = measurement before the intravenously administered contrast agent; Post-G-CA = measurement after the intravenously administered contrast agent.
Contrast-to-noise ratio (CNR) means ± standard deviations in images obtained before and after the administration of gadolinium based intravenous contrast agent (G-CA) using fat as the reference value.
| b Value Protocol | Pre-G-CA CRN | Post-G-CA CRN |
|
|---|---|---|---|
| ADC0,200,400,600,800 | 0.55 ± 0.07 | 0.55 ± 0.09 | ns. |
| ADC400,600,800 | 0.64 ± 0.13 | 0.62 ± 0.13 | ns. |
| ADC0,200,400 | 0.50 ± 0.18 | 0.51 ± 0.23 | ns. |
| ADC0,200,800 | 0.54 ± 0.07 | 0.55 ± 0.10 | ns. |
| ADC0,400,800 | 0.54 ± 0.09 | 0.54 ± 0.12 | ns. |
| ADC0,200 | 0.57 ± 0.53 | 0.64 ± 0.78 | ns. |
| ADC0,800 | 0.54 ± 0.09 | 0.54 ± 0.12 | ns. |
| ADC200,800 | 0.54 ± 0.26 | 0.54 ± 0.24 | ns. |
Pre-G-CA = measurement before the intravenously administered contrast agent; Post-G-CA = measurement after the intravenously administered contrast agent; CNR = Contrast-to-Noise Ratio; ADC = Apparent Diffusion Coefficient. P value indicates the statistical difference between the CNRs in pre- and post-contrast images.
Review of the literature about the effects of contrast agents on apparent diffusion coefficient (ADC) values in breast tumors.
| Number of patients (N of lesions) | Rubesova | Yuen | Janka R | Nguyen | Present study |
|---|---|---|---|---|---|
| 5 (7; 2 benign, 5 malignant) | 19 (19; 19 malignant) | n/a (35; 10 benign, 25 malignant) | 19 (19; 19 malignant of which 14 mass-like enhancing and 5 NMLE lesions) | 27 (32; 7 benign, 25 malignant, all lesions mass-like enhancing lesions) | |
| MRI field strength, vendor, manufacturer | 1.5 T, Magnetom Symphony, Siemens | 1.5 T, Gyroscan Intera Nova Dual, Philips | 1.5 T, Magnetom Avanto, Siemens | 3.0 T, Achieva Tx, Philips | Achieva TX, Philips, Achieva Tx, Philips |
| b value combinations | n/a | ADC0,1000 | ADC50,400,800 | ADC0,100,800; ADC0,800; ADC100,800 | ADC0,200,400,600,800; ADC400, 600,800; ADC0,200,400; ADC0,200,800; ADC0,400,800; ADC0,200; ADC0,800; ADC200,800 |
| Contrast agent (dose mmol/kg) | Gadodiamide (0.3) | Gadopentetate dimeglumine (0.1) | Gadobutrol (0.1) | Gadoteridol (0.1) | Gadoterate meglumine (0.2) |
| DCE-MRI acquisition time (minutes) | 15 | 8 | 6 | 9 | 6 |
| Median TR/TE (ms) | 2600/110 | 1206/71 | 4100/98 | 5336/61 | 6982/95 |
| Mean ± SD pre-contrast/post-contrast ADC values mm2/s ( | n/a (ns.) | 1.35 ± 0.38/1.04 ± 0.34 ( | Benign lesions: 1.99 ± 0.37/1.97 ± 0.30 ( | ADC0,100,800: 1.10/1.10 ( | All lesions ADC0,200,400,600,800: 1.05 ± 0.35/1.03 ± 0.36 ( |
N = number; n/a = not applicable information; ns. = not significant; TR = repetition time; TE = echo time; DCE-MRI = dynamic contrast enhanced magnetic resonance imaging.