Dosimetric quantification of dose fall-off in liver SBRT planning using dual photon energy IMRT.

J Henry1, C Moreno1, R L Crownover1, D Baacke1, N Papanikolaou1, A N Gutierrez1.   

Abstract

INTRODUCTION: The purpose of this study was to dosimetrically compare 6 MV, 10 MV and a hybrid of 6 and 10 MV photon beam energies in liver stereotactic body radiotherapy (SBRT) patients using a fixed-field IMRT technique. The objectives of the study were to evaluate dosimetric differences in the target volume and investigate if dose fall-off could be improved with 10MV beam energy. METHODS AND MATERIALS: Sixteen (n=16) liver SBRT patients previously treated using a non-coplanar, fixed-field IMRT technique with 6 MV were replanned using 10 MV and dual photon energy (DE). Plans were generated in Pinnacle3 using a Novalis Tx with HD120 MLC. For each patient, three plans with the same beam geometry were created using 6 MV, 10 MV and DE. For DE plans, the central axis effective depth from each beam was calculated and the values averaged. Beams with an effective depth greater than the average were assigned to 10 MV. All patients were optimized with the same planning objectives and normalized such that 98% of the target received 100% of prescription dose. Metrics used for comparison were the homogeneity index, conformity indices, and dose fall-off parameters at various isodose levels.
RESULTS: The three techniques showed comparable PTV conformity and inhomogeneity for all patients-differences in the median values 「0.6%. With regard to dose fall-off, no statistically significant differences were noted among the techniques for R80, R60 and R50; however, 10 MV showed statistical significance in the lowest median values of R40, R30, and R20. Ten MV and DE plans also demonstrated a statistically significant reduction in the total number of monitor units (MU) of 14.9% (p 「0.01) and 12.0% (p 「0.01) as compared to 6 MV, respectively.
CONCLUSION: Both dual energy and 10 MV photon beams had similar PTV dosimetric characteristics to 6 MV for liver SBRT but findings show faster dose fall-off for 10 MV and DE plans at the 40%, 30%, and 20% prescription isodose levels.

Entities:  

Keywords:  IMRT; SBRT; liver cancer; photon energy

Year:  2016        PMID: 29296439      PMCID: PMC5658876     

Source DB:  PubMed          Journal:  J Radiosurg SBRT


  16 in total

1.  Beam orientation optimization in intensity-modulated radiation treatment planning.

Authors:  A B Pugachev; A L Boyer; L Xing
Journal:  Med Phys       Date:  2000-06       Impact factor: 4.071

2.  Out-of-field photon and neutron dose equivalents from step-and-shoot intensity-modulated radiation therapy.

Authors:  Stephen F Kry; Mohammad Salehpour; David S Followill; Marilyn Stovall; Deborah A Kuban; R Allen White; Isaac I Rosen
Journal:  Int J Radiat Oncol Biol Phys       Date:  2005-07-15       Impact factor: 7.038

3.  Volumetric arc intensity-modulated therapy for spine body radiotherapy: comparison with static intensity-modulated treatment.

Authors:  Q Jackie Wu; Sua Yoo; John P Kirkpatrick; Danthai Thongphiew; Fang-Fang Yin
Journal:  Int J Radiat Oncol Biol Phys       Date:  2009-09-03       Impact factor: 7.038

4.  A conformation number to quantify the degree of conformality in brachytherapy and external beam irradiation: application to the prostate.

Authors:  A van't Riet; A C Mak; M A Moerland; L H Elders; W van der Zee
Journal:  Int J Radiat Oncol Biol Phys       Date:  1997-02-01       Impact factor: 7.038

5.  Commissioning and dosimetric characteristics of TrueBeam system: composite data of three TrueBeam machines.

Authors:  Zheng Chang; Qiuwen Wu; Justus Adamson; Lei Ren; James Bowsher; Hui Yan; Andrew Thomas; Fang-Fang Yin
Journal:  Med Phys       Date:  2012-11       Impact factor: 4.071

6.  Clinical evaluation of an immbolization system for stereotactic body radiotherapy using helical tomotherapy.

Authors:  Alonso N Gutiérrez; Sotirios Stathakis; Richard Crownover; Carlos Esquivel; Chengyu Shi; Niko Papanikolaou
Journal:  Med Dosim       Date:  2010-05-01       Impact factor: 1.482

7.  Increasing dose gradient and uniformity in small fields using modulation: theory and prototypes for cone-based stereotactic radiosurgery.

Authors:  Edward T Bender
Journal:  Med Phys       Date:  2014-05       Impact factor: 4.071

8.  Volumetric-modulated arc therapy for stereotactic body radiotherapy of lung tumors: a comparison with intensity-modulated radiotherapy techniques.

Authors:  Andrea Holt; Corine van Vliet-Vroegindeweij; Anton Mans; José S Belderbos; Eugène M F Damen
Journal:  Int J Radiat Oncol Biol Phys       Date:  2011-02-06       Impact factor: 7.038

9.  Effect of MLC leaf width and PTV margin on the treatment planning of intensity-modulated stereotactic radiosurgery (IMSRS) or radiotherapy (IMSRT).

Authors:  Jenghwa Chang; Kamil M Yenice; Kailiu Jiang; Margie Hunt; Ashwatha Narayana
Journal:  Med Dosim       Date:  2008-07-01       Impact factor: 1.482

10.  Multi-institutional phase I/II trial of stereotactic body radiation therapy for liver metastases.

Authors:  Kyle E Rusthoven; Brian D Kavanagh; Higinia Cardenes; Volker W Stieber; Stuart H Burri; Steven J Feigenberg; Mark A Chidel; Thomas J Pugh; Wilbur Franklin; Madeleine Kane; Laurie E Gaspar; Tracey E Schefter
Journal:  J Clin Oncol       Date:  2009-03-02       Impact factor: 44.544

View more
  1 in total

1.  Treatment plan comparison of volumetric-modulated arc therapy to intensity-modulated radiotherapy in lung stereotactic body radiotherapy using either 6- or 10-MV photon energies.

Authors:  Zhigong Wei; Xingchen Peng; Ling He; Jingjing Wang; Zheran Liu; Jianghong Xiao
Journal:  J Appl Clin Med Phys       Date:  2022-07-09       Impact factor: 2.243

  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.