The use of serial patient blood gas, electrolyte and glucose results to derive biologic variation: a new tool to assess the acceptability of intensive care unit testing.

George S Cembrowski1, David V Tran, Trefor N Higgins.   

Abstract

BACKGROUND: Most estimates of biologic variation (s(b)) are based on periodically acquiring and storing specimens, followed by analysis within a single analytic run. We demonstrate for many intensive care unit (ICU) tests, only patient results need be statistically analyzed to provide reliable estimates of s(b).
METHODS: Over 11 months, approximately 28,000 blood gas measurements (including electrolyte panels and glucose) were performed on one of two Radiometer ABL800 FLEX analyzers (Radiometer, Copenhagen, Denmark) from 1676 ICU patients. We tabulated the measurements of paired intra-patient blood samples drawn within 24 h of each other. After removal of outliers, we calculated the standard deviations of duplicates (SDD) of the intra-patient pairs grouped in 2-h intervals: 0-2 h, 2-4 h, 4-6 h, … 20-22 h and 22-24 h. The SDDs were then regressed against the time intervals of 2-14 h; extrapolation to zero time represents the sum of s(b) and short-term analytic variation (s(a)).
RESULTS: Substitution of experimentally derived analytic error permitted the calculation of coefficient of variation (biologic) (CV(b)) (100 s(b)/mean): pH, 0.3%; pCO(2), 5.7%; pO(2), 13%; Na(+), 0.6%; K(+), 4.8%; Cl(-), 0.8%; HCO(3)(-), 3.2%; iCa(++), 2.4%; and glucose, 10.3%. The CV(b) of the electrolytes very closely matches the lowest estimates obtained in the usual manner.
CONCLUSIONS: Derivation of the ratio of biologic to analytic variation indicates that the ABL800 is extremely suitable for ICU testing. This analysis should be extended to other point of care instrument systems.

Entities:  

Mesh:

Substances:

Year:  2010        PMID: 20716012     DOI: 10.1515/CCLM.2010.286

Source DB:  PubMed          Journal:  Clin Chem Lab Med        ISSN: 1434-6621            Impact factor:   3.694


  11 in total

1.  A new device to relieve venipuncture pain can affect haematology test results.

Authors:  Gabriel Lima-Oliveira; Giuseppe Lippi; Gian Luca Salvagno; Marise Danielle Raulino Campelo; Katharyne Soares Adala Tajra; Flavio dos Santos Gomes; Carlos David Valentim; Sylvio José Colonna Romano; Geraldo Picheth; Gian Cesare Guidi
Journal:  Blood Transfus       Date:  2013-07-18       Impact factor: 3.443

2.  The Glucose Measurement Industry and Hemoglobin A1c: An Opportunity for Creative Destruction.

Authors:  George Cembrowski
Journal:  J Diabetes Sci Technol       Date:  2015-10-18

3.  Quality impact on diagnostic blood specimen collection using a new device to relieve venipuncture pain.

Authors:  Gabriel Lima-Oliveira; Giuseppe Lippi; Gian Luca Salvagno; Martina Montagnana; Geraldo Picheth; Gian Cesare Guidi
Journal:  Indian J Clin Biochem       Date:  2013-04-03

4.  Impact of retrospective calibration algorithms on hypoglycemia detection in newborn infants using continuous glucose monitoring.

Authors:  Matthew Signal; Aaron Le Compte; Deborah L Harris; Philip J Weston; Jane E Harding; J Geoffrey Chase
Journal:  Diabetes Technol Ther       Date:  2012-08-02       Impact factor: 6.118

5.  Influence of a regular, standardized meal on clinical chemistry analytes.

Authors:  Gabriel Lima-Oliveira; Gian Luca Salvagno; Giuseppe Lippi; Matteo Gelati; Martina Montagnana; Elisa Danese; Geraldo Picheth; Gian Cesare Guidi
Journal:  Ann Lab Med       Date:  2012-06-20       Impact factor: 3.464

6.  Using stochastic modelling to identify unusual continuous glucose monitor measurements and behaviour, in newborn infants.

Authors:  Matthew Signal; Aaron Le Compte; Deborah L Harris; Phil J Weston; Jane E Harding; J Geoffrey Chase
Journal:  Biomed Eng Online       Date:  2012-08-06       Impact factor: 2.819

7.  Preanalytical management: serum vacuum tubes validation for routine clinical chemistry.

Authors:  Gabriel Lima-Oliveira; Giuseppe Lippi; Gian Luca Salvagno; Martina Montagnana; Geraldo Picheth; Gian Cesare Guidi
Journal:  Biochem Med (Zagreb)       Date:  2012       Impact factor: 2.313

8.  Incorrect order of draw could be mitigate the patient safety: a phlebotomy management case report.

Authors:  Gabriel Lima-Oliveira; Giuseppe Lippi; Gian Luca Salvagno; Martina Montagnana; Geraldo Picheth; Gian Cesare Guidi
Journal:  Biochem Med (Zagreb)       Date:  2013       Impact factor: 2.313

9.  Preanalytical conditions of point-of-care testing in the intensive care unit are decisive for analysis reliability.

Authors:  Adrien Auvet; Fabien Espitalier; Leslie Grammatico-Guillon; Mai-Anh Nay; Djilali Elaroussi; Marc Laffon; Christian R Andres; Annick Legras; Stephan Ehrmann; Pierre-François Dequin; Chantal Gendrot; Antoine Guillon
Journal:  Ann Intensive Care       Date:  2016-06-24       Impact factor: 6.925

10.  The effect of acute taurine ingestion on 4-km time trial performance in trained cyclists.

Authors:  Ryan Ward; Craig A Bridge; Lars R McNaughton; S Andy Sparks
Journal:  Amino Acids       Date:  2016-07-05       Impact factor: 3.520

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.