Comparison between two portable hemoglobinometers and a reference method to verify the reliability of screening in blood donors.

Renata Cristina Messores Rudolf-Oliveira1, Kauê Taneli Gonçalves, Mariana Lenhani Martignago, Vanessa Mengatto, Pâmela Cristina Gaspar, Jovino dos Santos Ferreira, Junara Cerutti de Castro, Maria Luiza Bazzo, Maria Cláudia Santos-Silva.   

Abstract

Portable hemoglobinometers are used to determine hemoglobin level, but there are conflicting reports regarding their accuracy. The aim of this study was to compare results from two portable hemoglobinometers (HemoCue® and Hemo-Control) with an automated hematology analyzer (Sysmex XE-2100D) to determine if the screening of blood donors is reliable. A total of 426 blood donors' samples were studied and on average the Hb content measured in capillary blood samples was higher than that found in venous blood samples. Hemoglobinometers can be employed as a method to screen blood donors, but critical values should be confirmed in an automated hematology analyzer.
Copyright © 2013 Elsevier Ltd. All rights reserved.

Keywords:  Automated hematology analyzer; Blood donors; Hemoglobin; Hemoglobinometers

Mesh:

Year:  2013        PMID: 24119714     DOI: 10.1016/j.transci.2013.09.004

Source DB:  PubMed          Journal:  Transfus Apher Sci        ISSN: 1473-0502            Impact factor:   1.764


  3 in total

Review 1.  Methods and analyzers for hemoglobin measurement in clinical laboratories and field settings.

Authors:  Ralph D Whitehead; Zuguo Mei; Carine Mapango; Maria Elena D Jefferds
Journal:  Ann N Y Acad Sci       Date:  2019-06-04       Impact factor: 5.691

Review 2.  Techniques used for the screening of hemoglobin levels in blood donors: current insights and future directions.

Authors:  Rajendra Chaudhary; Anju Dubey; Atul Sonker
Journal:  J Blood Med       Date:  2017-07-03

3.  Comparison of haemoglobin assessments by HemoCue and two automated haematology analysers in young Laotian children.

Authors:  Guy-Marino Hinnouho; Maxwell A Barffour; K Ryan Wessells; Kenneth H Brown; Sengchanh Kounnavong; Bigphone Chanhthavong; Kethmany Ratsavong; Chidchamai Kewcharoenwong; Sonja Y Hess
Journal:  J Clin Pathol       Date:  2017-12-02       Impact factor: 3.411

  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.