Literature DB >> 27807113

Breast Radiation Exposure in Female Orthopaedic Surgeons.

Lindsey C Valone1, Monique Chambers2, Lisa Lattanza3,4, Michelle A James3,2,4.   

Abstract

BACKGROUND: Breast cancer prevalence is higher among female orthopaedic surgeons compared with U.S. women. The most common breast cancer site, the upper outer quadrant (UOQ), may not be adequately shielded from intraoperative radiation. Factors associated with higher breast radiation exposure (protective apron size and type, surgeon position, and C-arm position) have yet to be established.
METHODS: An anthropomorphic torso phantom, simulating the female surgeon, was placed adjacent to a standard operating table. Dosimeters were placed over the UOQ and lower inner quadrant (LIQ) of the breast, bilaterally. Scatter radiation dose-equivalent rates were measured during continuous fluoroscopy to a pelvic phantom (simulating the patient). Four apron sizes (small, medium, large, and extra-large), 2 apron types (cross-back and vest), 2 surgeon positions (facing the table and 90° to the table), and 2 C-arm positions (anteroposterior and cross-table lateral projection) were tested.
RESULTS: The median dose-equivalent rate of scatter radiation to the UOQ (0.40 mrem/hr) was higher than that to the LIQ of the breast (0.06 mrem/hr) across all testing, although this was not statistically significant (p = 0.05). The cross-back aprons provided higher protection to the LIQ compared with the vests (p < 0.05). Lead protection in sizes that were too small or too large for the torso had higher breast radiation dose-equivalent rates. C-arm cross-table lateral projection was associated with higher breast radiation exposure (0.98 mrem/hr) compared with anteroposterior projection (0.13 mrem/hr) (p < 0.001).
CONCLUSIONS: Breast radiation exposure is higher in a C-arm lateral projection compared with an anteroposterior projection. Higher dose-equivalent rates were observed for the UOQ compared with the LIQ of the breast and for aprons that were too small or too large, although these differences did not reach significance. Factors that may reduce radiation exposure include lead protection of appropriate size and distancing the axilla from the patient and x-ray tube. CLINICAL RELEVANCE: Increased breast cancer prevalence has been reported for female orthopaedic surgeons. The UOQ of the breast may be at risk for intraoperative radiation exposure. Methods of reducing exposure are warranted.
Copyright © 2016 by The Journal of Bone and Joint Surgery, Incorporated.

Entities:  

Mesh:

Year:  2016        PMID: 27807113     DOI: 10.2106/JBJS.15.01167

Source DB:  PubMed          Journal:  J Bone Joint Surg Am        ISSN: 0021-9355            Impact factor:   5.284


  11 in total

Review 1.  Percutaneous treatment of hallux valgus: What's the evidence? A systematic review.

Authors:  S Caravelli; M Mosca; S Massimi; G G Costa; M Lo Presti; M Fuiano; A Grassi; S Zaffagnini
Journal:  Musculoskelet Surg       Date:  2017-10-28

2.  Elevated Radiation Exposure Associated With Above Surface Flat Detector Mini C-Arm Use.

Authors:  Dennis P Martin; Talia Chapman; Christopher Williamson; Brian Tinsley; Asif M Ilyas; Mark L Wang
Journal:  Hand (N Y)       Date:  2017-11-22

3.  Pedicle screw insertion with patient-specific 3D-printed guides based on low-dose CT scan is more accurate than free-hand technique in spine deformity patients: a prospective, randomized clinical trial.

Authors:  Riccardo Cecchinato; Pedro Berjano; Alberto Zerbi; Marco Damilano; Andrea Redaelli; Claudio Lamartina
Journal:  Eur Spine J       Date:  2019-04-20       Impact factor: 3.134

4.  A Preliminary Survey of Women Orthopaedic Surgeons on Awareness of Radiation Safety Practice and Breast Cancer Risk in India.

Authors:  Ansu Thankam John; Vrisha Madhuri
Journal:  Indian J Orthop       Date:  2021-09-22       Impact factor: 1.033

5.  Radiation Exposure and Case Characteristics in National Sample of Female Orthopaedic Trauma and Arthroplasty Surgeons.

Authors:  Cara H Lai; Andrea Finlay; Lisa K Cannada; Antonia F Chen; Loretta B Chou
Journal:  Iowa Orthop J       Date:  2020

6.  Procedural Outcomes of Double Vs. Single Fluoroscopy for Fixing Supracondylar Humerus Fractures in Children: A Case-Control Study.

Authors:  Burak Gunaydın; Ali Turgut; Abdulkadir Sarı; Yaşar Mahsut Dinçel; Cagatay Tekin; Yavuz Selim Kabukcuoglu
Journal:  Indian J Orthop       Date:  2020-05-31       Impact factor: 1.251

7.  A Report on Occupational Ionizing Radiation Exposure by an Orthopedic Surgeon in a National Health-care Setting - Clinical Case Perspective.

Authors:  Abdulbaset Abosala
Journal:  J Orthop Case Rep       Date:  2018 May-Jun

8.  Association between occupational testicular radiation exposure and lower male sex ratio of offspring among orthopedic surgeons.

Authors:  Yasukazu Hijikata; Masayuki Nakahara; Akira Kusumegi; Junji Morii; Naoki Okubo; Nozomi Hatano; Yuichi Takahashi
Journal:  PLoS One       Date:  2021-12-31       Impact factor: 3.240

Review 9.  Improved Radiation Exposure Monitoring of Orthopaedic Residents After Institution of a Personalized Lead Protocol.

Authors:  Patrick A Massey; Mitchell E Myers; Ryan D Guedry; Michael T Lowery; Kevin J Perry; R Shane Barton
Journal:  JB JS Open Access       Date:  2022-03-21

10.  Three-Dimensional Printing for Preoperative Planning and Pedicle Screw Placement in Adult Spinal Deformity: A Systematic Review.

Authors:  Cesar D Lopez; Venkat Boddapati; Nathan J Lee; Marc D Dyrszka; Zeeshan M Sardar; Ronald A Lehman; Lawrence G Lenke
Journal:  Global Spine J       Date:  2020-08-07
View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.