| Literature DB >> 29159206 |
Ninita Lindfors1,2, Henrik Lund1, Hans Johansson3, Annika Ekestubbe1.
Abstract
OBJECTIVES: The aim of this in vitro study was to evaluate how a deviation from the horizontal plane, affects the image quality in two different CBCT-devices.Entities:
Keywords: CBCT; FOV; Image quality; Patient position; Radiation dose
Year: 2017 PMID: 29159206 PMCID: PMC5675893 DOI: 10.1016/j.ejro.2017.10.001
Source DB: PubMed Journal: Eur J Radiol Open ISSN: 2352-0477
Fig. 1Illustrates the difference in the phantom head position.
Specifications for the different image parameter settings for the different CBCT-devices.
| Device | Voxel size (mm) | Slice thickness (mm) | No of axial slices in each scan | No of projections during scan | Exposure time (s) |
|---|---|---|---|---|---|
| Accuitomo | 0.125 | 0.375 | 107 | 586 | 17.5 |
| Veraview | 0.125 | 0.375 | 115 | 342 | 9.4 |
| Accuitomo | 0.160 | 0.480 | 167 | 586 | 17.5 |
| Veraview | 0.160 | 0.480 | 167 | 342 | 9.4 |
Accuitomo F80.
Veraviewepocs 3D R100.
Reuleaux Full Arch is an area shaped like a convex triangle. By more closely matching the natural dental arch form, this FOV allows a more complete scan of the maxilla and mandible.
Fig. 2Illustrates the difference between the areas of the large FOV in (A) Accuitomo and (B) Veraviewepocs. The Veraview has a rouleaux shape that is more adjusted the shape of the jaws.
Fig. 3The setting of different regions of interest (ROIs). Hounsfield value (HU) represent the CT value used to calculate the CNR and deviation (DEV) represent the standard deviation (SD) value also used in the same formula. Unilateral settings in the small FOV (A) and bilaterally in the large FOV (B).
Technical specifications used in the PCXMC simulation program (version 2.0).
| Accuitomo F80 | Veraviewepocs 3D R100 | |
|---|---|---|
| Source-to-detector distance (mm) | 710 | 519 |
| Source-to-isocenter distance (mm) | 500 | 346 |
| Anod angle (°) | 5 | 5 |
| Rotation angle (°) | 360 | 180 |
| Detector field (WxH, mm) | 67 × 67 (400 × 400) | 62 × 71 (400 × 400) |
| 117 × 123 (800 × 800) | 125 × 130 (100 × 80) | |
| Exposure time (s) | 17.5 | 9.4 |
| Total filtration | 3.9 mm Al | 2.5 mm Al + 0.2 mm Cu |
| Frames/basal images | 586 | 343 |
Exposure parameters and DAP values (mGycm2).
| 75 kV | 80 kV | 85 kV | 90 kV | 80 kV | 80 kV | 85 kV | 85 kV | 90 kV | 90 kV | |
|---|---|---|---|---|---|---|---|---|---|---|
| 5 mA | 5 mA | 5 mA | 5 mA | 2 mA | 8 mA | 2 mA | 8 mA | 2 mA | 8 mA | |
| Accuitomo | 288 | 329 | 367 | 410 | 137 | 519 | 153 | 581 | 164 | 656 |
| 40 mm × 40 mm | ||||||||||
| Veraview | 166 | 201 | 236 | 276 | 84 | 318 | 98 | 377 | 110 | 441 |
| 40 mm × 40 mm | ||||||||||
| Accuitomo | 1046 | 1212 | 1351 | 1531 | 504 | 1913 | 562 | 2139 | 612 | 2450 |
| 80 mm × 80 mm | ||||||||||
| Veraview | 546 | 657 | 764 | 890 | 274 | 1040 | 319 | 1209 | 356 | 1424 |
| 100 mm × 80 mm |
Calculated values (extrapolated from measured values).
Accuitomo F80.
Veraviewepocs 3D R100.
Calculated effective dose (E = DAP × EDLP) values (in mSv), based on weighting factors of ICRP Publication 103 (2007).
| EDLP | 75 kV | 80 kV | 85 kV | 90 kV | 80 kV | 80 kV | 85 kV | 85 kV | 90 kV | 90 kV | |
|---|---|---|---|---|---|---|---|---|---|---|---|
| 5 mA | 5 mA | 5 mA | 5 mA | 2 mA | 8 mA | 2 mA | 8 mA | 2 mA | 8 mA | ||
| Accuitomo | 0.156 | 0.045 | 0.051 | 0.057 | 0.064 | 0.021 | 0.081 | 0.024 | 0.091 | 0.026 | 0.102 |
| 40 mm × 40 mm | |||||||||||
| Veraview | 0.212 | 0.035 | 0.043 | 0.050 | 0.059 | 0.018 | 0.067 | 0.021 | 0.080 | 0.023 | 0.093 |
| 40 mm × 40 mm | |||||||||||
| Accuitomo | 0.139 | 0.145 | 0.168 | 0.188 | 0.213 | 0.070 | 0.266 | 0.078 | 0.297 | 0.085 | 0.340 |
| 80 mm × 80 mm | |||||||||||
| Veraview | 0.166 | 0.091 | 0.109 | 0.127 | 0.148 | 0.045 | 0.173 | 0.053 | 0.201 | 0.059 | 0.236 |
| 100 mm × 80 mm |
Conversion factor. Calculated with PCXMC 2.0 program (mSv/Gycm2).
Accuitomo F 80.
Veraviewepocs 3D R100.
Analysis of variance.
| Effect | p-value |
|---|---|
| CBCT-device and kV | 0.9709 |
| CBCT-device and mA | 0.9001 |
| CBCT-device and FOV | 0.0016 |
| CBCT-device and patient position | 0.0215 |
| kV and mA | 0.6936 |
| kV and FOV | 0.9851 |
| kV and patient position | 0.7487 |
| mA and FOV | 0.0424 |
| mA and patient position | 0.0829 |
| FOV and patient position | 0.0014 |
Significant covariance.
Mean CNR value analysing CBCT device and patient position.
| CBCT-device | Patient position | No of observations | Mean CNR | Std Error |
|---|---|---|---|---|
| Accuitomo F80 | Horisontally | 78 | 7.21 | 0.253 |
| Accuitomo F80 | Tilted | 78 | 9.01 | 0.253 |
| Veraviewepocs 3D R100 | Horisontally | 78 | 7.87 | 0.253 |
| Veraviewepocs 3D R100 | Tilted | 78 | 8.50 | 0.253 |
Mean CNR value analysing FOV and patient position.
| FOV | Patient position | Number of observations | Mean CNR | Std Error |
|---|---|---|---|---|
| Small | Horisontal | 78 | 7.33 | 0.253 |
| Small | Tilted | 78 | 9.35 | 0.253 |
| Large | Horisontal | 78 | 7.75 | 0.253 |
| Large | Tilted | 78 | 8.15 | 0.253 |
40 mm × 40 mm Accuitomo F80 and Veraviewepocs 3D R100.
80 mm × 80 mm Accuitomo F80 and 100 mm × 80 mm Veraviewepocs 3D R100.
Mean CNR value analysing mA and FOV.
| mA | FOV | No of observations | Mean CNR | Std Error |
|---|---|---|---|---|
| 2.00 | small | 36 | 6.44 | 0.373 |
| large | 36 | 7.22 | 0.373 | |
| 5.00 | small | 84 | 8.57 | 0.244 |
| large | 84 | 7.91 | 0.244 | |
| 8.00 | small | 36 | 9.71 | 0.373 |
| large | 36 | 8.79 | 0.373 |
40 mm × 40 mm Accuitomo F80 and Veraviewepocs 3D R100.
80 mm × 80 mm Accuitomo F80 and 100 mm × 80 mm Veraviewepocs 3D R100.