Accelerating volumetric cine MRI (VC-MRI) using undersampling for real-time 3D target localization/tracking in radiation therapy: a feasibility study.

Wendy Harris1, Fang-Fang Yin, Chunhao Wang, You Zhang, Jing Cai, Lei Ren.   

Abstract

PURPOSE: To accelerate volumetric cine MRI (VC-MRI) using undersampled 2D-cine MRI to provide real-time 3D guidance for gating/target tracking in radiotherapy.
METHODS: 4D-MRI is acquired during patient simulation. One phase of the prior 4D-MRI is selected as the prior images, designated as MRIprior. The on-board VC-MRI at each time-step is considered a deformation of the MRIprior. The deformation field map is represented as a linear combination of the motion components extracted by principal component analysis from the prior 4D-MRI. The weighting coefficients of the motion components are solved by matching the corresponding 2D-slice of the VC-MRI with the on-board undersampled 2D-cine MRI acquired. Undersampled Cartesian and radial k-space acquisition strategies were investigated. The effects of k-space sampling percentage (SP) and distribution, tumor sizes and noise on the VC-MRI estimation were studied. The VC-MRI estimation was evaluated using XCAT simulation of lung cancer patients and data from liver cancer patients. Volume percent difference (VPD) and Center of Mass Shift (COMS) of the tumor volumes and tumor tracking errors were calculated.
RESULTS: For XCAT, VPD/COMS were 11.93  ±  2.37%/0.90  ±  0.27 mm and 11.53  ±  1.47%/0.85  ±  0.20 mm among all scenarios with Cartesian sampling (SP  =  10%) and radial sampling (21 spokes, SP  =  5.2%), respectively. When tumor size decreased, higher sampling rate achieved more accurate VC-MRI than lower sampling rate. VC-MRI was robust against noise levels up to SNR  =  20. For patient data, the tumor tracking errors in superior-inferior, anterior-posterior and lateral (LAT) directions were 0.46  ±  0.20 mm, 0.56  ±  0.17 mm and 0.23  ±  0.16 mm, respectively, for Cartesian-based sampling with SP  =  20% and 0.60  ±  0.19 mm, 0.56  ±  0.22 mm and 0.42  ±  0.15 mm, respectively, for radial-based sampling with SP  =  8% (32 spokes).
CONCLUSIONS: It is feasible to estimate VC-MRI from a single undersampled on-board 2D cine MRI. Phantom and patient studies showed that the temporal resolution of VC-MRI can potentially be improved by 5-10 times using a 2D cine image acquired with 10-20% k-space sampling.

Entities:  

Mesh:

Year:  2017        PMID: 29087963      PMCID: PMC5756137          DOI: 10.1088/1361-6560/aa9746

Source DB:  PubMed          Journal:  Phys Med Biol        ISSN: 0031-9155            Impact factor:   3.609


  27 in total

1.  Controlled aliasing in parallel imaging results in higher acceleration (CAIPIRINHA) for multi-slice imaging.

Authors:  Felix A Breuer; Martin Blaimer; Robin M Heidemann; Matthias F Mueller; Mark A Griswold; Peter M Jakob
Journal:  Magn Reson Med       Date:  2005-03       Impact factor: 4.668

2.  Robustness of quantitative compressive sensing MRI: the effect of random undersampling patterns on derived parameters for DCE- and DSC-MRI.

Authors:  David S Smith; Xia Li; James V Gambrell; Lori R Arlinghaus; C Chad Quarles; Thomas E Yankeelov; E Brian Welch
Journal:  IEEE Trans Med Imaging       Date:  2011-10-14       Impact factor: 10.048

3.  On the Rician distribution of noisy MRI data.

Authors:  A H Andersen
Journal:  Magn Reson Med       Date:  1996-08       Impact factor: 4.668

4.  Respiration-based sorting of dynamic MRI to derive representative 4D-MRI for radiotherapy planning.

Authors:  Erik Tryggestad; Aaron Flammang; Sarah Han-Oh; Russell Hales; Joseph Herman; Todd McNutt; Teboh Roland; Steven M Shea; John Wong
Journal:  Med Phys       Date:  2013-05       Impact factor: 4.071

5.  4D respiratory motion-compensated image reconstruction of free-breathing radial MR data with very high undersampling.

Authors:  Christopher M Rank; Thorsten Heußer; Maria T A Buzan; Andreas Wetscherek; Martin T Freitag; Julien Dinkel; Marc Kachelrieß
Journal:  Magn Reson Med       Date:  2016-03-16       Impact factor: 4.668

6.  Accelerated Brain DCE-MRI Using Iterative Reconstruction With Total Generalized Variation Penalty for Quantitative Pharmacokinetic Analysis: A Feasibility Study.

Authors:  Chunhao Wang; Fang-Fang Yin; John P Kirkpatrick; Zheng Chang
Journal:  Technol Cancer Res Treat       Date:  2016-05-23

7.  4D XCAT phantom for multimodality imaging research.

Authors:  W P Segars; G Sturgeon; S Mendonca; Jason Grimes; B M W Tsui
Journal:  Med Phys       Date:  2010-09       Impact factor: 4.071

8.  The management of respiratory motion in radiation oncology report of AAPM Task Group 76.

Authors:  Paul J Keall; Gig S Mageras; James M Balter; Richard S Emery; Kenneth M Forster; Steve B Jiang; Jeffrey M Kapatoes; Daniel A Low; Martin J Murphy; Brad R Murray; Chester R Ramsey; Marcel B Van Herk; S Sastry Vedam; John W Wong; Ellen Yorke
Journal:  Med Phys       Date:  2006-10       Impact factor: 4.071

9.  A Technique for Generating Volumetric Cine-Magnetic Resonance Imaging.

Authors:  Wendy Harris; Lei Ren; Jing Cai; You Zhang; Zheng Chang; Fang-Fang Yin
Journal:  Int J Radiat Oncol Biol Phys       Date:  2016-02-06       Impact factor: 7.038

10.  Investigation of sagittal image acquisition for 4D-MRI with body area as respiratory surrogate.

Authors:  Yilin Liu; Fang-Fang Yin; Zheng Chang; Brian G Czito; Manisha Palta; Mustafa R Bashir; Yujiao Qin; Jing Cai
Journal:  Med Phys       Date:  2014-10       Impact factor: 4.071

View more
  5 in total

1.  A Novel method to generate on-board 4D MRI using prior 4D MRI and on-board kV projections from a conventional LINAC for target localization in liver SBRT.

Authors:  Wendy Harris; Chunhao Wang; Fang-Fang Yin; Jing Cai; Lei Ren
Journal:  Med Phys       Date:  2018-06-13       Impact factor: 4.071

2.  Predicting real-time 3D deformation field maps (DFM) based on volumetric cine MRI (VC-MRI) and artificial neural networks for on-board 4D target tracking: a feasibility study.

Authors:  Jonathan Pham; Wendy Harris; Wenzheng Sun; Zi Yang; Fang-Fang Yin; Lei Ren
Journal:  Phys Med Biol       Date:  2019-08-21       Impact factor: 3.609

Review 3.  Respiratory-Correlated (RC) vs. Time-Resolved (TR) Four-Dimensional Magnetic Resonance Imaging (4DMRI) for Radiotherapy of Thoracic and Abdominal Cancer.

Authors:  Guang Li; Yilin Liu; Xingyu Nie
Journal:  Front Oncol       Date:  2019-10-11       Impact factor: 6.244

4.  Evaluation of Multisource Adaptive MRI Fusion for Gross Tumor Volume Delineation of Hepatocellular Carcinoma.

Authors:  Andy Lai-Yin Cheung; Lei Zhang; Chenyang Liu; Tian Li; Anson Ho-Yin Cheung; Chun Leung; Angus Kwong-Chuen Leung; Sai-Kit Lam; Victor Ho-Fun Lee; Jing Cai
Journal:  Front Oncol       Date:  2022-02-25       Impact factor: 6.244

5.  Volumetric cine magnetic resonance imaging (VC-MRI) using motion modeling, free-form deformation and multi-slice undersampled 2D cine MRI reconstructed with spatio-temporal low-rank decomposition.

Authors:  Wendy Harris; Fang-Fang Yin; Jing Cai; Lei Ren
Journal:  Quant Imaging Med Surg       Date:  2020-02
  5 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.