Practical implementation of Channelized Hotelling Observers: Effect of ROI size.

Andrea Ferrero1, Christopher P Favazza1, Lifeng Yu1, Shuai Leng1, Cynthia H McCollough1.   

Abstract

Fundamental to the development and application of channelized Hotelling observer (CHO) models is the selection of the region of interest (ROI) to evaluate. For assessment of medical imaging systems, reducing the ROI size can be advantageous. Smaller ROIs enable a greater concentration of interrogable objects in a single phantom image, thereby providing more information from a set of images and reducing the overall image acquisition burden. Additionally, smaller ROIs may promote better assessment of clinical patient images as different patient anatomies present different ROI constraints. To this end, we investigated the minimum ROI size that does not compromise the performance of the CHO model. In this study, we evaluated both simulated images and phantom CT images to identify the minimum ROI size that resulted in an accurate figure of merit (FOM) of the CHO's performance. More specifically, the minimum ROI size was evaluated as a function of the following: number of channels, spatial frequency and number of rotations of the Gabor filters, size and contrast of the object, and magnitude of the image noise. Results demonstrate that a minimum ROI size exists below which the CHO's performance is grossly inaccurate. The minimum ROI size is shown to increase with number of channels and be dictated by truncation of lower frequency filters. We developed a model to estimate the minimum ROI size as a parameterized function of the number of orientations and spatial frequencies of the Gabor filters, providing a guide for investigators to appropriately select parameters for model observer studies.

Entities:  

Keywords:  Computed tomography (CT); channelized Hotelling observer (CHO); model observer; region of interest (ROI)

Year:  2017        PMID: 28943699      PMCID: PMC5607015          DOI: 10.1117/12.2255530

Source DB:  PubMed          Journal:  Proc SPIE Int Soc Opt Eng        ISSN: 0277-786X


  7 in total

1.  Automated computer evaluation and optimization of image compression of x-ray coronary angiograms for signal known exactly detection tasks.

Authors:  Miguel Eckstein; Jay Bartroff; Craig Abbey; James Whiting; Francois Bochud
Journal:  Opt Express       Date:  2003-03-10       Impact factor: 3.894

2.  Image covariance and lesion detectability in direct fan-beam x-ray computed tomography.

Authors:  Adam Wunderlich; Frédéric Noo
Journal:  Phys Med Biol       Date:  2008-04-18       Impact factor: 3.609

3.  Impact of number of repeated scans on model observer performance for a low-contrast detection task in computed tomography.

Authors:  Chi Ma; Lifeng Yu; Baiyu Chen; Christopher Favazza; Shuai Leng; Cynthia McCollough
Journal:  J Med Imaging (Bellingham)       Date:  2016-05-26

4.  Addition of a channel mechanism to the ideal-observer model.

Authors:  K J Myers; H H Barrett
Journal:  J Opt Soc Am A       Date:  1987-12       Impact factor: 2.129

5.  Model observers for assessment of image quality.

Authors:  H H Barrett; J Yao; J P Rolland; K J Myers
Journal:  Proc Natl Acad Sci U S A       Date:  1993-11-01       Impact factor: 11.205

6.  Correlation between human and model observer performance for discrimination task in CT.

Authors:  Yi Zhang; Shuai Leng; Lifeng Yu; Rickey E Carter; Cynthia H McCollough
Journal:  Phys Med Biol       Date:  2014-05-30       Impact factor: 3.609

7.  Prediction of human observer performance in a 2-alternative forced choice low-contrast detection task using channelized Hotelling observer: impact of radiation dose and reconstruction algorithms.

Authors:  Lifeng Yu; Shuai Leng; Lingyun Chen; James M Kofler; Rickey E Carter; Cynthia H McCollough
Journal:  Med Phys       Date:  2013-04       Impact factor: 4.071

  7 in total
  2 in total

1.  Use of a channelized Hotelling observer to assess CT image quality and optimize dose reduction for iteratively reconstructed images.

Authors:  Christopher P Favazza; Andrea Ferrero; Lifeng Yu; Shuai Leng; Kyle L McMillan; Cynthia H McCollough
Journal:  J Med Imaging (Bellingham)       Date:  2017-10-03

2.  Iterative reconstruction with multifrequency signal recognition technology to improve low-contrast detectability: A phantom study.

Authors:  Yoshinori Funama; Takashi Shirasaka; Taiga Goto; Yuko Aoki; Kana Tanaka; Ryo Yoshida
Journal:  Acta Radiol Open       Date:  2022-06-17
  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.