Wear testing of a canine hip resurfacing implant that uses highly cross-linked polyethylene.

Kevin J Warburton1, John B Everingham1, Jillian L Helms1, Andrew J Kazanovicz2, Katherine A Hollar1, Jeff D Brourman3, Steven M Fox2, Trevor J Lujan1.   

Abstract

Hip resurfacing offers advantages for young, active patients afflicted with hip osteoarthritis and may also be a beneficial treatment for adult canines. Conventional hip resurfacing uses metal-on-metal bearings to preserve bone stock, but it may be feasible to use metal-on-polyethylene bearings to reduce metal wear debris while still preserving bone. This study characterized the short-term wear behavior of a novel hip resurfacing implant for canines that uses a 1.5 mm thick liner of highly cross-linked polyethylene in the acetabular component. This implant was tested in an orbital bearing machine that simulated canine gait for 1.1 million cycles. Wear of the liner was evaluated using gravimetric analysis and by measuring wear depth with an optical scanner. The liners had a steady-state mass wear rate of 0.99 ± 0.17 mg per million cycles and an average wear depth in the central liner region of 0.028 mm. No liners, shells, or femoral heads had any catastrophic failure due to yielding or fracture. These results suggest that the thin liners will not prematurely crack after implantation in canines. This is the first hip resurfacing device developed for canines, and this study is the first to characterize the in vitro wear of highly cross-linked polyethylene liners in a hip resurfacing implant. The canine implant developed in this study may be an attractive treatment option for canines afflicted with hip osteoarthritis, and since canines are the preferred animal model for human hip replacement, this implant can support the development of metal-on-polyethylene hip resurfacing technology for human patients.
© 2017 Orthopaedic Research Society. Published by Wiley Periodicals, Inc. J Orthop Res 36:1196-1205, 2018. © 2017 Orthopaedic Research Society. Published by Wiley Periodicals, Inc.

Entities:  

Keywords:  cross-linked polyethylene; dog; gravimetrics; hip simulator; surface wear

Mesh:

Substances:

Year:  2017        PMID: 28941247      PMCID: PMC5866242          DOI: 10.1002/jor.23745

Source DB:  PubMed          Journal:  J Orthop Res        ISSN: 0736-0266            Impact factor:   3.494


  49 in total

1.  Analysis of temporal wear patterns of porous-coated acetabular components: distinguishing between true wear and so-called bedding-in.

Authors:  C J Sychterz; C A Engh; A Yang; C A Engh
Journal:  J Bone Joint Surg Am       Date:  1999-06       Impact factor: 5.284

2.  The effect of component size and orientation on the concentrations of metal ions after resurfacing arthroplasty of the hip.

Authors:  D J Langton; S S Jameson; T J Joyce; J Webb; A V F Nargol
Journal:  J Bone Joint Surg Br       Date:  2008-09

3.  Highly cross-linked polyethylene in hip resurfacing arthroplasty: long-term follow-up.

Authors:  Harlan C Amstutz; Karren M Takamura; Edward Ebramzadeh; Michel J Le Duff
Journal:  Hip Int       Date:  2014-10-17       Impact factor: 2.135

4.  Birmingham hip resurfacing at a mean of ten years: results from an independent centre.

Authors:  G Coulter; D A Young; R E Dalziel; A J Shimmin
Journal:  J Bone Joint Surg Br       Date:  2012-03

5.  Evaluation of a three-dimensional kinematic model for canine gait analysis.

Authors:  Yang-Chieh Fu; Bryan T Torres; Steven C Budsberg
Journal:  Am J Vet Res       Date:  2010-10       Impact factor: 1.156

6.  Early outcomes of sequentially cross-linked thin polyethylene liners with large diameter femoral heads in total hip arthroplasty.

Authors:  Siraj A Sayeed; Michael A Mont; Christopher R Costa; Aaron J Johnson; Qais Naziri; Peter M Bonutti; Ronald E Delanois
Journal:  Bull NYU Hosp Jt Dis       Date:  2011

7.  Wear behaviour of cross-linked polyethylene assessed in vitro under severe conditions.

Authors:  Saverio Affatato; Gianluca Bersaglia; Mirko Rocchi; Paola Taddei; Concezio Fagnano; Aldo Toni
Journal:  Biomaterials       Date:  2005-06       Impact factor: 12.479

8.  Wear resistance of the biocompatible phospholipid polymer-grafted highly cross-linked polyethylene liner against larger femoral head.

Authors:  Toru Moro; Yoshio Takatori; Masayuki Kyomoto; Kazuhiko Ishihara; Hiroshi Kawaguchi; Masami Hashimoto; Takeyuki Tanaka; Hirofumi Oshima; Sakae Tanaka
Journal:  J Orthop Res       Date:  2015-04-06       Impact factor: 3.494

9.  Long-term outcome of a metal-on-polyethylene cementless hip resurfacing.

Authors:  Timothy L Tan; Edward Ebramzadeh; Patricia A Campbell; Mariam Al-Hamad; Harlan C Amstutz
Journal:  J Arthroplasty       Date:  2013-10-03       Impact factor: 4.757

10.  Wear performance of neat and vitamin E blended highly cross-linked PE under severe conditions: The combined effect of accelerated ageing and third body particles during wear test.

Authors:  Saverio Affatato; Jonathan Salvatore De Mattia; Pierangiola Bracco; Eleonora Pavoni; Paola Taddei
Journal:  J Mech Behav Biomed Mater       Date:  2016-08-05
View more
  1 in total

1.  Quantifying wear depth in hip prostheses using a 3D optical scanner.

Authors:  Katherine A Hollar; Daniel S Ferguson; John B Everingham; Jillian L Helms; Kevin J Warburton; Trevor J Lujan
Journal:  Wear       Date:  2017-10-16       Impact factor: 3.892

  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.