Effectiveness of a New Lead-Shielding Device and Additional Filter for Reducing Staff and Patient Radiation Exposure During Videofluoroscopic Swallowing Study Using a Human Phantom.

Yoshiaki Morishima1,2, Koichi Chida3, Yoshikazu Muroya4, Yoshiya Utsumi4.   

Abstract

Interventional radiology procedures often involve lengthy exposure to fluoroscopy-derived radiation. We therefore devised a videofluoroscopic swallowing study (VFSS) procedure using a human phantom that proved to protect the patient and physician by reducing the radiation dose. We evaluated a new lead-shielding device and separately attached additional filters (1.0-, 2.0-, and 3.0-mm Al filters and a 0.5-mm Cu filter) during VFSS to reduce the patient's entrance skin dose (ESD). A monitor attached to the human phantom's neck measured the ESD. We also developed another lead shield (VFSS Shielding Box, 1.0-mm Pb equivalent) and tested its efficacy using the human phantom and an ionization chamber radiation survey meter with and without protection from scattered radiation at the physician's position on the phantom. We then measured the scattered radiation (at 90 and 150 cm above the floor) after combining the filters with the VFSS Shielding Box. With the additional filters, the ESD was reduced by 15.4-55.1%. With the VFSS Shielding Box alone, the scattered radiation was reduced by about 10% compared with the dose without additional shielding. With the VFSS Shielding Box and filters combined, the scattered radiation dose was reduced by a maximum of about 44% at the physician's position. Thus, the additional lead-shielding device effectively provided protection from scattered radiation during fluoroscopy. These results indicate that the combined VFSS Shielding Box and filters can effectively reduce the physician's and patient's radiation doses.

Entities:  

Keywords:  Additional filter; Deglutition; Deglutition disorders; Fluoroscopy; Original lead shielding; Scattered radiation protection; Video recording

Mesh:

Year:  2017        PMID: 28921404     DOI: 10.1007/s00455-017-9839-6

Source DB:  PubMed          Journal:  Dysphagia        ISSN: 0179-051X            Impact factor:   3.438


  32 in total

1.  Clinical utility of the modified barium swallow.

Authors:  B Martin-Harris; J A Logemann; S McMahon; M Schleicher; J Sandidge
Journal:  Dysphagia       Date:  2000       Impact factor: 3.438

2.  Comparison of image quality and radiation dose between an image-intensifier system and a newer-generation flat-panel detector system — technical phantom measurements and evaluation of clinical imaging in children.

Authors:  Meike Weis; Claudia Hagelstein; Theo Diehm; Stefan O Schoenberg; K Wolfgang Neff
Journal:  Pediatr Radiol       Date:  2016-02

3.  Radiation dose of interventional radiology system using a flat-panel detector.

Authors:  Koichi Chida; Yohei Inaba; Haruo Saito; Tadashi Ishibashi; Shoki Takahashi; Masahiro Kohzuki; Masayuki Zuguchi
Journal:  AJR Am J Roentgenol       Date:  2009-12       Impact factor: 3.959

4.  ICRP Publication 113. Education and training in radiological protection for diagnostic and interventional procedures.

Authors:  E Vañó; M Rosenstein; J Liniecki; M M Rehani; C J Martin; R J Vetter
Journal:  Ann ICRP       Date:  2011-04-15

5.  Patients' radiation dose during videofluoroscopic swallowing studies according to underlying characteristics.

Authors:  Hong Min Kim; Kyoung Hyo Choi; Tae Woo Kim
Journal:  Dysphagia       Date:  2012-09-09       Impact factor: 3.438

6.  Influence of the target vessel on the location and area of maximum skin dose during percutaneous coronary intervention.

Authors:  K Chida; K Fuda; Y Kagaya; H Saito; Y Takai; M Kohzuki; S Takahashi; S Yamada; M Zuguchi
Journal:  Acta Radiol       Date:  2007-10       Impact factor: 1.990

Review 7.  Fluoroscopically guided interventional procedures: a review of radiation effects on patients' skin and hair.

Authors:  Stephen Balter; John W Hopewell; Donald L Miller; Louis K Wagner; Michael J Zelefsky
Journal:  Radiology       Date:  2010-02       Impact factor: 11.105

8.  Patient dose during videofluoroscopy swallowing studies in a Hong Kong public hospital.

Authors:  Koon Ho Torrance Chau; Chak Man Andy Kung
Journal:  Dysphagia       Date:  2009-04-24       Impact factor: 3.438

9.  Occupational Radiation Exposure during Endoscopic Retrograde Cholangiopancreatography and Usefulness of Radiation Protective Curtains.

Authors:  Tomoyuki Minami; Tamito Sasaki; Masahiro Serikawa; Michihiro Kamigaki; Masanobu Yukutake; Takashi Ishigaki; Yasutaka Ishii; Teruo Mouri; Satoshi Yoshimi; Akinori Shimizu; Tomofumi Tsuboi; Keisuke Kurihara; Yumiko Tatsukawa; Eisuke Miyaki; Kazuaki Chayama
Journal:  Gastroenterol Res Pract       Date:  2014-11-13       Impact factor: 2.260

10.  Radiation Dose Reduction during Radial Cardiac Catheterization: Evaluation of a Dedicated Radial Angiography Absorption Shielding Drape.

Authors:  Andrew Ertel; Jeffrey Nadelson; Adhir R Shroff; Ranya Sweis; Dean Ferrera; Mladen I Vidovich
Journal:  ISRN Cardiol       Date:  2012-09-04
View more
  5 in total

Review 1.  What are useful methods to reduce occupational radiation exposure among radiological medical workers, especially for interventional radiology personnel?

Authors:  Koichi Chida
Journal:  Radiol Phys Technol       Date:  2022-05-24

2.  New Radioprotective Device that can be Used for Fluoroscopic Exam: Possibility to Contribute to Staff Exposure Protection During VFSS.

Authors:  Yoshiaki Morishima; Koichi Chida; Osamu Ito
Journal:  Dysphagia       Date:  2022-02-15       Impact factor: 3.438

3.  Usefulness of an Additional Filter Created Using 3D Printing for Whole-Body X-ray Imaging with a Long-Length Detector.

Authors:  Hyunsoo Seo; Wooyoung Kim; Bongju Han; Huimin Jang; Myeong Seong Yoon; Youngjin Lee
Journal:  Sensors (Basel)       Date:  2022-06-06       Impact factor: 3.847

Review 4.  Radiation Exposure to Staff and Patient During Videofluoroscopic Swallowing Studies and Recommended Protection Strategies.

Authors:  Victoria Jean Earl; Mohamed Khaldoun Badawy
Journal:  Dysphagia       Date:  2018-09-22       Impact factor: 3.438

5.  Cardiac catheterization real-time dynamic radiation dose measurement to estimate lifetime attributable risk of cancer.

Authors:  Chun-Yuan Tu; Chung-Jung Lin; Bang-Hung Yang; Jay Wu; Tung-Hsin Wu
Journal:  PLoS One       Date:  2020-06-16       Impact factor: 3.240

  5 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.