Literature DB >> 28862631

Strategies for high-throughput focused-beam ptychography.

Chris Jacobsen1, Junjing Deng1, Youssef Nashed2.   

Abstract

X-ray ptychography is being utilized for a wide range of imaging experiments with a resolution beyond the limit of the X-ray optics used. Introducing a parameter for the ptychographic resolution gain Gp (the ratio of the beam size over the achieved pixel size in the reconstructed image), strategies for data sampling and for increasing imaging throughput when the specimen is at the focus of an X-ray beam are considered. The tradeoffs between large and small illumination spots are examined.

Entities:  

Keywords:  high throughput; ptychographic resolution gain; ptychography

Year:  2017        PMID: 28862631      PMCID: PMC5580791          DOI: 10.1107/S1600577517009869

Source DB:  PubMed          Journal:  J Synchrotron Radiat        ISSN: 0909-0495            Impact factor:   2.616


Introduction

Ptychography (Hoppe, 1969 ▸) involves the use of overlapping coherent illumination regions on a specimen and the collection of diffraction data from each illumination spot, followed by reconstruction of an image. Following initial experimental demonstrations of ptychography (Rodenburg et al., 2007 ▸; Thibault et al., 2008 ▸) and related methods (Chapman, 1996 ▸), X-ray ptychography is finding increased utilization in X-ray microscopy because it can be used to deliver amplitude and phase-contrast images beyond the resolution limit of the coherent beam size. It does so without the small isolated specimen limitations of X-ray coherent diffraction imaging (Miao et al., 1999 ▸) which are intrinsic to the use of finite support iterative phase retrieval (Fienup, 1978 ▸) unless one uses a spatially restricted coherent beam to satisfy the support constraint (Abbey et al., 2008 ▸) (which begins to look like ptychography if one scans the beam).

Discussion

Because ptychography requires mostly coherent beams, and involves point-by-point scanning, it is usually regarded as a low-throughput imaging method even though several higher throughput examples exist (Guizar-Sicairos et al., 2014 ▸; Holler et al., 2014 ▸). We consider here the factors that can be used to increase that throughput, characterizing them in terms of the ptychographic resolution gain which we define as the ratio between the desired reconstructed image pixel size1 δ and the diameter d of the coherent beam spot that is scanned across the specimen, orIn order to have the possibility of a square half-period pixel size of δ using an illumination wavelength λ, Fig. 1 ▸ shows that one must record first-order diffraction out to an angle (in the small-angle approximation of ) ofThis requires coherent superposition between waves with a path length difference from the top and bottom edges of the illumination spot to a distant detector of , or a number m of wavelengths of longitudinal coherence (Spence et al., 2004 ▸; van der Veen & Pfeiffer, 2004 ▸; Enders et al., 2014 ▸) ofwhere we have again used the small-angle approximation. Therefore the spectral bandwidth should beIllumination from the coherent illumination spot of diameter d will diffract out by a semi-angle2 of = to cover a diameter D on the detector ofThe signal from small features within the illumination spot is recorded on a detector with N pixels (each of width Δ) on a side, so that it subtends a semi-angle ofThe illumination spot will therefore be spread out over a number of detector pixels ofwhere the final expression uses equations (1), (2) and (5). Finally, because the last two pixels on the detector must be able to record fringes caused by interference from points at the top and bottom of the illumination spot d with Nyquist sampling, we must have 2m pixels from the detector center to the edge or 4m pixels overall in each dimension. In other words, we find that the minimum number of detector pixels is given bywhere we have used equation (3). We note that ptychography can be performed with reduced detector sampling, though at a cost of image fidelity (Edo et al., 2013 ▸) or of finer real-space sampling (Batey et al., 2014 ▸; da Silva & Menzel, 2015 ▸) which effectively corresponds to larger values of the overlap factor o discussed below.
Figure 1

Data sampling in ptychography. The specimen is illuminated by beam spots of diameter d with divergence , leading to an incident illumination diameter D on the detector. The desired half-period pixel width δ (the limit of resolution) implies a diffraction numerical aperture out to the edges of the detector.

Achieving high resolution in ptychography requires the detection of scattering out to large angles, which is determined in part by the specimen’s optical properties at the chosen X-ray wavelength and the number of photons used to illuminate each pixel (Glaeser, 1971 ▸; Sayre et al., 1977 ▸; Schropp & Schroer, 2010 ▸). While high resolution can also be aided by having high-spatial-frequency content in the illuminating beam if a larger focal spot (larger value of ) is used (Guizar-Sicairos et al., 2012 ▸), we assume that the dominating factor is the required photon density per area, = , independent of the ptychographic spatial resolution gain . If a is the center-to-center position increment between illumination spots of diameter d, it is recommended to use an overlap factor = with ≃ 0.6 for robust reconstructions (Bunk et al., 2008 ▸) [this parameter has also been used in connection with continuous scan ptychography (Deng et al., 2015a ▸)]. Imaging of a square area A then requires the use of a number of illumination spots given byif one ignores incomplete illumination at the edges. Imaging the area A then requires a net photon count ofwhere we have used equation (9) to arrive at the final result. If the source delivers a coherent flux of photons per second, the exposure time for each of the illumination spots ( equals the transit time per distance a in the case of continuous scanning) can be found from = , leading to an exposure time per illumination spot ofBecause the coherent flux is an intrinsic property of the source [with the caveat that one can obtain higher coherent flux from a broad-band source as indicated by equation (4)], it is unchanged by the size d of the spot into which the coherent flux is delivered; that is, does not depend on the ptychographic spatial resolution gain . From the above, we see that decreasing the ptychographic spatial resolution gain (that is, using smaller coherent illumination spots d) has several implications: (i) Decreasing means detectors with fewer pixels can be used, as can be seen directly from equation (8). (ii) Decreasing leads to relaxed requirements for the monochromaticity in the illumination, as can be seen from equation (4). When using large-bandwidth X-ray sources, this could allow one to use multilayer monochromators with bandpass ≃ as compared with crystal monochromators with bandpass ≃ , thus leading to a usable flux gain increase of . This has already been demonstrated as a route to increased throughput in ptychography (Enders et al., 2014 ▸). (iii) Decreasing the gain has the effect of increasing the fractional number of detector pixels within which the incident beam is recorded, as can be seen from equation (7) divided by . Since the incident beam is often much stronger than the fraction of the beam scattered by the sample, putting the incident beam into a larger fraction of detector pixels reduces demands on the dynamic range of the detector. This is in contrast to the case of far-field coherent diffraction imaging of weakly scattering objects like biological cells, where the dynamic range required of the detector can be in excess of :1 so that multiple detector exposures must be acquired with various exposure times and direct-beam absorber positions (Shapiro et al., 2005 ▸). (iv) When smaller illumination spots d are used with smaller ptychographic gain , one has the option of simultaneously acquiring high-spatial-resolution scanning microscope images using other contrast modes, such as fluorescence from elemental content (Schropp et al., 2012 ▸; Deng et al., 2015b ▸, 2017 ▸). (v) Decreasing is associated with shorter exposure times per diffraction pattern recording, as can be see from equation (11). This means that higher detector frame rates are required if one uses smaller values of . However, the net ‘information rate’ to be transferred by the detector of total pixels per second, or , is not affected as can be seen by the fact that both [from the square of equation (8)] and the exposure time [from equation (11)] depend on , thus canceling out any dependence on . In other words, the total amount of data to be saved or transferred is the same in optimized experiments, independent of the ptychographic gain . We have not accounted for factors such as thermal noise or amplifier readout noise which might be present in charge-integrating detectors versus photon-counting detectors. (vi) Finally, if there is an overhead time associated with the collection of signal from each of the illumination spots, reducing and thus increasing [equation (9)] will lead to longer scan times. However, continuous-scan methods largely remove this overhead time between pixels (Pelz et al., 2014 ▸; Deng et al., 2015a ▸); in raster scans, the overhead between scan lines depends both on the step distance (favoring small ) and data transfer overheads (favoring large ). These advantages of using small illumination spots or small values of ptychographic spatial resolution gain can be compelling for certain applications. Finally, we note that increased experimental throughput should be coupled with fast computing; one example is to use efficient parallelization schemes for ptychographic image reconstruction (Nashed et al., 2014 ▸).
  15 in total

1.  Elementary signals in ptychography.

Authors:  Julio Cesar da Silva; Andreas Menzel
Journal:  Opt Express       Date:  2015-12-28       Impact factor: 3.894

2.  Hard-x-ray lensless imaging of extended objects.

Authors:  J M Rodenburg; A C Hurst; A G Cullis; B R Dobson; F Pfeiffer; O Bunk; C David; K Jefimovs; I Johnson
Journal:  Phys Rev Lett       Date:  2007-01-18       Impact factor: 9.161

3.  High-resolution scanning x-ray diffraction microscopy.

Authors:  Pierre Thibault; Martin Dierolf; Andreas Menzel; Oliver Bunk; Christian David; Franz Pfeiffer
Journal:  Science       Date:  2008-07-18       Impact factor: 47.728

4.  Reconstruction of an object from the modulus of its Fourier transform.

Authors:  J R Fienup
Journal:  Opt Lett       Date:  1978-07-01       Impact factor: 3.776

5.  Transmission microscropy of unmodified biological materials: comparative radiation dosages with electrons and ultrasoft x-ray photons.

Authors:  D Sayre; J Kirz; R Feder; D M Kim; E Spiller
Journal:  Ultramicroscopy       Date:  1977-08       Impact factor: 2.689

6.  Parallel ptychographic reconstruction.

Authors:  Youssef S G Nashed; David J Vine; Tom Peterka; Junjing Deng; Rob Ross; Chris Jacobsen
Journal:  Opt Express       Date:  2014-12-29       Impact factor: 3.894

7.  Simultaneous cryo X-ray ptychographic and fluorescence microscopy of green algae.

Authors:  Junjing Deng; David J Vine; Si Chen; Youssef S G Nashed; Qiaoling Jin; Nicholas W Phillips; Tom Peterka; Rob Ross; Stefan Vogt; Chris J Jacobsen
Journal:  Proc Natl Acad Sci U S A       Date:  2015-02-09       Impact factor: 11.205

8.  Continuous motion scan ptychography: characterization for increased speed in coherent x-ray imaging.

Authors:  Junjing Deng; Youssef S G Nashed; Si Chen; Nicholas W Phillips; Tom Peterka; Rob Ross; Stefan Vogt; Chris Jacobsen; David J Vine
Journal:  Opt Express       Date:  2015-03-09       Impact factor: 3.894

9.  High-throughput ptychography using Eiger: scanning X-ray nano-imaging of extended regions.

Authors:  Manuel Guizar-Sicairos; Ian Johnson; Ana Diaz; Mirko Holler; Petri Karvinen; Hans-Christian Stadler; Roberto Dinapoli; Oliver Bunk; Andreas Menzel
Journal:  Opt Express       Date:  2014-06-16       Impact factor: 3.894

10.  Limitations to significant information in biological electron microscopy as a result of radiation damage.

Authors:  R M Glaeser
Journal:  J Ultrastruct Res       Date:  1971-08
View more
  4 in total

1.  3D X-Ray Imaging of Continuous Objects beyond the Depth of Focus Limit.

Authors:  M A Gilles; Y S G Nashed; M DU; C Jacobsen; S M Wild
Journal:  Optica       Date:  2018-09-20       Impact factor: 11.104

2.  A comparison of absorption and phase contrast for X-ray imaging of biological cells.

Authors:  Colin Nave
Journal:  J Synchrotron Radiat       Date:  2018-08-27       Impact factor: 2.616

3.  Fast digital lossy compression for X-ray ptychographic data.

Authors:  Panpan Huang; Ming Du; Mike Hammer; Antonino Miceli; Chris Jacobsen
Journal:  J Synchrotron Radiat       Date:  2021-01-01       Impact factor: 2.616

Review 4.  Upscaling X-ray nanoimaging to macroscopic specimens.

Authors:  Ming Du; Zichao Wendy Di; Doǧa Gürsoy; R Patrick Xian; Yevgenia Kozorovitskiy; Chris Jacobsen
Journal:  J Appl Crystallogr       Date:  2021-02-19       Impact factor: 4.868

  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.