PURPOSE: The aim of the study was to determine the 5 year migratory and wear patterns, adverse events and clinical outcomes of a cementless, porous acetabular cup.
METHODS: RSA imaging of a cohort of 11 patients was retrospective analysed at 5 years post Total Hip Arthroplasty (THA). Changes in pain, function and symptoms of the hip at 5 years post-THA were compared to preoperative and 2 year postoperative assessments on the Harris Hip Score (HHS) and Hip dysfunction and Osteoarthritis Outcome Score (HOOS).
RESULTS: The majority of cup migration occurred up to 6 months and stabilised thereafter (6 months to 5 years, p = 0.091-0.866, Wilcoxon Signed Rank test). The direction of rotation around the 3 axes was evenly distributed among the cups between anterior-posterior rotation, internal-external rotation and increased-decreased inclination. The majority of the cups translated proximally, at an average migration of 0.36 mm (±95%CI 0.17) at 5-years post-THA. Following initial bedding in, up to 6 months, there was no detectable polyethylene wear between 6 months and 5 years. At 5 years postoperatively, a statistically significant difference was observed across all HOOS subscales in comparison to preoperative values, with higher means reported at 5 years (p < 0.001). The total mean HHS displayed a statistically significant improvement, increasing from 'poor' preoperatively to 'good' at 5 years post-THA.
CONCLUSION: Following initial migration between discharge and 6 months, the cementless porous acetabular cup demonstrated a tendency for stabilisation from 6 months up to 5 years post-THA, suggesting good mid-term fixation. Additionally, improvements in clinical outcome measures of pain, function and quality-of-life remained high following THA at 5 years post-surgery.

Entities:  

Keywords:  Acetabular cup; Migration; RSA; Radiostereometric analysis; THA; Total Hip Arthroplasty

Year:  2017        PMID: 28819343      PMCID: PMC5548368          DOI: 10.1016/j.jor.2017.07.004

Source DB:  PubMed          Journal:  J Orthop        ISSN: 0972-978X


  14 in total

1.  The 6-year migration characteristics of a hydroxyapatite-coated femoral stem: a radiostereometric analysis study.

Authors:  Stuart A Callary; David G Campbell; Graham E Mercer; Kjell G Nilsson; John R Field
Journal:  J Arthroplasty       Date:  2012-01-21       Impact factor: 4.757

Review 2.  Roentgen stereophotogrammetry. A method for the study of the kinematics of the skeletal system.

Authors:  G Selvik
Journal:  Acta Orthop Scand Suppl       Date:  1989

3.  Radiostereometry of hip prostheses. Review of methodology and clinical results.

Authors:  J Kärrholm; P Herberts; P Hultmark; H Malchau; B Nivbrant; J Thanner
Journal:  Clin Orthop Relat Res       Date:  1997-11       Impact factor: 4.176

4.  Porous titanium construct cup compared to porous coated titanium cup in total hip arthroplasty. A randomised controlled trial.

Authors:  Mats Salemyr; Olle Muren; Thomas Eisler; Henrik Bodén; Ghazi Chammout; André Stark; Olof Sköldenberg
Journal:  Int Orthop       Date:  2014-10-22       Impact factor: 3.075

5.  Traumatic arthritis of the hip after dislocation and acetabular fractures: treatment by mold arthroplasty. An end-result study using a new method of result evaluation.

Authors:  W H Harris
Journal:  J Bone Joint Surg Am       Date:  1969-06       Impact factor: 5.284

6.  Guidelines for standardization of radiostereometry (RSA) of implants.

Authors:  Edward R Valstar; Richie Gill; Leif Ryd; Gunnar Flivik; Niclas Börlin; Johan Kärrholm
Journal:  Acta Orthop       Date:  2005-08       Impact factor: 3.717

Review 7.  Measures of hip function and symptoms: Harris Hip Score (HHS), Hip Disability and Osteoarthritis Outcome Score (HOOS), Oxford Hip Score (OHS), Lequesne Index of Severity for Osteoarthritis of the Hip (LISOH), and American Academy of Orthopedic Surgeons (AAOS) Hip and Knee Questionnaire.

Authors:  Anna Nilsdotter; Ann Bremander
Journal:  Arthritis Care Res (Hoboken)       Date:  2011-11       Impact factor: 4.794

8.  Trabecular metal tibial knee component still stable at 10 years.

Authors:  Anders Henricson; Kjell G Nilsson
Journal:  Acta Orthop       Date:  2016-06-30       Impact factor: 3.717

Review 9.  Early proximal migration of cups is associated with late revision in THA: a systematic review and meta-analysis of 26 RSA studies and 49 survivalstudies.

Authors:  Bart G Pijls; Marc J Nieuwenhuijse; Marta Fiocco; Josepha Wm Plevier; Saskia Middeldorp; Rob Ghh Nelissen; Edward R Valstar
Journal:  Acta Orthop       Date:  2012-11-05       Impact factor: 3.717

10.  Hip disability and osteoarthritis outcome score (HOOS)--validity and responsiveness in total hip replacement.

Authors:  Anna K Nilsdotter; L Stefan Lohmander; Maria Klässbo; Ewa M Roos
Journal:  BMC Musculoskelet Disord       Date:  2003-05-30       Impact factor: 2.362

View more
  5 in total

1.  Minimum ten-year follow-up of a randomized trial comparing acetabular component fixation of two porous in-growth surfaces using radiosteriometric analysis.

Authors:  Luthfur Rahman; Mazin S Ibrahim; Lyndsay Somerville; Matthew G Teeter; Douglas D Naudie; Richard W McCalden
Journal:  Bone Jt Open       Date:  2020-10-20

2.  Higher early proximal migration of hemispherical cups with electrochemically applied hydroxyapatite (BoneMaster) on a porous surface compared with porous surface alone: a randomized RSA study with 53 patients.

Authors:  Peter Bo Jørgensen; Henrik Daugaard; Stig Storgaard Jakobsen; Martin Lamm; Kjeld Søballe; Maiken Stilling
Journal:  Acta Orthop       Date:  2019-11-08       Impact factor: 3.717

3.  In vitro and in vivo evaluations of the fully porous Ti6Al4V acetabular cups fabricated by a sintering technique.

Authors:  Ji Li; Wei Li; Zhongli Li; Yuxing Wang; Ruiling Li; Jiangping Tu; Gong Jin
Journal:  RSC Adv       Date:  2019-02-26       Impact factor: 4.036

4.  Larger 5-year migration but similar polyethylene wear of cementless hemispherical cups with electrochemically applied hydroxyapatite (BoneMaster) coating compared with porous plasma-spray titanium: a randomized 5-year RSA study.

Authors:  Sebastian Breddam Mosegaard; Peter Bo Jørgensen; Stig Storgaard Jakobsen; Henrik Daugaard; Kjeld Søballe; Maiken Stilling
Journal:  Acta Orthop       Date:  2022-07-14       Impact factor: 3.925

5.  Model-based roentgen stereophotogrammetric analysis using elementary geometrical shape models: 10 years results of an uncemented acetabular cup component.

Authors:  Anne Jacobsen; Frank Seehaus; Yutong Hong; Han Cao; Alexander Schuh; Raimund Forst; Stefan Sesselmann
Journal:  BMC Musculoskelet Disord       Date:  2018-09-18       Impact factor: 2.362

  5 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.