Literature DB >> 22266049

The 6-year migration characteristics of a hydroxyapatite-coated femoral stem: a radiostereometric analysis study.

Stuart A Callary1, David G Campbell, Graham E Mercer, Kjell G Nilsson, John R Field.   

Abstract

A prospective cohort of 30 patients undergoing primary total hip arthroplasty for treatment of osteoarthritis was enrolled in a study to characterize the migration behavior of a clinically successful cementless stem. At 6 years, the mean subsidence of the stem was 0.63 mm (range, -0.33 to 3.68 mm); the mean rotation into retroversion was 1.41° (range, -1.33° to 7.48°). No stems had additional subsidence of more than 0.25 mm between 6 months and 6 years. The resultant mean subsidence between 2 and 6 years was 0.03 mm, which is below the limit measurable by radiostereometric analysis. The data demonstrate that subsidence of this cementless stem occurs within the first 6 months, after which there is persistent stabilization.
Copyright © 2012 Elsevier Inc. All rights reserved.

Entities:  

Mesh:

Substances:

Year:  2012        PMID: 22266049     DOI: 10.1016/j.arth.2011.12.002

Source DB:  PubMed          Journal:  J Arthroplasty        ISSN: 0883-5403            Impact factor:   4.757


  6 in total

1.  Analysis of migration of the Nanos® short-stem hip implant within two years after surgery.

Authors:  Stefan Budde; Frank Seehaus; Michael Schwarze; Christof Hurschler; Thilo Floerkemeier; Henning Windhagen; Yvonne Noll; Max Ettinger; Fritz Thorey
Journal:  Int Orthop       Date:  2015-09-25       Impact factor: 3.075

2.  Is there any difference between tapered titanium stems with similar geometry and hydroxyapatite coating?

Authors:  G Cinotti; G Mazzotta; F Romana Ripani; G La Torre; G Giannicola
Journal:  Musculoskelet Surg       Date:  2019-01-08

3.  Mid-term migration of a cementless, porous acetabular cup: A 5 year Radiostereometric analysis.

Authors:  Kjell G Nilsson; Annika Theodoulou; Graham Mercer; Stephen J Quinn; Jeganath Krishnan
Journal:  J Orthop       Date:  2017-07-21

4.  Trabecular metal tibial knee component still stable at 10 years.

Authors:  Anders Henricson; Kjell G Nilsson
Journal:  Acta Orthop       Date:  2016-06-30       Impact factor: 3.717

5.  Good stability of a cementless, anatomically designed femoral stem in aging women: a 9-year RSA study of 32 patients.

Authors:  Erik Aro; Jessica J Alm; Niko Moritz; Kimmo Mattila; Hannu T Aro
Journal:  Acta Orthop       Date:  2018-07-10       Impact factor: 3.717

6.  The anatomical SP-CL stem demonstrates a non-progressing migration pattern in the first year: a low dose CT-based migration study in 20 patients.

Authors:  Olof Sandberg; Simon Tholén; Sofia Carlsson; Per Wretenberg
Journal:  Acta Orthop       Date:  2020-10-16       Impact factor: 3.717

  6 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.