Matthew C Wallace1, SSgt Tyler Britton2, Robbie Meek2, Sharon Walsh-Hart3, Col Todd E Carter1, Steven J Lisco4.
Abstract
BACKGROUND: The critically ill or injured patient undergoing military medical evacuation may require emergent intubation. Intubation may be life-saving, but it carries risks. The novice or infrequent laryngoscopist has a distinct disadvantage because experience is critical for the rapid and safe establishment of a secured airway. This challenge is compounded by the austere environment of the back of an aircraft under blackout conditions. This study determined which of five different video-assisted intubation devices (VAIDs) was best suited for in-flight use by U.S. Air Force Critical Care Air Transport Teams by comparing time to successful intubation between novice and expert laryngoscopists under three conditions, Normal Airway Lights on (NAL), Difficult Airway Lights on (DAL) and Difficult Airway Blackout (DAB), using manikins on a standard military transport stanchion and the floor with a minimal amount of setup time and extraneous light emission.Entities:
Keywords: Difficult airway; Expert; Laryngoscopy; Light emission; Military; Novice; Simulation; Video
Mesh:
Year: 2017 PMID: 28630743 PMCID: PMC5471909 DOI: 10.1186/s40779-017-0129-2
Source DB: PubMed Journal: Mil Med Res ISSN: 2054-9369
Tested video-assisted intubation devices (VAIDs)
| Device name | Manufacturer | Manufacturer location |
|---|---|---|
| Airtraq | Prodol Meditec, S.A. | Vizcaya, Spain |
| AWS | Pentax Medical Company | Montvale, NJ, USA |
| C-MAC | Karl Storz Endoskope GmbH & Co, KG | Tuttlingen, Germany |
| Coopdech VLP-100 | Daiken Medical Co, Ltd | Osaka, Japan |
| GlideScope Ranger | Verathon, Inc. | Bothell, WA, USA |
Fig. 1View of the simulation laboratory under full light conditions
Fig. 2View of the simulation laboratory under simulated blackout conditions
Fig. 3Diagram of the simulation laboratory as used for measuring device light output (scale is approximate)
Participants’ meantimes to successful intubation (seconds(SD) [medians])
| Condition | Name of devices | K-W | ||||
|---|---|---|---|---|---|---|
| Airtraq | AWS | C-MAC | Coopdech | GlideScope | ||
| Novice group | ||||||
| NAL Stanchion | 14.1 (7.5)[12.5] | 15.0 (13.1)[15.0] | 17.9 (29.8)[10.4] | 14.0 (9.9)[10.9] | 17.1 (11.6)[14.1] | 0.790 |
| NAL Ground | 9.3 (4.6)[9.0] | 13.4 (13.5)[9.2] | 31.6 (33.0)[16.4] | 20.1 (34.7)[11.8] | 17.9 (13.6)[13.1] | 0.002 |
| DAL Stanchion | 13.6 (7.6)[11.9] | 10.8 (6.4)[8.9] | 23.8 (30.1)[11.2] | 19.8 (27.9)[1.1] | 22.6 (21.6)[15.1] | 0.010 |
| DAL Ground | 42.8 (65.3)[14.5] | 17.9 (12.4)[16.0] | 48.4 (36.3)[39.5] | 47.5 (37.4)[35.1] | 34.3 (39.4)[22.0] | <0.001 |
| DAB Stanchion | 14.3 (7.6)[9.7] | 8.5 (5.2)[6.9] | 16.9 (11.3)[14.1] | 24.3 (58.1)[10.4] | 18.4 (16.4)[12.2] | <0.001 |
| DAB Ground | 23.0 (48.8)[11.2] | 10.8 (6.7)[8.7] | 31.0 (27.7)[25.4] | 19.0 (18.8)[14.3] | 19.8 (10.4)[17.4] | <0.001 |
| Expert group | ||||||
| NAL stanchion | 15.6 (10.5)[13.8] | 20.5 (21.0)[13.9] | 19.0 (46.3)[9.8] | 23.2 (18.8)[15.9] | 13.0 (9.1)[10.9] | 0.410 |
| NAL ground | 10.8 (6.3)[9.6] | 12.0 (10.6)[8.4] | 12.9 (7.8)[11.1] | 14.0 (9.1)[10.8] | 9.7 (3.3)[9.7] | 0.760 |
| DAL stanchion | 11.9 (5.6)[9.8] | 12.2 (9.7)[9.8] | 18.7 (14.5)[12.6] | 14.7 (8.5)[11.5] | 15.8 (9.3)[14.0] | 0.350 |
| DAL ground | 17.6 (7.6)[18.0] | 30.1 (93.3)[10.9] | 28.5 (21.9)[18.3] | 38.0 (37.9)[22.4] | 17.2 (5.7)[17.1] | 0.011 |
| DAB stanchion | 11.3 (6.0)[9.8] | 11.6 (13.5)[6.5] | 19.0 (22.5)[12.9] | 12.5 (5.6)[10.9] | 12.4 (5.2)[11.0] | 0.033 |
| DAB ground | 33.7 (67.7)[10.8] | 16.2 (23.8)[11.1] | 29.1 (22.9)[21.9] | 26.6 (18.4)[17.4] | 18.1 (15.8)[12.1] | 0.029 |
NAL Normal airway with lights on, DAL Difficult airway with lights on, DAB Difficult airway with blackout conditions
All values are expressed in seconds (SD). P < 0.05 indicated statistical significance
Paired analysis of devices where the mean time to intubation was statistically significantly different (only pairs where adjusted p–values were < 0.05 are shown)
| Group | Dunn’s Test | |
|---|---|---|
| Novice group | ||
| NAL ground | ||
| Airtraqb vs. C-MAC | 0.002 | |
| Airtraqb vs. GlideScope | 0.020 | |
| C-MAC vs. AWSb | 0.014 | |
| DAL stanchion | ||
| GlideScope vs. AWSb | 0.050 | |
| DAL ground | ||
| Airtraqb vs. C-MAC | 0.027 | |
| Airtraqb vs. Coopdech | 0.034 | |
| C-MAC vs. AWSb | 0.003 | |
| Coopdech vs. AWSb | 0.006 | |
| DAB stanchion | ||
| C-MAC vs. AWSb | 0.002 | |
| GlideScope vs. AWSb | 0.002 | |
| DAB ground | ||
| Airtraqb vs. C-MAC | 0.020 | |
| C-MAC vs. AWSb | <0.001 | |
| GlideScope vs. AWSb | 0.004 | |
| Expert group | ||
| DAL ground | ||
| Coopdech vs. AWSb | 0.009 | |
| DAB stanchion | ||
| C-MAC vs. AWSb | 0.043 | |
| Coopdech vs. AWSb | 0.046 | |
NAL Normal airway with lights on, DAL Difficult airway with lights on, DAB Difficult airway with blackout conditions. BH-adjusted p < 0.05 indicated statistical significance. aDunn’s Test p value adjusted for false discovery rate using Benjamin-Hochberg (BH) procedure; bThe device with shorter time to intubation in the paired results
Participants’proportion of attempts with one or more optimization maneuvers
| Condition | Airtraq | AWS | C-MAC | Coopdech | GlideScope | B-H adjusted |
|---|---|---|---|---|---|---|
| Novice group | ||||||
| NAL stanchion | 0.0% | 0.0% | 4.2% | 0.0% | 4.2% | 0.66 |
| NAL ground | 8.3% | 4.2% | 12.5% | 8.3% | 8.3% | 0.90 |
| DAL stanchion | 0.0% | 0.0% | 25.0% | 4.2% | 8.3% | 0.03 |
| DAL ground | 54.2% | 33.3% | 75.0% | 62.5% | 58.3% | 0.12 |
| DAB stanchion | 8.3% | 12.5% | 29.2% | 16.8% | 16.8% | 0.57 |
| DAB ground | 50.0% | 29.2% | 75.0% | 67.7% | 45.8% | 0.04 |
| Expert group | ||||||
| NAL Stanchion | 0.0% | 6.3% | 6.3% | 0.0% | 0.0% | 0.55 |
| NAL Ground | 6.3% | 0.0% | 18.8% | 12.5% | 12.5% | 0.55 |
| DAL Stanchion | 12.5% | 0.0% | 18.8% | 0.0% | 12.5% | 0.35 |
| DAL Ground | 27.5% | 18.8% | 75.0% | 50.0% | 50.0% | 0.08 |
| DAB Stanchion | 0.0% | 0.0% | 12.5% | 6.3% | 18.8% | 0.35 |
| DAB Ground | 43.8% | 12.5% | 50.0% | 68.8% | 25.0% | 0.07 |
NAL Normal airway with lights on, DAL Difficult airway with lights on, DAB Difficult airway with blackout conditions. BH-adjusted p < 0.05 indicated statistical significance
Survey results of participants’ subjective ease of use
| Question | BH-adjusted | Ranking of devices | ||||
|---|---|---|---|---|---|---|
| AWS | GlideScope | C-MAC | Coopdech | Airtraq | ||
| Novice group | ||||||
| Overall ease of use of the device | 0.180 | 4.7 | 4.8 | 4.3 | 3.8 | 3.7 |
| The ease of use of this device under the following circumstance | ||||||
| NAL Stanchion | 0.130 | 4.9 | 4.7 | 4.9 | 4.2 | 4 |
| NAL Ground | 0.110 | 4.6 | 4.4 | 4.1 | 3.5 | 3.9 |
| DAL Stanchion | 0.290 | 4.6 | 4.4 | 4.5 | 3.8 | 3.9 |
| DAL Ground | 0.001 | 4.6 | 3.9 | 3.2 | 3.2 | 3.6 |
| DAB Stanchion | 0.130 | 4.6 | 4.4 | 4.4 | 3.7 | 2.9 |
| DAB Ground | 0.130 | 4.6 | 4.6 | 3.3 | 3.2 | 3.2 |
| Expert group | ||||||
| Overall ease of use of the device | 0.400 | 4.4 | 4.7 | 4.6 | 4.2 | 3.6 |
| The ease of use of this device under the following circumstance | ||||||
| NAL stanchion | 0.047 | 4.7 | 4.7 | 4.4 | 4.0 | 4.6 |
| NAL ground | 0.120 | 4.5 | 4.7 | 4.2 | 3.3 | 4.4 |
| DAL stanchion | 0.120 | 4.6 | 4.3 | 4.0 | 3.9 | 4.0 |
| DAL ground | 0.088 | 4.5 | 3.8 | 3.4 | 2.2 | 4.0 |
| DAB stanchion | 0.001 | 4.4 | 4.3 | 3.6 | 3.9 | 3.0 |
| DAB ground | 0.004 | 4.5 | 4.2 | 3.6 | 3.3 | 3.1 |
NAL Normal airway with lights on, DAL Difficult airway with lights on, DAB Difficult airway with blackout conditions
All values are an average of ratings based on a 5-point Likert scale ranging from “5 = Extremely easy” to “1 = Extremely difficult”. BH-adjusted P-value <0.05 indicated statistical significance
Mean times for assembly and power-on in seconds [seconds (median)]
| Variable | Device name | BH-adjusted | ||||
|---|---|---|---|---|---|---|
| Airtraq | AWS | C-MAC | Coopdech | GlideScope | ||
| Assembly time | 245.8 (176.8) | 69.4 (42.6) | 91.2 (18.1) | 81.6 (47.9) | 44.4 (29.1) | 0.012 |
| Power-on time | 3.22 (3.3) | 4.19 (0.8) | 16.58 (0.3) | 2.90 (0.8) | 2.28 (2.5) | 0.0228 |
| Total average time | 249 (177.7) | 73.59 (41.9) | 107.8 (17.9) | 84.5 (48) | 46.68 (30.9) | 0.0228 |
All values are in seconds (SD). BH-adjusted p-value <0.05 indicated statistical significance. Data pooled from Novices and Experts