| Literature DB >> 31737366 |
Sedat Bilge1, Onur Tezel1, Yahya Ayhan Acar1, Guclu Aydin2, Attila Aydin3, Gokhan Ozkan4.
Abstract
BACKGROUND: The present study evaluates the success and efficacy of endotracheal intubation (ETI) using a modified intubation stylet and a magnet system to direct the stylet into the trachea. The system was developed by the researchers in an attempt to increase the success and efficacy of ETI.Entities:
Year: 2019 PMID: 31737366 PMCID: PMC6815604 DOI: 10.1155/2019/5804260
Source DB: PubMed Journal: Emerg Med Int ISSN: 2090-2840 Impact factor: 1.112
Figure 1(a) Modified intubation stylet produced by soldering an iron-rich metal ball onto the tip of a standard stylet; (b) the neodymium magnet designed for the study.
Figure 2Illustration of endotracheal intubation performed with an endotracheal tube fitted over a modified intubation stylet and a neodymium magnet placed on the anterior aspect of the neck at the level of the thyroid and cricoid cartilages.
Figure 3Flowchart of the study.
The holding power of the magnet on the intubation stylet measured from various distances.
| Distance (cm) | Measurement 1 (N) | Measurement 2 (N) | Mean (N) |
|---|---|---|---|
| 6 | 0.40 | 0.40 | 0.40 |
| 5 | 0.55 | 0d.60 | 0.58 |
| 4 | 0.80 | 0.80 | 0.80 |
| 3 | 1.20 | 1.20 | 1.20 |
| 2 | 1.70 | 1.65 | 1.68 |
| 1.5 | 2.10 | 2.10 | 2.10 |
| 1 | 2.40 | 2.50 | 2.45 |
cm, centimeter; N, newton.
Comparison of the procedure completion times and degree of difficulty between the two groups.
| Completion time (seconds) | Numeric rating scale# | ||||
|---|---|---|---|---|---|
| Mean ± SD |
| 95% CI | Mean ± SD |
| |
| METI | 18.31 ± 2.46 | 0.001 | 0.94–2.46 | 2.45 ± 0.76 | 0.001 |
| CETI | 20.01 ± 1.95 | 4.20 ± 0.83 | |||
Student's t-test; Wilcoxon rank sum test.