Calibration of GafChromic EBT3 for absorbed dose measurements in 5 MeV proton beam and (60)Co γ-rays.

M Vadrucci1, G Esposito2, C Ronsivalle1, R Cherubini3, F Marracino1, R M Montereali1, L Picardi1, M Piccinini1, M Pimpinella4, M A Vincenti1, C De Angelis2.   

Abstract

PURPOSE: To study EBT3 GafChromic film in low-energy protons, and for comparison purposes, in a reference (60)Co beam in order to use it as a calibrated dosimetry system in the proton irradiation facility under construction within the framework of the Oncological Therapy with Protons (TOP)-Intensity Modulated Proton Linear Accelerator for RadioTherapy (IMPLART) Project at ENEA-Frascati, Italy.
METHODS: EBT3 film samples were irradiated at the Istituto Nazionale di Fisica Nucleare-Laboratori Nazionali di Legnaro, Italy, with a 5 MeV proton beam generated by a 7 MV Van de Graaff CN accelerator. The nominal dose rates used were 2.1 Gy/min and 40 Gy/min. The delivered dose was determined by measuring the particle fluence and the energy spectrum in air with silicon surface barrier detector monitors. A preliminary study of the EBT3 film beam quality dependence in low-energy protons was conducted by passively degrading the beam energy. EBT3 films were also irradiated at ENEA-National Institute of Ionizing Radiation Metrology with gamma radiation produced by a (60)Co source characterized by an absorbed dose to water rate of 0.26 Gy/min as measured by a calibrated Farmer type ionization chamber. EBT3 film calibration curves were determined by means of a set of 40 film pieces irradiated to various doses ranging from 0.5 Gy to 30 Gy absorbed dose to water. An EPSON Expression 11000XL color scanner in transmission mode was used for film analysis. Scanner response stability, intrafilm uniformity, and interfilm reproducibility were verified. Optical absorption spectra measurements were performed on unirradiated and irradiated EBT3 films to choose the most sensitive color channel to the dose range used.
RESULTS: EBT3 GafChromic films show an under response up to about 33% for low-energy protons with respect to (60)Co gamma radiation, which is consistent with the linear energy transfer dependence already observed with higher energy protons, and a negligible dose-rate dependence in the 2-40 Gy/min range. Short- and long-term scanner stabilities were 0.5% and 1.5%, respectively; film uniformity and reproducibility were better than 0.5%.
CONCLUSIONS: The main purpose of this study was to implement EBT3 dosimetry in the proton low-energy radiobiology line of the TOP-IMPLART accelerator, having a maximum energy of 7 MeV. Low-energy proton and (60)Co calibrated sources were used to investigate the behavior of film response vs to be written in italicum dose. The calibration in 5 MeV protons is currently used for dose assessment in the radiobiological experiments at the TOP-IMPLART accelerator carried out at that energy value.

Entities:  

Mesh:

Substances:

Year:  2015        PMID: 26233195     DOI: 10.1118/1.4926558

Source DB:  PubMed          Journal:  Med Phys        ISSN: 0094-2405            Impact factor:   4.071


  6 in total

1.  Patient-Specific QA of Spot-Scanning Proton Beams using Radiochromic Film.

Authors:  Maria F Chan; Chin-Cheng Chen; Chengyu Shi; Jingdong Li; Xiaoli Tang; Xiang Li; Dennis Mah
Journal:  Int J Med Phys Clin Eng Radiat Oncol       Date:  2017-05-16

2.  LET dependent response of GafChromic films investigated with MeV ion beams.

Authors:  V Grilj; D J Brenner
Journal:  Phys Med Biol       Date:  2018-12-18       Impact factor: 3.609

3.  Radiation dose verification of an X-ray based blood irradiator using EBT3 radiochromic films calibrated using Gamma Knife machine.

Authors:  Khaled Soliman; Marouf Adili; Abdullah Alrushoud
Journal:  Rep Pract Oncol Radiother       Date:  2019-06-07

4.  Investigating Dependencies of Relative Biological Effectiveness for Proton Therapy in Cancer Cells.

Authors:  Michelle E Howard; Chris Beltran; Sarah Anderson; Wan Chan Tseung; Jann N Sarkaria; Michael G Herman
Journal:  Int J Part Ther       Date:  2018-03-21

5.  Dose- rather than fluence-averaged LET should be used as a single-parameter descriptor of proton beam quality for radiochromic film dosimetry.

Authors:  Andreas Franz Resch; Paul David Heyes; Hermann Fuchs; Niels Bassler; Dietmar Georg; Hugo Palmans
Journal:  Med Phys       Date:  2020-03-13       Impact factor: 4.071

6.  Expired EBT3 Films' Sensitivity for the Measurement of X-ray and UV Radiation: An Optical Analysis.

Authors:  Hamad Yahia Abu Mhanna; Ahmad Fairuz Omar; Yasmin Md Radzi; Hanan Fawaz Akhdar; Haytham Al Ewaidat
Journal:  Materials (Basel)       Date:  2022-04-15       Impact factor: 3.748

  6 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.