Tareq Meyer1,2, Omid Nikoubashman1,3, Lisa Kabelitz1, Marguerite Müller1, Ahmed Othman1, Saif Afat1, Martin Kramer4, Martin Wiesmann1, Marc A Brockmann1,2, Carolin Brockmann1,2.
Abstract
Feasibility of endovascular stentectomy using a snare over stent-retriever (SOS) technique was evaluated in a silicon flow model and an in vivo swine model. In vitro, stentectomy of different intracranial stents using the SOS technique was feasible in 22 out of 24 (92%) retrieval maneuvers. In vivo, stentectomy was successful in 10 out of 10 procedures (100%). In one case self-limiting vasospasm was observed angiographically as a technique related complication in the animal model. Endovascular stentectomy using the SOS technique is feasible in an experimental setting and may be transferred to a clinical scenario.Entities:
Mesh:
Year: 2017 PMID: 28542580 PMCID: PMC5444789 DOI: 10.1371/journal.pone.0178197
Source DB: PubMed Journal: PLoS One ISSN: 1932-6203 Impact factor: 3.240
Fig 1Stentectomy technique.
1.) The loop of a microsnare was slipped over the tip of a microcatheter, slightly tightened, (1A) and the microcatheter was inserted into a long sheath or guide catheter (1A) 2.) The microcatheter together with snare was positioned a few milimeters proximal of the mal-deployed stent (1C) 3.) A stent-retriever was advanced through the microcatheter and partially unfolded, so that the distal part of the stent-retriever attached to the vessel wall. 4.) The partially unfolded stent-retriever was pushed forward carefully, until the distal markers of the stent-retriever overlapped with the proximal markers of the mal-deployed stent (1D). Using a stent-retriever with a comparable or larger diameter than the target stent seems to be helpful. 5.) The stent-retriever is being resheathed while applying slight pressure, hereby „grabbing”the proximal end of the mal-deployed stent (1E). Doing so, some or all of the proximal markers of the maldeployed stent move towards the tip of the microcatheter (1E). 6.) The microsnare is being opened (1F), pushed forward over the microcatheter and over the proximal end of the maldeployed stent (1G). 7.) Finally, the microsnare is pulled close again (1H). 8.) The stent-retriever and the snare are „locked”in their position by tightening a torque on the wire of both devices directly adjacent to the hemostatic valve (not shown). 9.) To extract the maldeployed stent, the snare and the stent-retriever are being pulled back slowly into the sheath or guide catheter (1I) (aspirate to prevent embolism). 10.) If it should be impossible to pull a stent into the sheath or guide catheter, one should consider to “move” the stent into a less relevant vessel, like the external carotid artery or in a brachial or femoral vessel that can be reached more easily by a vascular surgeon.
In vitro procedures.
| Solitaire FR 6 x 30 mm | Enterprise 4.5 x 22 mm | 4.5 mm | ++ | — | ✓ | |
| Amplatz GooseNeck Microsnare 7 mm | ||||||
| Solitaire FR 6 x 30 mm | Enterprise 4.5 x 22 mm | 4.5 mm | +++ | Stent-retriever imposed on Enterprise after 2nd try | ✓ | |
| Amplatz GooseNeck Microsnare 7 mm | ||||||
| Solitaire FR 6 x 30 mm | Enterprise 4.5 x 22 mm | 4.5 mm | + | — | ✓ | |
| Amplatz GooseNeck Microsnare 7 mm | ||||||
| Solitaire FR 6 x 30 mm | Enterprise 4.5 x 22 mm | 4.5 mm | ++ | Stent-retriever imposed on Enterprise after 2nd try | ✓ | |
| Amplatz GooseNeck Microsnare 4 mm | ||||||
| Solitaire FR 6 x 30 mm | Enterprise 4.5 x 22 mm | 2.5 mm | ++ | — | ✓ | |
| Amplatz GooseNeck Microsnare 4 mm | ||||||
| Solitaire FR 6 x 30 mm | Enterprise 4.5 x 22 mm | 2.5 mm | +++ | — | ✓ | |
| Amplatz GooseNeck Microsnare 4 mm | ||||||
| Solitaire FR 6 x 30 mm | Enterprise 4.5 x 22 mm | 2.5 mm | +++ | — | ✓ | |
| Amplatz GooseNeck Microsnare 4 mm | ||||||
| Solitaire FR 6 x 30 mm | Enterprise 4.5 x 22 mm | 2.5 mm | + | — | ✓ | |
| Amplatz GooseNeck Microsnare 4 mm | ||||||
| Solitaire FR 6 x 30 mm | PED 4.5 x 25 mm | 2.5 mm | +++ | PED lost during extraction | × | |
| Amplatz GooseNeck Microsnare 4 mm | ||||||
| Trevo XP ProVue 3 x 20 mm | PED 4.5 x 25 mm | 2.5 mm | +++ | Stent-retriever did not interlock with PED | × | |
| Amplatz GooseNeck Microsnare 4 mm | ||||||
| Solitaire FR 4 x 20 mm | PED 4.5 x 25 mm | 2.5 mm | +++ | — | ✓ | |
| Amplatz GooseNeck Microsnare 4 mm | ||||||
| Solitaire FR 4 x 20 mm | PED 4.5 x 25 mm | 2.5 mm | +++ | — | ✓ | |
| Amplatz GooseNeck Microsnare 4 mm | ||||||
| Solitaire FR 4 x 20 mm | FRED 5.5 x 25 mm | 2.5 mm | +++ | — | ✓ | |
| Amplatz GooseNeck Microsnare 4 mm | ||||||
| Solitaire FR 4 x 20 mm | FRED 5.5 x 25 mm | 2.5 mm | + | — | ✓ | |
| Amplatz GooseNeck Microsnare 4 mm | ||||||
| Solitaire FR 6 x 30 mm | Acclino 4.5 x 22 mm | 4.3 mm | +++ | — | ✓ | |
| Amplatz GooseNeck Microsnare 4 mm | ||||||
| Solitaire FR 6 x 30 mm | Acclino 4.5 x 22 mm | 4.3 mm | ++ | — | ✓ | |
| Amplatz GooseNeck Microsnare 4 mm | ||||||
| Solitaire FR 6 x 30 mm | Acclino 4.5 x 22 mm | 4.3 mm | ++ | — | ✓ | |
| Amplatz GooseNeck Microsnare 4 mm | ||||||
| Solitaire FR 6 x 30 mm | Acclino 4.5 x 22 mm | 4.3 mm | +++ | — | ✓ | |
| Amplatz GooseNeck Microsnare 4 mm | ||||||
| Solitaire FR 6 x 30 mm | Acclino 4.5 x 22 mm | 4.3 mm | +++ | — | ✓ | |
| Amplatz GooseNeck Microsnare 4 mm | ||||||
| Solitaire FR 6 x 30 mm | Acclino 4.5 x 22 mm | 4.3 mm | + | — | ✓ | |
| Amplatz GooseNeck Microsnare 4 mm | ||||||
| Solitaire FR 6 x 30 mm | Acclino 4.5 x 22 mm | 4.3 mm | +++ | — | ✓ | |
| Amplatz GooseNeck Microsnare 4 mm | ||||||
| Solitaire FR 6 x 30 mm | Acclino 4.5 x 22 mm | 4.3 mm | +++ | — | ✓ | |
| Amplatz GooseNeck Microsnare 4 mm | ||||||
| Solitaire FR 6 x 30 mm | Acclino 4.5 x 22 mm | 4.3 mm | +++ | — | ✓ | |
| Amplatz GooseNeck Microsnare 4 mm | ||||||
| Solitaire FR 6 x 30 mm | Acclino 4.5 x 22 mm | 4.3 mm | +++ | — | ✓ | |
| Amplatz GooseNeck Microsnare 4 mm |
In vivo procedures.
| Trevo XP ProVue 4 x 20 mm | Enterprise 4.5 x 28 mm | 2.7 mm | + | — | ✓ | |
| Amplatz GooseNeck Microsnare 7 mm | ||||||
| Trevo XP ProVue 4 x 20 mm | Enterprise 4.5 x 28 mm | 3.3 mm | +++ | — | ✓ | |
| Amplatz GooseNeck Microsnare 7 mm | ||||||
| Trevo XP ProVue 4 x 20 mm | Enterprise 4.5 x 28 mm | 2.6 mm | + | — | ✓ | |
| Amplatz GooseNeck Microsnare 7 mm | ||||||
| Trevo XP ProVue 4 x 20 mm | Enterprise 4.5 x 28 mm | 2.3 mm | +++ | self limiting vasospasm | ✓ | |
| Amplatz GooseNeck Microsnare 4 mm | ||||||
| Solitaire FR 4 x 20 mm | Pharos Vitesse incompletely inflated | 2.2 mm | +++ | — | ✓ | |
| Amplatz GooseNeck Microsnare 4 mm | ||||||
| Solitaire FR 4 x 20 mm | Pharos Vitesse incompletely inflated | 2.1 mm | +++ | — | ✓ | |
| Amplatz GooseNeck Microsnare 4 mm | ||||||
| Solitaire FR 4 x 20 mm | Pharos Vitesse incompletely inflated | 2.4 mm | +++ | — | ✓ | |
| Amplatz GooseNeck Microsnare 4 mm | ||||||
| Solitaire FR 4 x 20 mm | Pharos Vitesse incompletely inflated | 2.9 mm | +++ | — | ✓ | |
| Amplatz GooseNeck Microsnare 4 mm | ||||||
| Solitaire FR 4 x 20 mm | Pharos Vitesse incompletely inflated | 2.8 mm | +++ | — | ✓ | |
| Amplatz GooseNeck Microsnare 4 mm | ||||||
| Trevo XP ProVue 4 x 20 mm | Neuroform Atlas 4.0 x 24 mm | 3.2 mm | +++ | — | ✓ | |
| Amplatz GooseNeck Microsnare 4 mm |