| Literature DB >> 32166403 |
Andreas Simgen1, Michael Kettner2, Philipp Dietrich2, Toshiki Tomori2, Ruben Mühl-Benninghaus2, Pervinder Bhogal3, Matthias W Laschke4, Michael D Menger4, Wolfgang Reith2, Umut Yilmaz2.
Abstract
PURPOSE: Treatment of wide-necked intracranial aneurysms using the Woven Endobridge (WEB) device has become broadly accepted. Feared complications with the potential of increased poor clinical outcome include dislocations and migration of the device. This study was carried out to determine the effectiveness of a variety of different strategies to rescue migrated WEB devices.Entities:
Keywords: Alligator retrieval device; JET 7; Microsnare; SOFIA Plus
Year: 2020 PMID: 32166403 PMCID: PMC8211610 DOI: 10.1007/s00062-020-00893-3
Source DB: PubMed Journal: Clin Neuroradiol ISSN: 1869-1439 Impact factor: 3.649
Fig. 1Microcomputed tomography (CT) fluoroscopy of the rescued WEB devices SL 3.5 × 2 mm (a) and SLS 8 (b). The white arrow indicates the slight deformation at the proximal marker. Scale bar = 1 mm
Fig. 2a Digital subtraction angiography (DSA) of the right subclavian artery, axillary arteries and their branches. b Roadmap image of the migrated WEB device (SL 3.5 × 2 mm) with its distal and proximal radiopaque marker (white arrowheads) and the JET 7 (white arrow). c Advancement of the JET 7 (white arrow) until the proximal marker (white arrowhead) was inside the JET 7. d After suction had been applied, the WEB device (white arrowheads) was fully aspirated into the JET 7 and the catheter was withdrawn (white arrow)
Overview of the results comparing the different rescue approaches presented as means ± standard deviations.
| Rescue devices | WEB Devices | Rescue rates [%] | Time of rescue [s] | Attempts, mean [ | Perforation | Dissection | Vasospasm | Entrapment at RC [ |
|---|---|---|---|---|---|---|---|---|
| SOFIA Plus | SL 3.5 × 2/4 × 3 | 100 | 69.62 ± 37.65 | 1.37 | 0 | 0 | 0 | 0 |
| JET 7 | SL 3.5 × 2/4 × 3 | 100 | 54.25 ± 36.51 | 1.25 | 0 | 0 | 0 | 0 |
| – | – | 0.421 | 0.619 | – | – | – | – | |
| Microsnare | SL 8 × 5/SLS 8 | 100 | 122.75 ± 41.15 | 1 | 0 | 0 | 0 | 0 |
| Alligator | SL 8 × 5/SLS 8 | 100 | 137.50 ± 54.46 | 1.37 | 0 | 0 | 0 | 0 |
| – | – | 0.551 | 0.060 | – | – | – | – |
RC reperfusion catheter
Fig. 3a Rescue of the WEB device (SLS 8) using the Microsnare leading to an hourglass configuration. b Rescue of the WEB device (SLS 8) using the ARD at the proximal marker with partially resheated WEB device within the VIA-27 microcatheter