| Literature DB >> 28500459 |
Anna M Sailer1,2, Laura Vergoossen3, Leonie Paulis3, Willem H van Zwam3, Marco Das3,4, Joachim E Wildberger3,4, Cécile R L P N Jeukens3.
Abstract
PURPOSE: Radiation safety and protection are a key component of fluoroscopy-guided interventions. We hypothesize that providing weekly personal dose feedback will increase radiation awareness and ultimately will lead to optimized behavior. Therefore, we designed and implemented a personalized feedback of procedure and personal doses for medical staff involved in fluoroscopy-guided interventions.Entities:
Keywords: Endovascular procedures; Interventional radiology; Occupational dose; Radiation dosimetry; Radiation exposure; Radiation monitoring; Radiation protection
Mesh:
Year: 2017 PMID: 28500459 PMCID: PMC5651709 DOI: 10.1007/s00270-017-1690-5
Source DB: PubMed Journal: Cardiovasc Intervent Radiol ISSN: 0174-1551 Impact factor: 2.740
Fig. 1Example of a weekly personalized dose feedback form of a physician
Fig. 2Closed questions of the questionnaire to evaluate the personalized dose feedback among medical staff
Distribution of staff doses during phase 1 (pre feedback) and phase 2 (with feedback)
| Median | 25 Percentile | 75 Percentile | Mean | SD | Min | Max | |||
|---|---|---|---|---|---|---|---|---|---|
| Normalized dose FO (E/DAP) | Prefeedback | [μSv/Gycm²] | 0.52 | 0.17 | 1.45 | <0.01 | 11.73 | 1.19 | 1.81 |
| With feedback | [μSv/Gycm²] | 0.4 | 0.15 | 1.27 | <0.01 | 6.82 | 0.95 | 1.32 | |
| Normalized dose FT (E/DAP) | Prefeedback | [μSv/Gycm²] | 0.12 | 0.04 | 0.5 | <0.01 | 13.2 | 0.6 | 1.42 |
| With feedback | [μSv/Gycm²] | 0.08 | 0.02 | 0.24 | <0.01 | 10.37 | 0.32 | 0.95 | |
| Absolute over-lead dose FO (E) | Prefeedback | [μSv] | 12.11 | 2.12 | 33.11 | 0.02 | 614.1 | 37.05 | 73.11 |
| With feedback | [μSv] | 11.68 | 3.11 | 35.91 | <0.01 | 603.92 | 37.17 | 77.54 | |
| Absolute over-lead dose FT (E) | Prefeedback | [μSv] | 2.64 | 0.76 | 5.91 | <0.01 | 136.92 | 6.45 | 14.04 |
| With feedback | [μSv] | 2.05 | 0.767 | 5.49 | <0.01 | 547 | 9.7 | 50.07 |
E effective dose. DAP dose area product. SD standard deviation
Staff doses presented as normalized for procedure dose as well as absolute values
Fig. 3Graphical representation of scores of medical staff’ answers on the questionnaire’s closed questions. Numbers in colored bars indicate percentages