| Literature DB >> 28487829 |
Vahideh Nazari1, Seied R Mahdavi1, Ahmad Mostaar2, Hassan Nedaei3, Mohammad A Mosleh Shirazi4, Alireza Nikoofar4, Golbarg Esmailie5.
Abstract
Dosimetric verification of radiation treatment has recently been extended by the introduction of electronic portal imaging devices (EPIDs). Detailed dose response specifications of EPID should be addressed prior to any dosimetric application. The present study evaluates improvements of dosimetric properties of the low elbow camera-based EPID Theraview (Cablon Medical, Leusden, The Netherlands) equipped with a cooled charge coupled device (CCD) for portal dosimetry. The dose response, warm-up behavior, stability over long- and short-term scales (throughout a day) were studied. The field size dependency of the EPID response was also investigated and compared with ion chamber measurements under the same conditions. The EPID response without saturation for doses up to 2 Gy was linear for both beam qualities (6 and 15 MV). There was no evident warm-up characteristic. The detector sensitivity showed excellent stability in short term [standard deviation (SD) 0.38%]. In long-term stability (over a period of approximately 3 months), a negligible linear decline of 0.01% per day was observed. It was concluded that the cooled CCD camera-based EPID could be used for portal dosimetry, after accurate corrections for the field size dependency and sensitivity loss.Entities:
Keywords: Electronic portal imaging devices; elbow; portal dosimetry; radiometry
Year: 2017 PMID: 28487829 PMCID: PMC5394802
Source DB: PubMed Journal: J Med Signals Sens ISSN: 2228-7477
Figure 1Dose response of electronic portal imaging device for 6 MV photons
Figure 2Buildup effects on EPID behavior to dose, for 6 and 15 MV photons
Temporal changes EPID response in 6 MV photon beams in short-term and long-term scales*
Figure 3Sensitivity changes of EPID after 3 months, normalized to linac output
Figure 4EPID warm-up behavior in 6 MV photons, detector response normalized to mean pixel intensity in each series
Figure 5EPID response to dose in different field sizes for 6 MV photons
Figure 6Field size dependency of EPID response in comparison to 2D array for the same geometry and fixed level, both of detector response normalized to 10 cm × 10 cm field size