Dosimetric properties of the Theraview fluoroscopic electronic portal imaging device.

A G Glendinning1, D E Bonnett.   

Abstract

Electronic portal imaging devices (EPIDs) can be used for non-imaging applications in radiotherapy such as patient dosimetry. Of the systems available, the fluoroscopic camera-based EPID Theraview (InfiMed Inc.) has not been studied to date, and a review of the dosimetric properties of the system is presented here. In the "single set-up" mode of image acquisition, pixel intensity increases sublinearly with applied dose. The response was dependent on the system's video signal gain and showed a threshold dose to the detector in the range 0.05-0.35 cGy, and pixel saturation at detector doses in the range 1.2-1.6 cGy. Repeated exposures of the EPID were observed to be extremely reproducible (standard deviation 0.5%). The sensitivity of the system showed a linear decline of 0.04% day-1 over a 68-day period, during which time the relative off-axis response within 10 x 10 cm2 field was constant to within a standard deviation of 0.56%. The system shows spatial non-uniformity, which requires correction for application to dose measurements in two-dimensions. Warm-up of the camera control unit required a period of at least 40 min and was associated with an enhancement in pixel intensity of up to 12%. A radiation dose history effect was observed at doses as low as 0.2 Gy. Camera dark current was shown to be negligible at normal accelerator operation. No discernible image distortion was found. Mechanical stability on gantry rotation was also assessed and image displacement of up to 5 mm at the isocentre was observed. It was concluded that the device could be used for dosimetry provided necessary precautions were observed and corrections made.

Entities:  

Mesh:

Year:  2000        PMID: 10884749     DOI: 10.1259/bjr.73.869.10884749

Source DB:  PubMed          Journal:  Br J Radiol        ISSN: 0007-1285            Impact factor:   3.039


  2 in total

1.  Third party EPID with IGRT capability retrofitted onto an existing medical linear accelerator.

Authors:  D O Odero; D S Shimm
Journal:  Biomed Imaging Interv J       Date:  2009-07-01

2.  Multi-parametric Improvements in the CCD Camera-based EPID for Portal Dosimetry.

Authors:  Vahideh Nazari; Seied R Mahdavi; Ahmad Mostaar; Hassan Nedaei; Mohammad A Mosleh Shirazi; Alireza Nikoofar; Golbarg Esmailie
Journal:  J Med Signals Sens       Date:  2017 Jan-Mar
  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.