Literature DB >> 27535443

Catastrophic Femoral Head-Stem Trunnion Dissociation Secondary to Corrosion.

Laura Matsen Ko1, Antonia F Chen2, Gregory K Deirmengian2, William J Hozack2, Peter F Sharkey3.   

Abstract

BACKGROUND: Modular femoral heads provide procedural enhancement by allowing accurate restoration of hip offset and limb-length equalization. However, corrosion may lead to adverse local tissue reactions. Severe trunnion corrosion can also lead to femoral head dissociation and catastrophic implant failure following primary total hip arthroplasty.
METHODS: We describe 5 cases, from our institution, in which the femoral head became dissociated from the femoral stem trunnion secondary to severe corrosion. Possible causes are evaluated.
RESULTS: Demographic commonalities among the 5 patients included a body mass index (BMI) of ≥30 kg/m(2) and male sex. All femoral heads were made of cobalt-chromium alloy and were larger-diameter implants (≥36 mm). Four of the 5 patients had a femoral head that increased the neck length above the default on a so-called standard head and 3 of the 5 had a stem with a 127° neck-shaft angle.
CONCLUSIONS: Although dissociation of the femoral head from the femoral trunnion following total hip arthroplasty is exceedingly rare, the prevalence may increase with longer follow-up. The dissociation is likely related to multiple factors, including a BMI of ≥30 kg/m(2), male sex, and corrosion resulting from the use of a larger metal head with a neck length of greater than the default and a stem with high offset. It is critical that surgeons be able to recognize this mode of implant failure and appropriately prepare to remove the femoral component during revision surgery. LEVEL OF EVIDENCE: Therapeutic Level IV. See Instructions for Authors for a complete description of levels of evidence.
Copyright © 2016 by The Journal of Bone and Joint Surgery, Incorporated.

Entities:  

Mesh:

Year:  2016        PMID: 27535443     DOI: 10.2106/JBJS.15.00914

Source DB:  PubMed          Journal:  J Bone Joint Surg Am        ISSN: 0021-9355            Impact factor:   5.284


  19 in total

1.  Trunnion Failure of the Recalled Low Friction Ion Treatment Cobalt Chromium Alloy Femoral Head.

Authors:  Kenneth L Urish; Brian R Hamlin; Anton Y Plakseychuk; Timothy J Levison; Genymphas B Higgs; Steven M Kurtz; Anthony M DiGioia
Journal:  J Arthroplasty       Date:  2017-04-08       Impact factor: 4.757

Review 2.  [Hip dislocation after revision arthroplasty : Risk assessment and treatment strategies].

Authors:  P M Prodinger; J Schauwecker; H Mühlhofer; N Harrasser; F Pohlig; C Suren; R von Eisenhart-Rothe
Journal:  Orthopade       Date:  2017-02       Impact factor: 1.087

3.  Head-stem trunnion dissociation due to corrosion in total hip arthroplasty.

Authors:  Michael R Whitehouse
Journal:  Ann Transl Med       Date:  2017-09

4.  Blood cobalt ion level in patients with different sizes of cobalt/chrome femoral head with the Accolade TMZF stem.

Authors:  Seiya Ishii; Yasuhiro Homma; Takehisa Matsukawa; Tomonori Baba; Kazuo Kaneko; Muneaki Ishijima
Journal:  Int Orthop       Date:  2022-07-15       Impact factor: 3.479

5.  Ceramic Bearings with Titanium Adapter Sleeves Implanted During Revision Hip Arthroplasty Show Minimal Fretting or Corrosion: a Retrieval Analysis.

Authors:  Chelsea N Koch; Mark Figgie; Mark P Figgie; Marcella E Elpers; Timothy M Wright; Douglas E Padgett
Journal:  HSS J       Date:  2017-06-22

Review 6.  Gross trunnion failure in metal on polyethylene total hip arthroplasty-a systematic review of literature.

Authors:  Tungish Bansal; Sameer Aggarwal; Mandeep Singh Dhillon; Sandeep Patel
Journal:  Int Orthop       Date:  2020-01-04       Impact factor: 3.075

7.  Comparison of Asymptomatic and Symptomatic Adverse Local Tissue Reaction in Patients With Head-Neck Taper Corrosion.

Authors:  Margaret A Weber; Matthew J Snyder; Kalain K Workman; Margaret M Sims; Clair N Smith; Deepak Kumar; Akshay V Daji; Camilo G Borrero; Andrew C Cordle; Anthony M DiGioia; Brian R Hamlin; Anton Y Plakseychuk; Kenneth L Urish
Journal:  J Arthroplasty       Date:  2020-09-17       Impact factor: 4.435

8.  What Is the Risk of THA Revision for ARMD in Patients with Non-metal-on-metal Bearings? A Study from the Australian National Joint Replacement Registry.

Authors:  R N de Steiger; Alesha Hatton; Yi Peng; Stephen Graves
Journal:  Clin Orthop Relat Res       Date:  2020-06       Impact factor: 4.755

9.  Bead debonding from a modern cementless total hip femoral component with concomitant taper corrosion.

Authors:  Adam I Edelstein; Duy L Phan; Scott M Sporer
Journal:  Arthroplast Today       Date:  2017-05-06

10.  Head size in primary total hip arthroplasty.

Authors:  Georgios Tsikandylakis; Maziar Mohaddes; Peter Cnudde; Antti Eskelinen; Johan Kärrholm; Ola Rolfson
Journal:  EFORT Open Rev       Date:  2018-05-21
View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.