Literature DB >> 28087160

MRI, Retrieval Analysis, and Histologic Evaluation of Adverse Local Tissue Reaction in Metal-on-Polyethylene Total Hip Arthroplasty.

Scott R Nodzo1, Christina I Esposito2, Hollis G Potter3, Chitranjan S Ranawat1, Timothy M Wright2, Douglas E Padgett1.   

Abstract

BACKGROUND: The severity and location of adverse local tissue reaction (ALTR) seen in metal-on-polyethylene total hip arthroplasty (THA) is not well described.
METHODS: We retrospectively reviewed the records of patients who underwent a revision THA using our biomechanics database. We included all patients who underwent revision surgery for the diagnosis of ALTR with THA implants that had modularity solely at the head-neck junction and excluded patients with implant modularity at sites other than the head-neck junction. Magnetic resonance imaging (MRI) was evaluated by a fellowship-trained radiologist who specializes in evaluating metal artifact reducing MRI sequences to quantify the ALTR lesions. Histology was evaluated for findings of ALTR using the Campbell score.
RESULTS: We identified 11 patients in the database. Eight patients had an MRI ALTR grade of severe and 7 did based on the histology score. The mean synovial volume was 218,658 mm3 (range 23,461-451,435 mm3) with a mean maximal synovial thickness of 15.3 mm (range 3-34.3 mm). A disruptive infiltration of the abductors due to pseudocapsule invasion was seen in 67% of the patients with 3 not having abductor involvement. Mean preoperative cobalt and chromium levels were 5.4 ppb (range 1-12.3 ppb) and 1.1 ppb (range 0.6-2.4 ppb), respectively.
CONCLUSION: Patients with ALTR from head-neck junction corrosion in metal-on-polyethylene THA may present with large pseudotumors that have previously been under appreciated. Mean preoperative serum cobalt and chromium ion levels remained relatively low, and MRI was an effective way to characterize the size and location of these lesions.
Copyright © 2016 Elsevier Inc. All rights reserved.

Entities:  

Keywords:  adverse local tissue reaction; magnetic resonance imaging; pseudotumor; total hip arthroplasty; tribocorrosion

Mesh:

Substances:

Year:  2016        PMID: 28087160     DOI: 10.1016/j.arth.2016.11.046

Source DB:  PubMed          Journal:  J Arthroplasty        ISSN: 0883-5403            Impact factor:   4.757


  5 in total

1.  Trunnion Failure of the Recalled Low Friction Ion Treatment Cobalt Chromium Alloy Femoral Head.

Authors:  Kenneth L Urish; Brian R Hamlin; Anton Y Plakseychuk; Timothy J Levison; Genymphas B Higgs; Steven M Kurtz; Anthony M DiGioia
Journal:  J Arthroplasty       Date:  2017-04-08       Impact factor: 4.757

Review 2.  [Tribology in hip arthroplasty : Benefits of different materials].

Authors:  J Philippe Kretzer; Maximilian Uhler; Sebastian Jäger; Therese Bormann; Robert Sonntag; Mareike Schonhoff; Stefan Schröder
Journal:  Orthopade       Date:  2021-02-25       Impact factor: 1.087

3.  How Useful Is Magnetic Resonance Imaging in Evaluating Adverse Local Tissue Reaction?

Authors:  Douglas E Padgett; Edwin P Su; Timothy M Wright; Alissa J Burge; Hollis G Potter
Journal:  J Arthroplasty       Date:  2020-01-15       Impact factor: 4.757

4.  MRI of THA Correlates With Implant Wear and Tissue Reactions: A Cross-sectional Study.

Authors:  Matthew F Koff; Christina Esposito; Parina Shah; Mauro Miranda; Elexis Baral; Kara Fields; Thomas Bauer; Douglas E Padgett; Timothy Wright; Hollis G Potter
Journal:  Clin Orthop Relat Res       Date:  2019-01       Impact factor: 4.176

5.  Synovial Fluid Metal Ion Levels are Superior to Blood Metal Ion Levels in Predicting an Adverse Local Tissue Reaction in Failed Total Hip Arthroplasty.

Authors:  Matthew T Houdek; Michael J Taunton; Cody C Wyles; Paul J Jannetto; David G Lewallen; Daniel J Berry
Journal:  J Arthroplasty       Date:  2021-05-01       Impact factor: 4.435

  5 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.