Since recovery time of the RF coil is long at low field MRI, the rising and the ring-down times of the square pulse are also long, which means the applied sinc pulse can easily be distorted from the changing amplitude. However, both the rising time and ring-down time can be calculated using Q-factor. Using this information, an RF square pulse were compensated by appending two square pulses before and after the RF pulse. The durations of these RF square pulses were calculated using the Q-factor. Since the amplitude of the sinc pulse changes continuously, a series of square pulses were applied to apply sinc pulse to the coil. The minimum number of square pulses and the amplitude of the square pulses were calculated. It was successfully demonstrated that the sinc pulse can be compensated using a series of square pulses. The more number of square pulses were used, the smoother sinc pulse was applied to the RF coil. The Q-factor was experimentally calculated from the ring-down time of a signal induced in a sniffer loop which was connected to an oscilloscope. The resulting Q-factor was then used to calculate both the duration and amplitude of the square pulses for compensation. Echo trains were also acquired in an inhomogeneous B0 field using the compensated RF pulses. In order to enhance the SNR of the echo trains, a pre-polarization pulse was added to the CPMG spin echo sequence. The SNRs of the echo signal acquired using compensated pulses were compared with those of signal obtained with uncompensated pulses and showed significant improvements of 61.1% and 51.5% for the square and sinc shaped pulses respectively.
Since recovery time of the RF coil is long at low field MRI, the rising and the ring-down times of the square pulse are also long, which means the applied sinc pulse can easily be distorted from the changing amplitude. However, both the rising time and ring-down time can be calculated using Q-factor. Using this information, an RF square pulse were compensated by appending two square pulses before and after the RF pulse. The durations of these RF square pulses were calculated using the Q-factor. Since the amplitude of the sinc pulse changes continuously, a series of square pulses were applied to apply sinc pulse to the coil. The minimum number of square pulses and the amplitude of the square pulses were calculated. It was successfully demonstrated that the sinc pulse can be compensated using a series of square pulses. The more number of square pulses were used, the smoother sinc pulse was applied to the RF coil. The Q-factor was experimentally calculated from the ring-down time of a signal induced in a sniffer loop which was connected to an oscilloscope. The resulting Q-factor was then used to calculate both the duration and amplitude of the square pulses for compensation. Echo trains were also acquired in an inhomogeneous B0 field using the compensated RF pulses. In order to enhance the SNR of the echo trains, a pre-polarization pulse was added to the CPMG spin echo sequence. The SNRs of the echo signal acquired using compensated pulses were compared with those of signal obtained with uncompensated pulses and showed significant improvements of 61.1% and 51.5% for the square and sinc shaped pulses respectively.
The signal-to-noise ratio (SNR) in magnetic resonance imaging (MRI) is proportional to the static magnetic field, and is related to image contrast, resolution, and acquisition time. To achieve high SNR, higher magnetic field strengths generated from superconducting magnets (1.5–7 T) are commonly used for clinical and research studies. However, such magnets are expensive to purchase, site, install, and maintain (e.g. helium refills, cold head replacement), such that access to MRI, especially in underdeveloped and developing countries is limited [1]. In response there has been a renewed push for alternative small footprint low field MRIs. Resistive and permanent magnets are candidates for low field MRI [2-7]. While moving to low field sacrifices some of the image quality achievable at high field MRI, there are still many benefits including: 1) the magnetic field generated by main magnet is safer as it reduces the risk of projectiles; 2) the specific absorption rate (SAR) associated with RF pulses is low (proportional to the square of magnetic field); 3) there can be greatly reduced acoustic noise from gradient coils since the forces on the gradients are proportional to the magnetic field strength; and 4) magnets can be made with a much smaller footprint and lower costs. The acoustic noise can be further reduced by using radiofrequency (RF) pulses for spatial encoding [8-11] which becomes much more feasible at low field whereas it is usually limited by SAR at high field MRI [12].Accepting a lower main field also allows for the design of magnets with bigger bore sizes and/or open magnet designs, both of which can be helpful for people who suffer from claustrophobia in conventional high field MRI systems. An electromagnet system consisting of two elements with permanent magnets has been previously proposed [2]. Unlike typical MRI systems, this system generates inhomogeneous Bo magnetic field. Inhomogeneity of the B0 field makes spin dephasing faster, leading to rapid loss of coherence at the polarizing field strength. However, to mitigate the inhomogeneous Bo field, the design utilizes an electromagnet the main field is ramp-able, allowing polarization to take place at high field (thereby maximizing signal), while signal acquisition is performed at low field [3-5]. In the presence of any gradient, including those that arise from non-uniform B0 fields, selective spin excitation can be performed in a manner identical to that using slice select gradient coils by adjusting both resonance frequency and bandwidth of the RF pulse [2].However, at low fields, the recovery time for RF coils is long because of the narrow bandwidth of the coil. It is proportional to the quality factor of the coil (Q-factor), and inversely proportional to the resonance frequency [13]. After switching off the RF transmit pulse, stored energy in the coil during transmit takes some time to de-energize, known as the coil ring-down time [14]. Since transmit RF power is much higher than receive power (typically eight orders of magnitude), signal detection is impossible during the ring-down time. In addition, the RF coil’s recovery time also affects the rise time of the pulse [13]. For RF pulses of insufficient duration, long ring down times and slow rise times may not allow for the desired shape of the waveform, thereby affecting the applied flip angle. For example, in a spin echo sequence this would mean that the 90° and 180° pulses could not be applied appropriately for square RF pulses. The ring-down time can be shortened by either using a Q-switch [15-18], or through the application of a damping pulse which is 180° out-of-phase with main pulse [13, 18]. Similarly, the rise time can be shortened by applying pre-emphasis pulses with greater amplitude [13, 18].Since the frequency profile of a sinc pulse is rectangular, it is often used for slice-selective excitation [19]. However, unlike with a rectangular pulse, the amplitude of the sinc pulse is not constant. When bandwidth of the sinc pulse is wide, its amplitude changes quickly. As a result, the actual pulse applied to the RF coil is often distorted and delayed. Waveform distortion is usually insignificant for the sinc pulse with narrow bandwidth but there can still be a delay in the applied RF pulse. The input pulse needed to compensate the sinc pulse can be calculated using the Laplace-transform of the impulse response function [20]. In this work, only time domain analysis was adapted to calculate the input pulse.We calculated the duration of pre-emphasis and post-damping pulses for a given amplitude using the measured Q-factor. We also demonstrate a method for calculating an input pulse consisting of a series of square pulses designed to compensate the sinc pulse. For the calculation of input pulses for compensating both square and sinc pulses, Q-factor was measured from ring-down time [21]. The waveforms of the compensated RF pulses were measured under different conditions and compared to the original RF pulses. MR signals were acquired by applying compensated pulse using a low field magnet with a non-uniform B0 field and compared with the signal acquired by uncompensated pulse.
Theory
A. Square pulse
An RF coil can be modeled as an RLC series circuit and energy considerations in RLC circuit lead to a second-order differential equation:
where L, R, C, i(t), and V(t) are inductance, resistance, capacitance, the time-varying current, and applied time-varying voltage, respectively. When a square pulse (Fig 1a) is applied to an RF coil at low frequency, the amplitude of the resultant transmitted RF pulse is shown in Fig 1(b) demonstrating a slow rise time rather than a rapid onset box-car shape. The amplitude of the actual pulse during τ1 is given by the solution of Eq 1.
Where A1, ω, and Q are equilibrium amplitude of the RF signal, the resonance frequency, and the Q-factor of the RF coil, respectively. The A, and Q satisfy the formulas below.
where Z (Z = R + j(ωL − 1⁄ωC)) is the total impedance of the RF coil. After switching off the input to the RF coil, there is also a long ring-down time before the signal dissipates. The amplitude of the RF signal (A) after switching off the pulse is given as the solution of Eq 1 with the condition of V(t) = 0.
Where A is amplitude of the RF signal when the pulse is switched off at the time τ1. The RF ring-down time is reduced by applying additional square pulses that are 180° out-of-phase as shown in Fig 1c and 1d [13, 19]. If τ1 is long enough, A reaches equilibrium (A) and the equilibrium amplitude is proportional to the applied voltage (V). For the given voltages for square pulse (V) and damping pulse (V), time (τ2) to zero amplitude can be calculated by the below equation.
where t = 0. Thus, RF current is eliminated at time τ1 + τ2 13].
Fig 1
Three different RF pulses applied in the power amplifier (left column) and the actual signal generated by the RF coil (right column).
(a) A square pulse and (b) its actual signal on the RF coil, (c) a square pulse with a post-pulse-damping lobe and (d) its actual signal on the RF coil, and (e) a square pulse with both pre-emphasis and a damping pulse and (f) its actual signal on the RF coil.
Three different RF pulses applied in the power amplifier (left column) and the actual signal generated by the RF coil (right column).
(a) A square pulse and (b) its actual signal on the RF coil, (c) a square pulse with a post-pulse-damping lobe and (d) its actual signal on the RF coil, and (e) a square pulse with both pre-emphasis and a damping pulse and (f) its actual signal on the RF coil.However, with relatively short τ1, A does not reach equilibrium and the time required to reach zero is rewritten asLikewise, the transition time is shortened by applying pre-emphasis pulse with greater amplitude (V s V) as shown in Fig 1e and 1f. τ0 is given as a function of V and V.
B. Sinc pulse
The sinc pulse differs from the square pulse above in that the amplitude of the sinc pulse is not constant. But it too, can be compensated using multiple square pulses. The number of samples (N) in the signal can be defined as N = 4iN0 + 1, where N indicates the number of zero crossings on each side of the center before truncation, and i is integer. Then, sampled points include all peaks and zero crossing points. For example, for the sinc with N = 3 and i = 1, N is equal to 13 and the peak and zero amplitude points are included in sampled points as shown in Fig 2.
Fig 2
Example of the sinc pulse (N = 3, i = 1, N = 13).
When a square pulse is applied with peak voltage of V which satisfies the equation below, the amplitude of the actual signal is A at t (t = n∆t).If a square RF pulse is not applied during the time interval between t and t, the amplitude of the actual signal decays and can be calculated using Eq 5 at t. In order to make the amplitude to A at t, a square pulse with peak voltage of V should be applied to the RF coil. V is defined as:Similarly, the peak voltage (Vn) of the square pulse needed to compensate for the nth sampled signal is given by the following general formAs shown in Fig 3, smoother sinc pulses can be generated by increasing the number of samples, N.
Fig 3
Input RF pulses (left column) and the signal generated by the RF coil (right column) for sinc pulse (top row), calculated pulses with N = 13 (second row), N = 25 (third row), N = 37 (bottom row).
C. Q-factor measurement
As shown in the equations above, the Q-factor of the RF coil is needed to design compensation RF pulses. Q-factor is alternatively defined as the ratio of a resonance frequency to its bandwidth and the 3 dB bandwidth is defined as the frequency at which the power level of the signal decreases by half (3 dB) from its maximum value. Q-factor of the RF coil can be measured either using two decoupled sniffer loops (Fig 4a and 4b) [22] or calculated from the return loss of the RF coil (Fig 4c and 4d) [23]. When it is measured using two decoupled sniffer loops, both low and high cutoff frequency are measured at 3 dB down point from the peak. On the other hand, for the return loss method, those are measured at -3 dB points. This is because the half power of the signal is reflected at the –3 dB points while no signal is reflected at the peak with perfect matching condition. The Q-factor of the RF coil is given as
where f, f, and f are resonance frequency, lower cut-off frequency, and upper cut-off frequency, respectively.
Fig 4
Three different methods for measuring Q-factor.
(a) Diagram using two decoupled sniffer loops and (b) example insertion loss between two sniffer loops. (c) Diagram to measure return loss of the RF coil and (d) example return loss of the RF coil. (e) Diagram to measure ring-down time and (f) example waveform of the applied RF pulse.
Three different methods for measuring Q-factor.
(a) Diagram using two decoupled sniffer loops and (b) example insertion loss between two sniffer loops. (c) Diagram to measure return loss of the RF coil and (d) example return loss of the RF coil. (e) Diagram to measure ring-down time and (f) example waveform of the applied RF pulse.The distance between the coil and the sniffer loops affects the measurement. Similarly, the matching condition of the RF coil can affect the Q-factor when measured from the return loss of the coil. However, Q-factor of the coil can be measured from the ring-down time (Fig 4e and 4f) [21]. As shown in Fig 8f, the amplitude of the RF signal decays after switching off the RF pulse, and the rate of this decay is given by Eq 5. By counting how many cycles it takes to halve the amplitude (N), the Q-factor can be calculated as [21]:
Methods
All experiments were carried out on an open, table-top, MRI scanner [2]. Fig 5a shows the schematic representation of the major system on scanner and the major interconnections.
Fig 5
Table-top MRI system.
(a) Schematic diagram of an open, table-top, MRI system with photos of (b) the RF solenoid coil and (c) the magnet.
Table-top MRI system.
(a) Schematic diagram of an open, table-top, MRI system with photos of (b) the RF solenoid coil and (c) the magnet.This MRI system consists of two main components: a two-element electromagnet (Fig 5c), and an RF system. Since the electromagnet generates a non-uniform polarizing B0 field, slice selective excitation (exciting nonuniform isofrequency bands) can be performed without using gradient coils. The maximum B0 field was 300 mT at the bottom of the magnet, and 200 mT at a height of 10 cm. The RF solenoid coil is tuned at 1 MHz (23.4 mT) and generates an alternating B1 magnetic field which is perpendicular to B0 (Fig 5b). The RF coil also detects the NMR signal created by the sample. The RF pulse was generated by a spectrometer (Redstone, Tecmag Inc., TX, USA), and amplified using an RF power amplifier (AR 0.5-15-1E-3-3C, CPC, NY, USA). A band pass filter built in-house was used to eliminate any harmonics in the applied RF pulse. A passive T/R switch with crossed-diode pair was also used.A 40 μs square pulse with an amplitude of 80.2 mV was applied to the RF coil, and the resulting signal measured using a sniffer loop connected to an oscilloscope (Picoscope 4824, Pico technology, UK). From the ring-down signal of the measured pulse, the Q-factor was calculated using Eq 14. Using the Q-factor measured from ring-down time, τ0, and τ2 of the compensation pulse were calculated. The compensated square pulses were then applied to the RF coil, and the resulting signal was also measured using an oscilloscope.Sinc pulses with different bandwidth (30 kHz, 90 kHz, and 150 kHz) were also applied to the RF coil. The actual signal and frequency spectrums were measured using an oscilloscope. The amplitude and duration for the compensated sinc pulse were calculated for different N (13, 25, and 37) using Python (Python Software Foundation). Then, compensated sinc pulses were applied to the RF coil and the actual pulse frequency spectrums were measured and compared using a sniffer loop connected to an oscilloscope. For this purpose, relative error was calculated using Eq 15 by dividing the total sum of difference between magnitude and mean of the magnitude through the total points within the bandwidth.
Phantom experiments
1H echo trains of a uniform cylindrical phantom (110 mm length and 70 mm diameter, distilled water) were obtained using compensated square and sinc pulses. Fig 6 shows the diagram of applied sequence.
Fig 6
A spin-echo sequence with pre-polarization.
All the scan parameters used in this study are described in Table 1. For the square pulse experiment, 20 μs and 40 μs of pulse duration was applied for 90° and 180° pulses, respectively. For the scan with compensated sinc pulse, a pulse with double amplitude of 90° pulse was applied as the 180° pulse. The SNR of the NMR signal was also compared with and without using of RF pulse distortion compensation. For the SNR calculations, the signal mean (maximum intensity) of the first 20 echoes was divided by the noise standard deviation of the last 20 echoes (the NMR signal at the last 20 echoes had been decayed).
Table 1
NMR signal acquisition parameters used in this study.
Pre-polarization time
Dwell time
Sampling time
TR
TE
number of echoes
Averages
Square pulse
5 s
10 μs
320 μs
~ 6 s
1 ms
1,000
32
Sinc pulse
~ 6.5 s
1.5 ms
1,000
32
Results
Fig 7 shows the measured square input pulse without (a) and with (b) a compensated input.
Fig 7
Measured actual pulses of (a) square pulse input and (b) compensated square pulse input.
Vertical solid lines indicate the input pulse ends.
Measured actual pulses of (a) square pulse input and (b) compensated square pulse input.
Vertical solid lines indicate the input pulse ends.Q-factor was calculated using Fig 7a and found to be 49.83. For the square pulse, the amplitudes and durations of the compensation pulse were found to be 120.3 mV, 40.1 mV, 120.3 mV, 4.68 μs, 40 μs. and 3.32 μs for V, V, V, τ0, τ1, and τ2, respectively.Fig 8 shows the calculated amplitude for compensating the sinc pulses for the different bandwidths.
Fig 8
Sinc pulses (top row) and calculated input pulses for compensation calculated with different N (13 (second row), 25 (third row), and 37 (bottom row)) for different bandwidth (30 kHz (left column), 90 kHz (middle column), and 150 kHz (right column)).
Since the amplitude changes rapidly for the wide band pulse, higher amplitudes are required for the compensation. Fig 9 shows the measured actual pulses when sinc and compensated sinc pulses were applied.
Fig 9
Measured actual pulses (output pulses, vertical dashed lines indicate the input pulse ends).
And frequency spectrums of the sinc pulses (top row) and compensated sinc pulses calculated with N values of 13 (second row), 25 (third row), and 37 (bottom row) and bandwidths of 30 kHz (left column), 90 kHz (middle column), and 150 kHz (right column). The y-axis shows normalized values (from -1 to 1 for waveforms, from 0 to 1 for frequency spectrums).
Measured actual pulses (output pulses, vertical dashed lines indicate the input pulse ends).
And frequency spectrums of the sinc pulses (top row) and compensated sinc pulses calculated with N values of 13 (second row), 25 (third row), and 37 (bottom row) and bandwidths of 30 kHz (left column), 90 kHz (middle column), and 150 kHz (right column). The y-axis shows normalized values (from -1 to 1 for waveforms, from 0 to 1 for frequency spectrums).The long response time of the RF coil also affects the shape of the sinc pulse, especially for broad band pulses. The compensation helps with both reduction of the amplitude after the input pulse end and frequency profile. Table 2 shows the relative errors of the signal magnitude of the frequency spectrum.
Table 2
Relative errors of the signal magnitude of the frequency spectrum.
Bandwidth
Relative error
30 kHz
90 kHz
150 kHz
Sinc pulse
0.33
0.51
0.62
Compensated pulse
NS = 13
0.25
0.23
0.24
NS = 25
0.27
0.23
0.23
NS = 37
0.26
0.23
0.24
Fig 10 shows the 1H NMR signal from a water phantom. The x-axis shows time in microseconds and y axis shows the normalized signal intensity.
Fig 10
1H 25th, 50th, 75th, 100th, and 125th echoes from a train of 1000 echoes, of water obtained using square pulse (top row), compensated square pulse (second row), sinc pulse with 90 kHz bandwidth (third row), and compensated sinc pulse, 90 kHz bandwidth, n = 37 (bottom row).
The uncompensated and the compensated square pulses shown in Fig 8 was used as a 90° excitation pulse followed by a series of 180° refocusing RF pulses (the same pulse doubled in duration) in order to acquire the echo train. Fig 10 (third and bottom rows) shows the 1H NMR signal of water phantom acquired using uncompensated and compensated sinc pulses (the same pulse doubled in amplitude) which was calculated with N = 37. Table 3 summarizes the SNR values of the NMR signals.
Table 3
SNR values measured by using uncompensated and compensated pulses (square and sinc pulses).
SNR
Without distortion compensation
With distortion compensation
Square pulse
9.9
16.2
Sinc pulse
14.0
27.2
Discussion
When the Q-factor of the RF coil is high, the bandwidth of the RF coil can be narrower than the bandwidth of the applied RF pulse. For this reason, especially in low magnetic field MRI (low resonance frequency), short RF pulses can be easily distorted due to the narrow bandwidth of the RF coil. This study demonstrates that compensation pulses (including the required duration and amplitude of such a pulse) can be calculated utilizing the measured Q-factor, and these compensation pulses lead to RF shapes closer to the desired output. The SNR of the NMR signal acquired by compensated RF pulses were compared with those obtained by using uncompensated RF pulses.For the square RF pulse, three square pulses were applied rather than applying a single square RF pulse [13]. While the amplitude of a square pulse only changes once at the beginning and once at the end of the pulse, the signal amplitude of a sinc pulse changes throughout the duration of the pulse. These amplitude changes can cause time delays and waveform distortions as they propagate through the coil. It was demonstrated that the distortion of a sinc pulse can be compensated by using series of square pulses, and that a smoother waveform can be acquired by using more square pulses with short durations. Furthermore, the frequency profiles of the compensated sinc pulses were closer to a square than that of the original sinc pulses especially for the wide bandwidth. The time delay of the pulse was also compensated using this method, which is helpful if a hardware Q-switch is used. If a Q-switch is used without compensation, the end of applied pulse can be truncated due to the time delay. Although only a square pulse and sinc pulse, the latter with N0 = 1 to 3, were considered here, the suggested method can be used for any shape of transmit pulses such as sinc1, sinc5, and Gaussian pulses. As demonstrated here both the square and the sinc pulse shapes can be improved by calculating compensatory input pulses using Q-factor. Thus, it is important to accurately measure the Q-factor. The Q-factor of the RF coil is usually measured either using two decoupled sniffer loops [22] or from a measurement of the return loss [23] of the RF coil. However, the Q-factor measured from the ring-down time is different from the Q-factor measured using vector network analyzer since other factors such as the distance between the RF coil and sniffer loops and the matching condition of the RF coil can influence the results. Moreover, RF recovery time is also affected by the RF chain (filter, T/R switch, and co-axial cable) in addition to the RF coil itself. This Q-factor measurement worked well for generating RF compensation pulses here. For the higher precision of Q-factor, N in Eq 14 can be estimated using decimal rather than using only integer values.Since the Q value is directly related to the bandwidth, the signal distortion had a greater effect when a wide bandwidth sin pulse was applied. This was confirmed by the waveform, especially the wider the bandwidth, the longer the ring down time. It was confirmed that this long ring-down signal was effectively reduced through the proposed method. Through the frequency profile, it is confirmed that the wide bandwidth signal is more distorted, and the proposed method has a correction effect. It was also confirmed through the calculated relative error.By applying compensated pulses (to both square and sinc pulses), multi-echo spin echo NMR signal were acquired on inhomogeneous magnetic field. A CPMG spin echo sequence was used to acquire 1000 spin-echoes at low magnetic field (1 MHz) after the spins had been pre-polarized at approximately 300 mT (12.77 MHz) [3-5] in a novel field cycling open MR system. As shown in Fig 9, the output pulse can excite more spins within the slice to the desired flip angle. As a result, the SNR of the echo signals obtained using compensated pulses were 61.1% and 51.5% higher relative to that obtained using uncompensated pulses for square pulse and sinc pulse, respectively, indicating substantial improvement with this compensation strategy.In conclusion, we have introduced an approach to low field imaging for measuring the Q-factor of an RF coil, in order to derive a set of compensation pulses to reduce distortions in the shape of the excitation profile. For the accurate calculation of the compensating pulses, the Q-factor was measured from the ring-down time of the coil. It was demonstrated that both square and sinc pulse shapes could be improved by applying compensating RF pulses at the RF input. Echo trains 1000 echoes long were acquired using these pulses at low readout field B0. Improvement in the echo SNR was demonstrated by using RF pulse compensation for both square and sinc pulses.(PY)Click here for additional data file.24 May 2022
PONE-D-22-10094
Practical method for RF pulse distortion compensation using multiple square pulses for low-field MRI
PLOS ONE
Dear Dr. Constable,Thank you for submitting your manuscript to PLOS ONE. After careful consideration, we feel that it has merit but does not fully meet PLOS ONE’s publication criteria as it currently stands. Therefore, we invite you to submit a revised version of the manuscript that addresses the points raised during the review process.Please submit your revised manuscript by Jul 08 2022 11:59PM. If you will need more time than this to complete your revisions, please reply to this message or contact the journal office at plosone@plos.org. When you're ready to submit your revision, log on to https://www.editorialmanager.com/pone/ and select the 'Submissions Needing Revision' folder to locate your manuscript file.Please include the following items when submitting your revised manuscript:
A rebuttal letter that responds to each point raised by the academic editor and reviewer(s). You should upload this letter as a separate file labeled 'Response to Reviewers'.A marked-up copy of your manuscript that highlights changes made to the original version. You should upload this as a separate file labeled 'Revised Manuscript with Track Changes'.An unmarked version of your revised paper without tracked changes. You should upload this as a separate file labeled 'Manuscript'.If you would like to make changes to your financial disclosure, please include your updated statement in your cover letter. Guidelines for resubmitting your figure files are available below the reviewer comments at the end of this letter.If applicable, we recommend that you deposit your laboratory protocols in protocols.io to enhance the reproducibility of your results. Protocols.io assigns your protocol its own identifier (DOI) so that it can be cited independently in the future. For instructions see: https://journals.plos.org/plosone/s/submission-guidelines#loc-laboratory-protocols. Additionally, PLOS ONE offers an option for publishing peer-reviewed Lab Protocol articles, which describe protocols hosted on protocols.io. Read more information on sharing protocols at https://plos.org/protocols?utm_medium=editorial-email&utm_source=authorletters&utm_campaign=protocols.We look forward to receiving your revised manuscript.Kind regards,Yefeng YaoAcademic EditorPLOS ONEJournal Requirements:When submitting your revision, we need you to address these additional requirements.1. Please ensure that your manuscript meets PLOS ONE's style requirements, including those for file naming. The PLOS ONE style templates can be found athttps://journals.plos.org/plosone/s/file?id=wjVg/PLOSOne_formatting_sample_main_body.pdf and https://journals.plos.org/plosone/s/file?id=ba62/PLOSOne_formatting_sample_title_authors_affiliations.pdf2. Please note that PLOS ONE has specific guidelines on code sharing for submissions in which author-generated code underpins the findings in the manuscript. In these cases, all author-generated code must be made available without restrictions upon publication of the work. Please review our guidelines at https://journals.plos.org/plosone/s/materials-and-software-sharing#loc-sharing-code and ensure that your code is shared in a way that follows best practice and facilitates reproducibility and reuse.3. In your Data Availability statement, you have not specified where the minimal data set underlying the results described in your manuscript can be found. PLOS defines a study's minimal data set as the underlying data used to reach the conclusions drawn in the manuscript and any additional data required to replicate the reported study findings in their entirety. All PLOS journals require that the minimal data set be made fully available. For more information about our data policy, please see http://journals.plos.org/plosone/s/data-availability.Upon re-submitting your revised manuscript, please upload your study’s minimal underlying data set as either Supporting Information files or to a stable, public repository and include the relevant URLs, DOIs, or accession numbers within your revised cover letter. For a list of acceptable repositories, please see http://journals.plos.org/plosone/s/data-availability#loc-recommended-repositories. Any potentially identifying patient information must be fully anonymized.Important: If there are ethical or legal restrictions to sharing your data publicly, please explain these restrictions in detail. Please see our guidelines for more information on what we consider unacceptable restrictions to publicly sharing data: http://journals.plos.org/plosone/s/data-availability#loc-unacceptable-data-access-restrictions. Note that it is not acceptable for the authors to be the sole named individuals responsible for ensuring data access.We will update your Data Availability statement to reflect the information you provide in your cover letter.[Note: HTML markup is below. Please do not edit.]Reviewers' comments:Reviewer's Responses to Questions
Comments to the Author1. Is the manuscript technically sound, and do the data support the conclusions?The manuscript must describe a technically sound piece of scientific research with data that supports the conclusions. Experiments must have been conducted rigorously, with appropriate controls, replication, and sample sizes. The conclusions must be drawn appropriately based on the data presented. Reviewer #1: YesReviewer #2: Yes********** 2. Has the statistical analysis been performed appropriately and rigorously? Reviewer #1: I Don't KnowReviewer #2: Yes********** 3. Have the authors made all data underlying the findings in their manuscript fully available?The PLOS Data policy requires authors to make all data underlying the findings described in their manuscript fully available without restriction, with rare exception (please refer to the Data Availability Statement in the manuscript PDF file). The data should be provided as part of the manuscript or its supporting information, or deposited to a public repository. For example, in addition to summary statistics, the data points behind means, medians and variance measures should be available. If there are restrictions on publicly sharing data—e.g. participant privacy or use of data from a third party—those must be specified. Reviewer #1: YesReviewer #2: Yes********** 4. Is the manuscript presented in an intelligible fashion and written in standard English?PLOS ONE does not copyedit accepted manuscripts, so the language in submitted articles must be clear, correct, and unambiguous. Any typographical or grammatical errors should be corrected at revision, so please note any specific errors here. Reviewer #1: YesReviewer #2: Yes********** 5. Review Comments to the AuthorPlease use the space provided to explain your answers to the questions above. You may also include additional comments for the author, including concerns about dual publication, research ethics, or publication ethics. (Please upload your review as an attachment if it exceeds 20,000 characters) Reviewer #1: This manuscript proposed a method for RF pulse distortion compensation using multiple square pulses for low-field MRI. The authors demonstrated that compensation pulses could be calculated utilizing the measured Q-factor, and these compensation pulses leaded to RF shapes closer to the desired output. The results are interesting and worthy of publication. The manuscript can be improved by addressing the following concerns.(1) The quality of figures in the manuscript should be improved. The resolutions of the figures are too low.(2) I suggest that the authors provide more information for the phantom and more scanning parameters for spin-echo sequence (receiver bandwidth, sampling points, etc.).(3) The method of calculating SNR of signal should be provided.(4) The authors should explain why the SNRs were improved after applying the RF with distortion compensation.(5) I suggest that the authors discuss the advantages of the proposed method for human body low-field MRI. In human body low-field MRI, two coils are used for RF transmission and signal receive, respectively. The duration of RF pulse is usually several milliseconds. Is it necessary to use the proposed method?(6) Minor concernsPage#15, Line#15: 180s ->180° refocused RF pulsesReviewer #2: This work introduces a RF square pulse compensated method by appending two square pulses before and after the RF pulse, with durations of these RF square pulses calculated using the Q-factor. It was also successfully demonstrated that the sinc pulse can be compensated using a series of square pulses. The more number of square pulses were used, the smoother sinc pulse was applied to the RF coil. Echo trains were also acquired in an inhomogeneous B0 field using the compensated RF pulses. In order to enhance the SNR of the echo trains, a pre-polarization pulse was added to the CPMG spin echo sequence. The SNRs of the echo signal acquired using compensated pulses were compared with those of signal obtained with uncompensated pulses and showed significant improvements of 61.1% and 51.5% for the square and sinc shaped pulses respectively.Some questions are as follows:Page 15, line 4, table 2, why the sinc pulse deviation for larger radiofrequency is bigger comparing lower frequency condition, which seems contradictory with equation 2.In fig.1, how to control the phase of the pre-emphasis pulse(figure 1c)&e)) to keep the phase of the pulse constant?In fig.8, why the sinc pulse in Ns =25&37 are asymmetric after the compensation? while this also appears in fig.9.In fig.10, the axis in the figure are vague, please make it more readable.This work is interesting and applicable in low-field NMR/MRI.********** 6. PLOS authors have the option to publish the peer review history of their article (what does this mean?). If published, this will include your full peer review and any attached files.If you choose “no”, your identity will remain anonymous but your review may still be made public.Do you want your identity to be public for this peer review? For information about this choice, including consent withdrawal, please see our Privacy Policy. Reviewer #1: Yes: Jianqi LiReviewer #2: Yes: Xiaodong YANG[NOTE: If reviewer comments were submitted as an attachment file, they will be attached to this email and accessible via the submission site. Please log into your account, locate the manuscript record, and check for the action link "View Attachments". If this link does not appear, there are no attachment files.]While revising your submission, please upload your figure files to the Preflight Analysis and Conversion Engine (PACE) digital diagnostic tool, https://pacev2.apexcovantage.com/. PACE helps ensure that figures meet PLOS requirements. To use PACE, you must first register as a user. Registration is free. Then, login and navigate to the UPLOAD tab, where you will find detailed instructions on how to use the tool. If you encounter any issues or have any questions when using PACE, please email PLOS at figures@plos.org. Please note that Supporting Information files do not need this step.27 Jun 2022Reviewer #1: This manuscript proposed a method for RF pulse distortion compensation using multiple square pulses for low-field MRI. The authors demonstrated that compensation pulses could be calculated utilizing the measured Q-factor, and these compensation pulses leaded to RF shapes closer to the desired output. The results are interesting and worthy of publication. The manuscript can be improved by addressing the following concerns.(1) The quality of figures in the manuscript should be improved. The resolutions of the figures are too low.We have regenerated figures (Fig. 2, 4, and 10). The resolution of the figures in the pdf file looks low. However, high resolution images are available by clicking the links (downloading the figures) in the manuscript document.(2) I suggest that the authors provide more information for the phantom and more scanning parameters for spin-echo sequence (receiver bandwidth, sampling points, etc.).We have added details about phantom as:“H echo trains of a uniform cylinder phantom (110 mm length and 70 mm diameter, distilled water) were obtained using compensated square and sinc pulses.”We have also added dwell time (10 us) and sampling time (320 us) to table 1, so receiver bandwidth and sampling points are provided.(3) The method of calculating SNR of signal should be provided.We have added details about SNR calculation as:“For the SNR calculations, the signal mean (maximum intensity) of the first 20 echoes was divided by the noise standard deviation of the last 20 echoes (the NMR signal at the last 20 echoes had decayed to essentially zero).”(4) The authors should explain why the SNRs were improved after applying the RF with distortion compensation.When a distorted signal is applied, some spins within the slice are excited with lower flip angle than expected. This can be corrected through the proposed compensation method, resulting in increased SNR. We have added details in discussions as:As shown in Fig. 8, the calculated input pulses for distortion compensation require higher power. However, as shown in Fig. 9, the output pulse can excite more spins within the slice to the desired flip angle. As a result, the SNR of the echo signals obtained using compensated pulses were 61.1 % and 51.5 % higher relative to that obtained using uncompensated pulses for square pulse and sinc pulse, respectively, indicating substantial improvement with this compensation strategy.(5) I suggest that the authors discuss the advantages of the proposed method for human body low-field MRI. In human body low-field MRI, two coils are used for RF transmission and signal receive, respectively. The duration of RF pulse is usually several milliseconds. Is it necessary to use the proposed method?RF distortion may not be a problem in human body low-field MRI with several milliseconds duration of RF pulse. However, it depends on not only the field strength (carrier frequency) and the pulse duration (BW of the pulse), but also bandwidth (Q) of the RF coil as described in Eq. 2 and Eq. 5. We have modified a sentence in the introduction section as:When the Q-factor of the RF coil is high, the bandwidth of the RF coil can be narrower than the bandwidth of the applied RF pulse. For this reason, especially in low magnetic field MRI (low resonance frequency), short RF pulses can be easily distorted due to the narrow bandwidth of the RF coil (long RF coil recovery times).(6) Minor concernsPage#15, Line#15: 180s ->180° refocused RF pulsesThank you for your careful proofreading. We have made the suggested change.Reviewer #2: This work introduces a RF square pulse compensated method by appending two square pulses before and after the RF pulse, with durations of these RF square pulses calculated using the Q-factor. It was also successfully demonstrated that the sinc pulse can be compensated using a series of square pulses. The more number of square pulses were used, the smoother sinc pulse was applied to the RF coil. Echo trains were also acquired in an inhomogeneous B0 field using the compensated RF pulses. In order to enhance the SNR of the echo trains, a pre-polarization pulse was added to the CPMG spin echo sequence. The SNRs of the echo signal acquired using compensated pulses were compared with those of signal obtained with uncompensated pulses and showed significant improvements of 61.1% and 51.5% for the square and sinc shaped pulses respectively.Some questions are as follows:Page 15, line 4, table 2, why the sinc pulse deviation for larger radiofrequency is bigger comparing lower frequency condition, which seems contradictory with equation 2.Equation 2 describes the rise time of the applied pulse. This shows that the signal takes time to increase in magnitude. The wide bandwidth of a sine pulse means that the amplitude of the signal changes rapidly. The wider the bandwidth, the greater the signal distortion, because the change in signal amplitude of a wide bandwidth sinc pulse is shorter than the time required for the change in signal amplitude.To avoid confusion, we added that frequency represents bandwidth in Table 2.In fig.1, how to control the phase of the pre-emphasis pulse(figure 1c)&e)) to keep the phase of the pulse constant?Ring-down is not affecting the carrier frequency but the envelope of the pulse. However, during the τ2, a square pulse with 180º out-of-phase is applied to cancel out the ring-down signal.In fig.8, why the sinc pulse in Ns =25&37 are asymmetric after the compensation? while this also appears in fig.9.Fig. 8 shows the calculated input pulse for compensation. By applying these calculated asymmetric input pulses to the RF coil, compensated sinc pulses were generated. When the input pulse is a sinc (symmetric) shape, the applied pulse can be asymmetrically distorted as shown in Fig. 9 first row.To avoid confusion, we modified the figure caption of Fig. 9 as:Measured actual pulse (vertical dashed lines indicate the input pulse ends.) → Measured actual pulse (output pulses, vertical dashed lines indicate the input pulse ends.)In fig.10, the axis in the figure are vague, please make it more readable.Thank you for your careful proofreading.We have adjusted the figure size.This work is interesting and applicable in low-field NMR/MRI.Submitted filename: Response to Reviewers.docxClick here for additional data file.9 Aug 2022Practical method for RF pulse distortion compensation using multiple square pulses for low-field MRIPONE-D-22-10094R1Dear Dr. Constable,We’re pleased to inform you that your manuscript has been judged scientifically suitable for publication and will be formally accepted for publication once it meets all outstanding technical requirements.Within one week, you’ll receive an e-mail detailing the required amendments. When these have been addressed, you’ll receive a formal acceptance letter and your manuscript will be scheduled for publication.An invoice for payment will follow shortly after the formal acceptance. To ensure an efficient process, please log into Editorial Manager at http://www.editorialmanager.com/pone/, click the 'Update My Information' link at the top of the page, and double check that your user information is up-to-date. If you have any billing related questions, please contact our Author Billing department directly at authorbilling@plos.org.If your institution or institutions have a press office, please notify them about your upcoming paper to help maximize its impact. If they’ll be preparing press materials, please inform our press team as soon as possible -- no later than 48 hours after receiving the formal acceptance. Your manuscript will remain under strict press embargo until 2 pm Eastern Time on the date of publication. For more information, please contact onepress@plos.org.Kind regards,Yefeng YaoAcademic EditorPLOS ONEAdditional Editor Comments (optional):Reviewers' comments:Reviewer's Responses to Questions
Comments to the Author1. If the authors have adequately addressed your comments raised in a previous round of review and you feel that this manuscript is now acceptable for publication, you may indicate that here to bypass the “Comments to the Author” section, enter your conflict of interest statement in the “Confidential to Editor” section, and submit your "Accept" recommendation. Reviewer #1: All comments have been addressedReviewer #2: All comments have been addressed********** 2. Is the manuscript technically sound, and do the data support the conclusions?The manuscript must describe a technically sound piece of scientific research with data that supports the conclusions. Experiments must have been conducted rigorously, with appropriate controls, replication, and sample sizes. The conclusions must be drawn appropriately based on the data presented. Reviewer #1: YesReviewer #2: Yes********** 3. Has the statistical analysis been performed appropriately and rigorously? Reviewer #1: YesReviewer #2: Yes********** 4. Have the authors made all data underlying the findings in their manuscript fully available?The PLOS Data policy requires authors to make all data underlying the findings described in their manuscript fully available without restriction, with rare exception (please refer to the Data Availability Statement in the manuscript PDF file). The data should be provided as part of the manuscript or its supporting information, or deposited to a public repository. For example, in addition to summary statistics, the data points behind means, medians and variance measures should be available. If there are restrictions on publicly sharing data—e.g. participant privacy or use of data from a third party—those must be specified. Reviewer #1: YesReviewer #2: Yes********** 5. Is the manuscript presented in an intelligible fashion and written in standard English?PLOS ONE does not copyedit accepted manuscripts, so the language in submitted articles must be clear, correct, and unambiguous. Any typographical or grammatical errors should be corrected at revision, so please note any specific errors here. Reviewer #1: YesReviewer #2: Yes********** 6. Review Comments to the AuthorPlease use the space provided to explain your answers to the questions above. You may also include additional comments for the author, including concerns about dual publication, research ethics, or publication ethics. (Please upload your review as an attachment if it exceeds 20,000 characters) Reviewer #1: (No Response)Reviewer #2: The authors have well addressed my questions, and the current version is fine for me. The method and result in the paper is attractive for the readers, i believe.********** 7. PLOS authors have the option to publish the peer review history of their article (what does this mean?). If published, this will include your full peer review and any attached files.If you choose “no”, your identity will remain anonymous but your review may still be made public.Do you want your identity to be public for this peer review? For information about this choice, including consent withdrawal, please see our Privacy Policy. Reviewer #1: Yes: Jianqi LiReviewer #2: No**********7 Sep 2022PONE-D-22-10094R1Practical method for RF pulse distortion compensation using multiple square pulses for low-field MRIDear Dr. Constable:I'm pleased to inform you that your manuscript has been deemed suitable for publication in PLOS ONE. Congratulations! Your manuscript is now with our production department.If your institution or institutions have a press office, please let them know about your upcoming paper now to help maximize its impact. If they'll be preparing press materials, please inform our press team within the next 48 hours. Your manuscript will remain under strict press embargo until 2 pm Eastern Time on the date of publication. For more information please contact onepress@plos.org.If we can help with anything else, please email us at plosone@plos.org.Thank you for submitting your work to PLOS ONE and supporting open access.Kind regards,PLOS ONE Editorial Office Staffon behalf ofDr. Yefeng YaoAcademic EditorPLOS ONE
Authors: Ralf Kartäusch; Toni Driessle; Thomas Kampf; Thomas Christian Basse-Lüsebrink; Uvo Christoph Hoelscher; Peter Michael Jakob; Florian Fidler; Xavier Helluy Journal: MAGMA Date: 2013-11-20 Impact factor: 2.310