Wenzhao Sun1, Bin Wang1, Bo Qiu1, Jian Liang1, Weihao Xie1, Xiaowu Deng1, Zhenyu Qi1.
Abstract
OBJECTIVE: To assess the breast dose during a routine thoracic cone-beam CT (CBCT) check with the efforts to explore the possible dose reduction strategy.Entities:
Keywords: breast dose; cone beam computed tomography (CBCT); image registration; metal oxide semiconductor field-effect transistor (MOSFET); radiotherapy
Mesh:
Year: 2017 PMID: 28423624 PMCID: PMC5386753 DOI: 10.18632/oncotarget.15555
Source DB: PubMed Journal: Oncotarget ISSN: 1949-2553
Results and parameters for MOSFET calibration
| Peak Voltage (kV) | Scanning configuration | HVLa (mm Al) | Effective Energy (keV) | (μen/ρ)w,air | BW | Pstem | Sensitivity (mV/cGy) |
|---|---|---|---|---|---|---|---|
| 120 | F0+ S20 | 8.1 | 56.5 | 1.043 | 1.502 | 1 | 15.77 |
| 120 | F1+ S20 | 9.0 | 60.0 | 1.047 | 1.509 | 1 | 14.88 |
a Half value layer.
Breast surface doses of different breast volumes for thoracic CBCT scans (A, B, C, and D corresponds to the breast volume of 200, 400, 600, and 1200 cc respectively)
| Filtration | A (200 cc) | B (400 cc) | C (600 cc) | D (1200 cc) |
|---|---|---|---|---|
| F0 | 19.44±1.05 | 20.29±0.00 | 19.44±1.05 | 20.92±0.00 |
| F1 | 12.17±0.00 | 12.17±0.00 | 11.33±0.00 | 10.91±0.00 |
Protocol B with or without filter was used. The breast dose was presented in the form of Dmean±SD mGy.
Comparison of radiation doses delivered to ipsilateral breast vs. contralateral breast for full arc and partial arc CBCT
| Filtration | Protocol B | Protocol C | ||
|---|---|---|---|---|
| Ipsilateral breast | Contralateral breast | Ipsilateral breast | Contralateral breast | |
| F0 | 19.44±1.05 | 19.44±1.05 | 17.97±0.40 | 15.43±1.43 |
| F1 | 12.17±0.00 | 12.17±0.00 | 11.33±0.00 | 6.30±0.69 |
The breast doses were measured using ART phantom with 200 cc artificial breast. The measured dose was presented in the form of Dmean±SD mGy.
Results of uniformity and LCV tests for thoracic CBCT images
| Full Arc (F0M20) | Full Arc (F1M20) | Full Arc (F0S20) | Full Arc (F1S20) | |
|---|---|---|---|---|
| Non-uniformity(%) | 2.38 | 5.16 | 2.77 | 1.09 |
| LCV(%) | 1.16 | 1.11 | 1.24 | 0.64 |
Comparison of registration results between full arc CBCT and partial arc CBCT for thoracic cancer patient positioning verification
| Nominal shifted distance (cm) | Protocol B (full arc) | Protocol C (partial arc) | ||||
|---|---|---|---|---|---|---|
| LAT | LONG | VERT | LAT | LONG | VERT | |
| Normal setup (0) | 0.04 | 0.14 | 0.04 | 0.06 | 0.16 | 0.06 |
| Anterior 1cm | 0.03 | −0.11 | −1.05 | 0.03 | −0.05 | −1.07 |
| Superior 1cm | −0.01 | −0.96 | 0.00 | −0.02 | −0.97 | −0.01 |
| Left 1cm | −0.96 | −0.04 | 0.00 | −0.96 | −0.12 | 0.00 |
The systematic setup and registration error was subtracted from the registration results.
Scanning parameters for thorax CBCT. (Protocol A was used for MOSFET calibration, protocol B was used for nominal thorax CBCT scan, and protocol C was used for partial-arc scan investigation.)
| Protocol A | Protocol B | Protocol C | |
|---|---|---|---|
| Tube voltage [kV] | 120 | 120 | 120 |
| Nonminal mA/frame | 25 | 25 | 25 |
| Nominal mS/frame | 40 | 40 | 40 |
| Frames | 650 | 650 | 397 |
| Total mAs | 650 | 650 | 397 |
| FOV [mm] | S | M | M |
| Collimator | 20 | 20 | 20 |
| Filter | F0/F1 | F0/F1 | F0/F1 |
| Gantry Rotation | stationary x-ray tube (0°) | rotational x-ray tube (−180°~180°) | rotational x-ray tube (−110°~110°) |
Figure 1Setup of breast CBCT dose measurement using a female-configured anthropomorphic phantom and MOSFET detectors
Figure 2(A) Catphan 503 phantom, (B) images of 486 module used for image uniformity test and (C) 404 module used for low contrast visibility evaluation