Sven G Hyberts1, Scott A Robson1, Gerhard Wagner2.
Abstract
Non-Uniform Sampling has the potential to exploit the optimal resolution of high-field NMR instruments. This is not possible in 3D and 4D NMR experiments when using traditional uniform sampling due to the long overall measurement time. Nominally, uniformly sampled time domain data acquired to a maximum evolution time tmax can be extended to high resolution via a virtual maximum evolution time t*max while extrapolating with linear prediction or iterative soft thresholding (IST). At the high resolution obtainable with extrapolation of US data, however, the accuracy of peak positions is compromised as observed when comparing inter- and intra-residue peaks in a 3D HNCA experiment. However, the accuracy of peak positions is largely improved by spreading the same number of acquired time domain data points non-uniformly over a larger evolution time to an optimal tmax followed by extrapolation to a total t*max and processing the data with an appropriate reconstruction method, such as hmsIST. To explore the optimum value of experimentally measured tmax to be reached non-uniformly with a given number of sampling points we have created test situations of time-equivalent experiments and evaluate sensitivity and accuracy of peak positions. Here we use signal-to-maximum-noise ratio as the decisive measure of sensitivity. We find that both sensitivity and resolution are optimal when PoissonGap sampling to a tmax of about ½*T2*. Digital resolution is further enhanced by extrapolating the range of acquired time domain data to 2*T2* but without measuring experimental points beyond ½*T2*.Entities:
Keywords: Iterative soft threshold reconstruction; NUS; Non-uniform sampling; Protein backbone chemical shift assignments; Reduced time multidimensional NMR spectroscopy
Mesh:
Year: 2017 PMID: 28332026 PMCID: PMC5614452 DOI: 10.1007/s10858-017-0103-z
Source DB: PubMed Journal: J Biomol NMR ISSN: 0925-2738 Impact factor: 2.835