| Literature DB >> 28320451 |
Joshua Guag1, Bisrat Addissie2, Donald Witters2.
Abstract
BACKGROUND: There have been concerns that Electromagnetic security systems such as walk-through metal detectors (WTMDs) can potentially cause electromagnetic interference (EMI) in certain active medical devices including implantable cardiac pacemakers and implantable neurostimulators. Incidents of EMI between WTMDs and active medical devices also known as personal medical electronic devices (PMED) continue to be reported. This paper reports on emission measurements of sample WTMDs and testing of 20 PMEDs in a WTMD simulation system.Entities:
Keywords: EMC; EMI; Electromagnetic compatibility; Electromagnetic interference; Medical device; Metal detector; Personal medical electronic devices; Security system; Walk through metal detector
Mesh:
Substances:
Year: 2017 PMID: 28320451 PMCID: PMC5359895 DOI: 10.1186/s12938-017-0328-9
Source DB: PubMed Journal: Biomed Eng Online ISSN: 1475-925X Impact factor: 2.819
Fig. 1WTMD field mapping area. Left front view, right top view
Magnetic field characteristics from the sample WTMDs
| WTMD | Waveform | Frequency/pulse repetition rate | Pulse width (µS) | Rise/fall time (µS) | Absolute maximum field measured (A/m RMS) |
|---|---|---|---|---|---|
| WTMD-A | Multiple CW | 3–14 kHz | N/A | N/A | 21.3 |
| WTMD-B | Pulsed | 200–250 Hz | 100–200 | 5–50 | 21.1 |
| WTMD-C | 27.9 |
Fig. 2Example test setting with WTMD simulator
Fig. 3Block diagram of example test setting for cardiac devices
Maximum field strength of WTMD simulator generated for each waveform from each WTMD
| WTMD source | Waveform | Maximum field strength generated in the WMTD simulator (A/m RMS) |
|---|---|---|
| WTMD-A | A1 | 22.9 |
| A2 | 80 | |
| A3 | 79.7 | |
| WTMD-B | B1 | 24.9 |
| B2 | 45.9 | |
| WTMD-C | C1 | 30.3 |
| C2 | 74.9 |
Implantable cardiac pacemaker device setting
| Device | Mode | Lead configuration | Sensitivity range (maximum) | Sensitivity range (nominal) |
|---|---|---|---|---|
| P1 | DDD | Bipolar/unipolar | Atrial: 0.1–0.6 mV | Atrial: 0.4–1 mV |
| P2 | VVI | Bipolar/unipolar | ||
| P3 | VVI | Bipolar/unipolar | ||
| P4 | DDD | Bipolar/unipolar | ||
| P5 | DDD | Bipolar | ||
| P6 | DDD | Bipolar/unipolar |
Implantable cardioverter defibrillator device setting
| Device | Mode | Lead configuration | Sensitivity range (maximum) |
|---|---|---|---|
| D1 | DDD | Bipolar | Atrial: 0.1–0.3 mV |
| D2 | VVI | Bipolar | |
| D3 | DDD | Bipolar | |
| D4 | DDD | Bipolar | |
| D5 | DDD | Bipolar | |
| D6 | VVI | Bipolar |
Fig. 4Example magnetic field emission measurement from WTMD-A (A/m RMS). Magnetic field value of vector magnitude over entire mapping area. Left three slices represent center vertical (Y–Z) plane, 1 m height horizontal (X–Y) plane, and vertical (X–Z) plane parallel and 8 cm outside away from the transmitting pylon. Right vertical (X–Z) plane parallel and 8 cm inside away from the transmitting pylon
EMI observed during the testing
| Device type | EMI observed |
|---|---|
| Implantable cardiac pacemakers | Pacing inhibition |
| Pacing induced | |
| Pacing rate change | |
| Atrial/ventricular pacing interval changes | |
| Implantable neurostimulators | Pulse inhibition |
| Pulse rate changes | |
| Pulse waveform changes | |
| Implantable cardioverter defibrillators (ICDs) | No EMI |
| Body worn insulin pumps | No EMI |
Fig. 5Scatter points are the threshold for EMI (if observed) for six implantable cardiac pacemakers (P1–P6) tested across the seven different waveforms in WTMD simulator. For each waveform, PMEDs were exposed to maximum field generated in WTMD simulator (see Table 1) in order to investigate any potential EMI. Bars represent measured exposure field strengths in actual WTMDs. The “averaged max measured” fields are averaged over a pacemaker lead loop area of 225 cm2
Fig. 6Scatter points in red are the threshold for EMI for the implantable neurostimulator (N5) tested across the seven different waveforms in WTMD simulator. For each waveform, PMEDs were exposed to maximum field generated in WTMD simulator (see Table 1) in order to investigate any potential EMI. Bars represent measured exposure field strengths in actual WTMDs. The “averaged max measured” fields are averaged over the neurostimulator lead loop area