Interference of low frequency magnetic fields with implantable cardioverter-defibrillators.

Maria Tiikkaja1, Tommi Alanko, Harri Lindholm, Maila Hietanen, Lauri Toivonen, Juha Hartikainen.   

Abstract

OBJECTIVES: The aim of this study was to find the electromagnetic interference (EMI) thresholds for several commonly used implantable cardioverter-defibrillators (ICD).
DESIGN: Seventeen ICDs were exposed to magnetic fields with different intensities produced by the Helmholtz coil system. Sinusoidal, pulse, ramp, and square-waveforms with a frequency range of 2 Hz to 1 kHz were used.
RESULTS: ICD malfunctions occurred in 11 of the 17 ICDs tested. The ICD malfunctions that occurred were false detections of ventricular tachycardia (6/17 ICDs) and ventricular fibrillation (3/17 ICDs), false detection of atrial tachycardia (4/6 dual chamber ICDs) and tachycardia sensing occurring during atrial or ventricular refractory periods (1/17 ICD). In most cases, no interference occurred at magnetic field levels below the occupational safety limits of the International Commission on Non-Ionizing Radiation Protection (ICNIRP). Nevertheless, some frequencies using sine, ramp or square waveforms did interfere with certain ICDs at levels below these limits. No EMI occurred with any of the ICDs below the ICNIRP limits for public exposure.
CONCLUSION: Evaluation of EMI should be part of the risk assessment of an employee returning to work after an ICD implantation. The risk assessment should consider magnetic field intensities, frequencies and waveforms.

Entities:  

Mesh:

Year:  2012        PMID: 22834765     DOI: 10.3109/14017431.2012.716525

Source DB:  PubMed          Journal:  Scand Cardiovasc J        ISSN: 1401-7431            Impact factor:   1.589


  4 in total

1.  Electromagnetic interference between a three-dimensional cardiac mapping system and an implantable cardioverter defibrillator.

Authors:  Alessandra Buiatti; Herribert Pavaci; Isabel Deisenhofer; Christof Kolb
Journal:  Clin Res Cardiol       Date:  2013-07-31       Impact factor: 5.460

2.  Assessment of Electromagnetic Interference with Active Cardiovascular Implantable Electronic Devices (CIEDs) Caused by the Qi A13 Design Wireless Charging Board.

Authors:  Tobias Seckler; Kai Jagielski; Dominik Stunder
Journal:  Int J Environ Res Public Health       Date:  2015-05-27       Impact factor: 3.390

3.  Personal medical electronic devices and walk-through metal detector security systems: assessing electromagnetic interference effects.

Authors:  Joshua Guag; Bisrat Addissie; Donald Witters
Journal:  Biomed Eng Online       Date:  2017-03-20       Impact factor: 2.819

4.  Electromagnetic interference in cardiac electronic implants caused by novel electrical appliances emitting electromagnetic fields in the intermediate frequency range: a systematic review.

Authors:  Sarah Driessen; Andreas Napp; Kristina Schmiedchen; Thomas Kraus; Dominik Stunder
Journal:  Europace       Date:  2019-02-01       Impact factor: 5.214

  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.