Literature DB >> 12798581

Do airport metal detectors interfere with implantable pacemakers or cardioverter-defibrillators?

Christof Kolb1, Sebastian Schmieder, Günter Lehmann, Bernhard Zrenner, Martin R Karch, Andreas Plewan, Claus Schmitt.   

Abstract

OBJECTIVES: The aim of this study was to determine whether airport metal detector gates (AMDGs) interfere with pacemakers (PMs) or implantable cardioverter-defibrillators (ICDs).
BACKGROUND: It is currently unknown whether AMDGs interfere with implanted PMs or ICDs.
METHODS: A total of 348 consecutive patients (200 PM and 148 ICD recipients) have been tested for the occurrence of electromagnetic interference (EMI) within the electromagnetic field of a worldwide-used airport metal detector.
RESULTS: No interference, such as pacing or sensing abnormalities, was observed in any of the 200 PM and 148 ICD patients; also no reprogramming occurred.
CONCLUSIONS: In vivo testing of PM and ICD systems showed no EMI with a standard AMDG. Clinically relevant interactions with implanted PMs or ICDs seem unlikely.

Entities:  

Mesh:

Year:  2003        PMID: 12798581     DOI: 10.1016/s0735-1097(03)00424-8

Source DB:  PubMed          Journal:  J Am Coll Cardiol        ISSN: 0735-1097            Impact factor:   24.094


  15 in total

1.  Experimental study on malfunction of pacemakers due to exposure to different external magnetic fields.

Authors:  Maria Tiikkaja; Tommi Alanko; Harri Lindholm; Maila Hietanen; Juha Hartikainen; Lauri Toivonen
Journal:  J Interv Card Electrophysiol       Date:  2012-01-11       Impact factor: 1.900

2.  Small bowel capsule endoscopy in patients with cardiac pacemakers and implantable cardioverter defibrillators: Outcome analysis using telemetry review.

Authors:  Justin R Cuschieri; Mohammed N Osman; Richard Ck Wong; Amitabh Chak; Gerard A Isenberg
Journal:  World J Gastrointest Endosc       Date:  2012-03-16

3.  Experiences and Attitudes of International Travelers with Cardiovascular Disease: A Qualitative Analysis.

Authors:  Chee Hwui Liew; Gerard Thomas Flaherty
Journal:  Am J Trop Med Hyg       Date:  2020-03       Impact factor: 2.345

Review 4.  Navigating air travel and cardiovascular concerns: Is the sky the limit?

Authors:  Muhammad Hammadah; Bryan R Kindya; Marc P Allard-Ratick; Sammer Jazbeh; Danny Eapen; W H Wilson Tang; Laurence Sperling
Journal:  Clin Cardiol       Date:  2017-06-08       Impact factor: 2.882

Review 5.  [Electromagnetic interference : Pacemakers, cardiac resynchronization therapy devices, implantable cardioverter-defibrillator].

Authors:  Bernd Nowak; Oliver Przibille; Andreas Napp
Journal:  Herzschrittmacherther Elektrophysiol       Date:  2022-07-04

6.  Effects of external electrical and magnetic fields on pacemakers and defibrillators: from engineering principles to clinical practice.

Authors:  Roy Beinart; Saman Nazarian
Journal:  Circulation       Date:  2013-12-24       Impact factor: 29.690

7.  Interference between pacemakers/implantable cardioverter defibrillators and video capsule endoscopy.

Authors:  Dirk Bandorski; Johannes Gehron; Reinhard Höltgen
Journal:  World J Gastrointest Endosc       Date:  2013-04-16

8.  Electromagnetic interference in implantable cardioverter defibrillators: present but rare.

Authors:  Gesa von Olshausen; Ina-Christine Rondak; Carsten Lennerz; Verena Semmler; Christian Grebmer; Tilko Reents; Sonia Ammar-Busch; Alessandra Buiatti; Felix Bourier; Isabel Deisenhofer; Christof Kolb
Journal:  Clin Res Cardiol       Date:  2016-01-29       Impact factor: 5.460

9.  Pacemaker patients' perception of unsafe activities: a survey.

Authors:  Masooma Aqeel; Azam Shafquat; Nawal Salahuddin
Journal:  BMC Cardiovasc Disord       Date:  2008-11-14       Impact factor: 2.298

10.  Capsule Endoscopy in Patients with Implantable Electromedical Devices is Safe.

Authors:  Lucinda A Harris; Stephanie L Hansel; Elizabeth Rajan; Komandoor Srivathsan; Robert Rea; Michael D Crowell; David E Fleischer; Shabana F Pasha; Suryakanth R Gurudu; Russell I Heigh; Arthur D Shiff; Janice K Post; Jonathan A Leighton
Journal:  Gastroenterol Res Pract       Date:  2013-04-03       Impact factor: 2.260

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.