Asgeir Gudnason1, Gunnar Adalberth1, Kjell-Gunnar Nilsson2, Nils P Hailer1.
Abstract
Background and purpose - Maximal total point motion (MTPM) measured by radiostereometry (RSA) is widely used as a predictor of total knee arthroplasty (TKA) loosening. We compared the ability of different RSA measurements at different time points to predict loosening of tibial TKA components in the long term. Patients and methods - 116 TKAs in 116 patients were included in our analysis. 16 (14.8-17.4) years after surgery, 5 tibial components had been revised due to aseptic loosening. Receiver operating characteristic curves were calculated in order to investigate the specificity and sensitivity of different RSA parameters at different thresholds. Results - Rotation around the transverse (x-) axis measured 2 years postoperatively had the best predictive value of all parameters, with an area under the curve (AUC) of 80%. Using a threshold of 0.8 degrees, a specificity of 85% and a sensitivity of 50% were reached. The AUC for tibial component distal translation was 79% and it was 77% for proximal translation, whereas it was only 68% for MTPM. Interpretation - Rotation of the cemented tibial component around the transverse axis, proximal translation, and distal translation are slightly better at predicting aseptic loosening than MTPM, and tibial component migration measured after 2 years gives a good prediction of aseptic loosening up to 15 years.Entities:
Mesh:
Substances:
Year: 2017 PMID: 28266241 PMCID: PMC5434596 DOI: 10.1080/17453674.2017.1297001
Source DB: PubMed Journal: Acta Orthop ISSN: 1745-3674 Impact factor: 3.717
Figure 1.Flow chart describing the 3 initial study populations and the pooled study cohort. AGC – Anatomic Graduated Components knee (Biomet). See also abbreviations in Table 1.
Demographics of the 5 patients who were revised due to aseptic loosening
| Case | Sex | BMI | Age at index operation | Years to revision | Type of prosthesis |
|---|---|---|---|---|---|
| 1 | F | 36 | 76 | 12 | FS–MB |
| 2 | M | 39 | 65 | 10 | FS–AP |
| 3 | M | 25 | 58 | 7 | FS–MB |
| 4 | M | 30 | 68 | 11 | FS–MB |
| 5 | M | 27 | 65 | 12 | AMK–CMW1 |
FS – Freeman-Samuelson knee (Biomet),
MB – metal backed tibial component,
AP – all polyethylene tibial component,
Tibial component migration and rotation measured by RSA, up to 2 years. For definitions of migration along and rotation around the 3 axes, see Patients and methods
| Not revised | Revised | ||||||||
|---|---|---|---|---|---|---|---|---|---|
| Median | IQR | Mean | 95% CI | Median | IQR | Mean | 95% CI | ||
| X-axis rotation | 4 months | 0.15 | 0.25 | 0.25 | 0.19 to 0.31 | 0.34 | 0.27 | 0.35 | 0.12 to 0.57 |
| 12months | 0.24 | 0.42 | 0.37 | 0.27 to 0.47 | 0.54 | 0.55 | 0.47 | –0.01 to 0.95 | |
| 24months | 0.32 | 0.45 | 0.46 | 0.35 to 0.57 | 0.69 | 1.05 | 0.90 | –0.05 to 1.85 | |
| Y-axis rotation | 4 months | 0.18 | 0.19 | 0.27 | 0.20 to 0.34 | 0.14 | 0.14 | 0.12 | 0.05 to 0.24 |
| 12 months | 0.21 | 0.29 | 0.28 | 0.23 to 0.34 | 0.31 | 0.24 | 0.25 | 0.02 to 0.48 | |
| 24 months | 0.22 | 0.37 | 0.30 | 0.24 to 0.36 | 0.57 | 0.82 | 0.57 | –0.11 to 1.25 | |
| Z-axis rotation | 4 months | 0.17 | 0.26 | 0.25 | 0.19 to 0.32 | 0.13 | 0.20 | 0.14 | −0.03 to 0.30 |
| 12 months | 0.23 | 0.30 | 0.32 | 0.25 to 0.39 | 0.29 | 0.68 | 0.43 | −0.19 to 1.04 | |
| 24 months | 0.27 | 0.30 | 0.37 | 0.28 to 0.45 | 0.74 | 1.15 | 0.79 | −0.22 to 1.80 | |
| MTPM | 4 months | 0.40 | 0.32 | 0.53 | 0.43 to 0.64 | 0.35 | 0.13 | 0.32 | 0.19 to 0.44 |
| 12 months | 0.53 | 0.46 | 0.64 | 0.55 to 0.73 | 0.57 | 0.41 | 0.57 | 0.21 to 0.94 | |
| 24 months | 0.56 | 0.61 | 0.70 | 0.60 to 0.81 | 0.98 | 1.52 | 1.11 | −0.16 to 2.38 | |
| Proximal translation | 4 months | 0.15 | 0.21 | 0.21 | 0.17 to 0.25 | 0.27 | 1.16 | 0.54 | −0.56 to 1.63 |
| 12 months | 0.20 | 0.27 | 0.28 | 0.22 to 0.34 | 0.37 | 0.95 | 0.67 | −0.14 to 1.47 | |
| 24 months | 0.28 | 0.37 | 0.38 | 0.31 to 0.46 | 1.01 | 0.99 | 0.84 | 0.17 to 1.51 | |
| Distal translation | 4 months | –0.18 | 0.26 | –0.24 | –0.29 to –0.19 | –0.27 | 1.46 | –0.70 | –2.11 to 0.70 |
| 12 months | –0.22 | 0.41 | –0.32 | –0.38 to –0.25 | –0.39 | 1.45 | –0.79 | –2.02 to 0.44 | |
| 24 months | –0.26 | 0.46 | –0.36 | –0.43 to –0.29 | –0.64 | 0.78 | –0.71 | –1.23 to –0.19 | |
Median differences in tibial component migration and rotation between revised and unrevised tibial components.
| 4 months | 12 months | 24 months | |||||||
|---|---|---|---|---|---|---|---|---|---|
| Median diff. | 95% CI | p-value | Median diff. | 95% CI | p-value | Median diff. | 95% CI | p-value | |
| X-axis rotation (°) | −0.16 | −0.31 to 0.05 | 0.1 | −0.23 | −0.49 to 0.18 | 0.3 | −0.38 | −1.03 to −0.03 | 0.04 |
| Y-axis rotation (°) | 0.06 | −0.06 to 0.22 | 0.3 | −0.03 | −0.22 to 0.23 | 0.8 | −0.26 | −0.71 to 0.11 | 0.1 |
| Z-axis rotation (°) | 0.04 | −0.09 to 0.26 | 0.5 | −0.06 | −0.49 to 0.19 | 0.4 | −0.23 | −1.11 to 0.07 | 0.1 |
| MTPM (mm | 0.08 | −0.09 to 0.31 | 0.4 | −0.03 | −0.32 to 0.37 | 0.8 | −0.29 | −1.13 to 0.24 | 0.2 |
| Proximal translation (mm | −0.10 | −1.23 to 0.09 | 0.3 | −0.20 | −0.55 to 0.00 | 0.053 | −0.49 | −0.91 to −0.01 | 0.04 |
| Distal translation (mm | 0.06 | −0.14 to 0.30 | 0.5 | 0.17 | −0.09 to 0.69 | 0.2 | 0.31 | 0.03 to 0.70 | 0.03 |
Sensitivity and specificity, with different thresholds of migration or rotation, for prediction of aseptic tibial component loosening. All measurements were performed 2 years after the index procedure
| Specificity, % | Sensitivity, % | ||||
|---|---|---|---|---|---|
| Threshold | 95% CI | 95% CI | |||
| X-axis rotation | 0.2° | 39 | 28–49 | 100 | 100–100 |
| 0.6° | 77 | 67–86 | 50 | 0–100 | |
| 0.8° | 85 | 76–92 | 50 | 0–100 | |
| Y-axis rotation | 0.5° | 78 | 69–87 | 50 | 0–100 |
| 0.7° | 90 | 82–96 | 50 | 0–100 | |
| 0.9° | 96 | 91–100 | 25 | 0–75 | |
| Z-axis rotation | 0.7° | 86 | 78–94 | 50 | 0–100 |
| 0.9° | 92 | 86–97 | 50 | 0–100 | |
| 1.1° | 96 | 91–100 | 50 | 0–100 | |
| MTPM | 1.0 mm | 76 | 67–85 | 50 | 0–100 |
| 1.2 mm | 87 | 80–94 | 50 | 0–100 | |
| 1.4 mm | 91 | 85–96 | 25 | 0–75 | |
| Proximal translation | 0.7 mm | 87 | 80–95 | 60 | 20–100 |
| 0.9 mm | 91 | 84–96 | 60 | 20–100 | |
| 1.1 mm | 96 | 92–99 | 20 | 0–60 | |
| Distal translation | −0.4 mm | 65 | 55–75 | 60 | 20–100 |
| −0.6 mm | 77 | 68–86 | 60 | 20–100 | |
| −0.8 mm | 89 | 82–95 | 40 | 0–80 | |
Figure 2.ROC curve for tibial component rotation around the transverse (x-) axis as measured by RSA after 2 years, showing thresholds at 0.2° and 0.8°. The blue error bars indicate the 95% confidence intervals around the specificity and sensitivity for the indicated thresholds of 0.2° and 0.8°.
Figure 3.ROC curve for maximal total point motion (MTPM) of the tibial component as measured by RSA after 2 years, showing threshold at 1.0 mm. The blue error bar indicates the 95% confidence intervals around the specificity and sensitivity for the indicated threshold of 1 mm.